Toward Robotically Automated Femoral Vascular Access

Nico Zevallos, Evan Harber, Abhimanyu, Kirtan Patel, Yizhu Gu, Kenny Sladick, Francis Guyette, Leonard Weiss, Michael R. Pinsky, Hernando Gomez, John M. Galeotti, Howie Choset

Abstract

Advanced resuscitative technologies, such as Extra Corporeal Membrane Oxygenation (ECMO) cannulation or Resuscitative Endovascular Balloon Occlusion of the Aorta (REBOA), are technically difficult even for skilled medical personnel. This paper describes the core technologies that comprise a teleoperated system capable of granting femoral vascular access, which is an important step in both of these procedures and a major roadblock in their wider use in the field. These technologies include a kinematic manipulator, various sensing modalities, and a user interface. In addition, we evaluate our system on a surgical phantom as well as in-vivo porcine experiments. These resulted in, to the best of our knowledge, the first robot-assisted arterial catheterizations; a major step towards our eventual goal of automatic catheter insertion through the Seldinger technique.

Introduction

In emergency medicine, procedures like ECMO and REBOA are vital for patient survival but are often limited by the difficulty of obtaining femoral vascular access. This access typically requires the Seldinger technique, a multi-step process involving ultrasound-guided needle insertion, guidewire threading, and catheter placement.

The Seldinger technique: (a) needle insertion, (b) guidewire introduction, (c) needle removal, (d) dilator/sheath advancement, (e) sheath placement, and (f) final catheter position.

The femoral artery is particularly challenging due to its depth (often 2–8 cm) and the variability of the surrounding anatomy. While automated systems exist for superficial veins, deep arterial access in austere environments remains a significant hurdle. This research presents a robotic system designed to assist clinicians in performing these high-stakes procedures with greater precision and reliability.

System Design and Methods

The developed system utilizes a "single-manipulator" approach, integrating both imaging and insertion capabilities into a unified robotic platform.

System architecture overview, highlighting the UR3e robotic arm, the custom end effector with linear and rotary actuators, and the integrated ultrasound probe and force sensor.

Hardware and Sensing

The core of the system is a 6-DOF Universal Robotics UR3e arm equipped with a custom 2-DOF end effector. This end effector features a linear actuator for controlling needle depth and a curved guide track for adjusting the insertion angle. To ensure high-quality ultrasound imaging, an ATI Nano25 6-axis force sensor is positioned between the robot and the ultrasound probe, allowing for constant, controlled pressure on the tissue.

Control and User Interface

The system is operated via a hybrid control scheme:

  • Image-space Controller: Translates user clicks on the ultrasound interface into robotic movements to center the target vessel.
  • Needle Tweak Controller: Provides a manual override for real-time adjustments, compensating for tissue deformation or needle deflection.
  • Perception Pipeline: Currently relies on user-selected targets to guide the robot's autonomous centering and puncture maneuvers.

Experimental Evaluation

The system underwent rigorous testing in two distinct environments:

  • Phantom Trials: Using a Blue Phantom Generation II model, three non-medical users achieved a 100% success rate in guidewire insertion, with completion times between 4 and 6 minutes.
  • In-Vivo Porcine Trials: The system was validated in three live pig experiments. These trials successfully demonstrated the entire sequence: needle puncture, guidewire placement, and catheter threading. Notably, no complications such as hematomas or lacerations occurred, despite the operators being engineers rather than trained surgeons.

Discussion and Future Work

This work demonstrates that a teleoperated robotic system can successfully bridge the gap between manual expertise and automated vascular access. By focusing on the "rate-limiting step" of initial cannulation, the system enables advanced resuscitative treatments in environments where expert clinicians may not be available.

Future developments will focus on increasing autonomy through deep learning-based vessel segmentation and 3D anatomical reconstruction. Additionally, the team plans to automate the guidewire insertion phase and integrate advanced sensors to detect vessel entry in low-blood-pressure scenarios where traditional visual cues are absent.

Summary updated

Abstract

Advanced resuscitative technologies, such as Extra Corporeal Membrane Oxygenation (ECMO) cannulation or Resuscitative Endovascular Balloon Occlusion of the Aorta (REBOA), are technically difficult even for skilled medical personnel. This paper describes the core technologies that comprise a teleoperated system capable of granting femoral vascular access, which is an important step in both of these procedures and a major roadblock in their wider use in the field. These technologies include a kinematic manipulator, various sensing modalities, and a user interface. In addition, we evaluate our system on a surgical phantom as well as in-vivo porcine experiments. These resulted in, to the best of our knowledge, the first robot-assisted arterial catheterizations; a major step towards our eventual goal of automatic catheter insertion through the Seldinger technique.
BibTeX
@inproceedings{zevallos2021toward,
  author = {Nico Zevallos and Evan Harber and Abhimanyu and Kirtan Patel and Yizhu Gu and Kenny Sladick and Francis Guyette and Leonard Weiss and Michael R. Pinsky and Hernando Gomez and John M. Galeotti and Howie Choset},
  title = {{Toward Robotically Automated Femoral Vascular Access}},
  booktitle = {ISMR},
  year = {2021},
  url = {https://openreview.net/forum?id=hQE-9-al9Bo},
}
Figure