Waveform Digitizing for LaBr$_3$/NaI Phoswich Detector

Rui She, Jianfeng Ji, Feng Hua, Hong Li, Han Dong

Abstract

The detection efficiency of phoswich detectors starts to decrease when Compton scattering becomes significant. Events with energy deposit in both scintillators, if not rejected, are not useful for spectral analysis as the full energy of the incident photon cannot be reconstructed with conventional readout. We show that once the system response is carefully calibrated, the full energy of those double deposit events can be reconstructed using a waveform digitizer as the readout. Our experiment suggests that the efficiency of the photopeak at 662 keV can be increased by a factor of 2 given our LaBr$_3$/NaI phoswich detector.

Introduction

In X-ray and gamma-ray astronomy, phoswich (phosphor sandwich) detectors are essential tools. They typically consist of two stacked scintillation crystals—a thin sensitive layer and a thick veto layer—each with distinct light decay times. Traditionally, photons that undergo Compton scattering and deposit energy in both layers produce overlapping signals that standard electronics cannot easily disentangle. Consequently, these "double deposit" events are usually rejected to maintain data purity. However, at energies above 200–300 keV, Compton scattering becomes the dominant interaction mechanism. Rejecting these events creates a significant "blind spot" for high-energy astrophysical phenomena, such as the 511 keV Galactic electron-positron annihilation line.

Methods

The researchers employed a phoswich pair consisting of $\text{LaBr}_3$:Ce (fast decay, ~16 ns) and $\text{NaI}$(Tl) (slow decay, ~250 ns). By replacing traditional peak-sensing electronics with a high-speed waveform digitizer (1 GS/s), they could capture the raw pulse shapes of every event.

The core of the method involves modeling the output signal as a convolution of the light decay and the measured system response ($R_s$). For events hitting both crystals, a two-component exponential fit is applied: $$I = I_{1,0} e^{-t/\tau_1} + I_{2,0} e^{-t/\tau_2}$$ By fixing the known decay constants ($\tau_1$ and $\tau_2$), the energy deposited in each layer ($I_{1,0}$ and $I_{2,0}$) can be determined and summed to recover the total incident photon energy.

Figure 6: Waveform fitting of a double-deposit event, showing the total fit (solid) and the individual fast and slow scintillation components (dashed).

Key Insights and Results

A critical insight was the use of pulse shape parameters to classify events. By plotting the Full Width at Half Maximum (FWHM) against the pulse amplitude, the team could distinguish between pure $\text{LaBr}_3$ events, pure $\text{NaI}$ events, and the "bridge" of Compton events connecting them. This classification allows for the targeted reconstruction of mixed-deposit events that would otherwise be discarded.

Figure 4: Scatter plot of FWHM vs. Peak channel, showing the clear separation between pure LaBr3 events, pure NaI events, and the 'bridge' of mixed-deposit events.

Using standard radioactive sources like $^{137}\text{Cs}$ (662 keV), the team demonstrated that the detection efficiency for the full energy peak increased by a factor of 1.9 to 2.0. This experimental result aligns with GEANT4 simulations, which predicted that capturing double-deposit events would roughly double the effective area at these energies.

Conclusion

This work demonstrates that with careful calibration and waveform digitizing, the "veto" layer of a phoswich detector can be transformed into an active part of the calorimeter. While this approach requires more computational power for real-time fitting, it offers a path to significantly enhance the scientific return of future space missions by extending the effective energy range of phoswich-based instruments.

Summary updated

Abstract

The detection efficiency of phoswich detector starts to decrease when Compton scattering becomes significant. Events with energy deposit in both scintillators, if not rejected, are not useful for spectral analysis as the full energy of the incident photon cannot be reconstructed with conventional readout. We show that once the system response is carefully calibrated, the full energy of those double deposit events can be reconstructed using a waveform digitizer as the readout. Our experiment suggests that the efficiency of photopeak at 662 keV can be increased by a factor of 2 given our LaBr$_3$/NaI phoswich detector.
BibTeX
@inproceedings{she2014waveform,
  author = {Rui She and Jianfeng Ji and Feng Hua and Hong Li and Han Dong},
  title = {{Waveform Digitizing for LaBr$\_3$/NaI Phoswich Detector}},
  booktitle = {Chinese},
  year = {2014},
  doi = {10.1088/1674-1137/38/8/086005},
}