Waveform digitizing for LaBr 3 /NaI phoswich detector

Rui She, Jianfeng Ji, Hua Feng, Hong Li, Dong Han

TL;DR

This paper demonstrates a novel method for reconstructing the full energy of X-ray and gamma-ray events in a LaBr₃/NaI phoswich detector. By using a waveform digitizer, they can now analyze events where the photon's energy is deposited in both detector layers, significantly increasing the detector's overall efficiency for high-energy photons.

What Problem Does This Paper Address?

Phoswich detectors, combining two scintillators with different decay times (like LaBr₃ and NaI), are excellent for X-ray astronomy. Typically, the outer layer acts as a veto to reject background. However, for high-energy photons (above ~200-300 keV), Compton scattering can cause the photon's energy to be deposited in both layers. Conventionally, these "double-deposit" events are discarded because their differing signal decay times make it impossible to accurately determine the total energy deposited. This limits the useful energy range of phoswich detectors, potentially missing important astrophysical signals like electron-positron annihilation at 511 keV or high-energy gamma-ray bursts.

Methods

The core innovation here is the use of a waveform digitizer (CAEN V1742) sampling at 1 GS/s. Instead of just recording the peak amplitude of the signal, this digitizer captures the entire pulse shape over time. The authors then developed a method to fit these digitized waveforms. They model the signal as a sum of two exponential decays, corresponding to the LaBr₃ and NaI layers, with their known decay times (16 ns and 250 ns, respectively) fixed. By fitting this model to the measured waveform, they can disentangle the energy deposited in each layer and sum them to reconstruct the photon's total energy. This process requires careful calibration of the system's response.

Datasets

The experiments were conducted using a ¹³⁷Cs radioactive source, which emits gamma-rays at 662 keV. This source was used to generate events for calibration and spectral analysis.

Key Insights and Ideas

  • Waveform Digitization is Key: Moving beyond simple charge integration to full waveform analysis unlocks the potential of phoswich detectors for higher energies.
  • Double-Deposit Events are Recoverable: Events previously considered "unusable" due to energy sharing between layers can now be analyzed.
  • Improved High-Energy Efficiency: By including these recovered events, the detector's ability to register full-energy photopeaks at higher energies is substantially boosted.
  • Trade-off in Resolution: While efficiency increases, there's a slight degradation in energy resolution when including double-deposit events, likely due to the poorer resolution of the NaI layer compared to LaBr₃.

Key Results

The authors demonstrated that by using waveform digitizing and fitting, they could reconstruct the full energy of ¹³⁷Cs (662 keV) events where energy was deposited in both the LaBr₃ and NaI layers. This method increased the efficiency of detecting the 662 keV photopeak by a factor of approximately 1.9, which is consistent with their GEANT4 simulations. They also showed that the energy resolution for these reconstructed events was slightly worse than for events fully contained within the LaBr₃ layer.

Limitations and Future Work

The paper acknowledges that the energy resolution is slightly degraded when including the double-deposit events. Future work could focus on optimizing the fitting algorithms or exploring detector configurations that might mitigate this resolution loss. The authors suggest that this improved high-energy response could significantly broaden the scientific return of phoswich detectors in X-ray and gamma-ray astronomy, enabling new discoveries in areas like transient events and high-energy astrophysical phenomena.

Summary updated

Abstract

The detection efficiency of phoswich detector starts to decrease when Compton scattering becomes significant. Events with energy deposit in both scintillators, if not rejected, are not useful for spectral analysis as the full energy of the incident photon cannot be reconstructed with conventional readout. We show that once the system response is carefully calibrated, the full energy of those double deposit events can be reconstructed using a waveform digitizer as the readout. Our experiment suggests that the efficiency of photopeak at 662 keV can be increased by a factor of 2 given our LaBr3/NaI phoswich detector.
BibTeX
@inproceedings{she2014waveform,
  author = {Rui She and Jianfeng Ji and Hua Feng and Hong Li and Dong Han},
  title = {{Waveform digitizing for LaBr <sub>3</sub> /NaI phoswich detector}},
  booktitle = {Chinese},
  year = {2014},
  doi = {10.1088/1674-1137/38/8/086005},
}
Figure