Toward Robotically Automated Femoral Vascular Access

Nico Zevallos, Evan Harber, Abhimanyu, Kirtan Patel, Yizhu Gu, Kenny Sladick, Francis Guyette, Leonard Weiss, Michael R. Pinsky, Hernando Gomez, John M. Galeotti, Howie Choset

TL;DR

Researchers have developed a teleoperated robotic system designed to perform the Seldinger technique—a complex medical procedure for accessing deep blood vessels. By combining a robotic arm with a custom end effector and ultrasound guidance, the team achieved the first successful robot-assisted arterial catheterizations in live porcine models, marking a major step toward fully autonomous emergency vascular access.

What Problem Does This Paper Address?

In emergency medicine, procedures like ECMO (Extra Corporeal Membrane Oxygenation) or REBOA (stopping internal bleeding with a balloon) are life-savers, but they have a major bottleneck: femoral vascular access.

To perform these, a clinician must use the "Seldinger technique"—inserting a needle into the femoral artery under ultrasound, threading a guidewire through it, and then sliding a catheter over that wire. This is incredibly difficult even for experts, especially in "austere environments" (like a battlefield or a chaotic ER). Current automated systems for drawing blood work well for shallow veins in the arm, but the femoral artery is deep (often buried under 2–8 cm of fat and tissue) and sits in a highly variable anatomical region. This paper aims to bridge that gap with a robotic system that can handle the depth, tissue deformation, and precision required for the groin.

Methods

The authors developed a "single-manipulator" approach that handles both the "eyes" (ultrasound) and the "hands" (needle insertion).

  • Hardware: They mounted a custom 2-degree-of-freedom (DOF) end effector onto a 6-DOF Universal Robotics UR3e arm. The end effector includes a linear actuator for needle depth and a curved guide track for adjusting the insertion angle.
  • Sensing: A 6-axis force sensor (ATI Nano25) sits between the robot and the ultrasound probe. This allows the robot to maintain constant, gentle pressure on the skin, which is critical for clear ultrasound imaging.
  • Control Software:
    • Image-space controller: This is the "bridge" between the user and the robot. A user can click a point on the ultrasound screen, and the system calculates the necessary robot movements to center the vessel or aim the needle.
  • Needle Tweak Controller: Recognizing that tissue shifts and needles bend, they included a manual "tweak" mode that allows the operator to fine-tune the needle's position in real-time based on visual feedback.
  • The Procedure: The robot scans the area, the user selects the target vessel on a GUI, the robot centers the vessel in the image, and then the user triggers the puncture.

Datasets

The system was evaluated using two primary setups:

  1. Blue Phantom Generation II: A high-fidelity silicone training model of the human lower torso and femoral region used to test repeatability and user interface efficiency.
  2. In-vivo Porcine Models: Three live pig experiments were conducted to validate the system in a dynamic, "real-world" biological environment where vessels roll, deform, and tissue is non-uniform.

Key Insights and Ideas

  • Hybrid Automation: The paper argues that we don't need full autonomy yet. By creating a "teleoperated" system where a human makes the high-level decisions (like picking the vessel) and the robot handles the precision (steadying the probe and hitting the target), they can achieve clinical success today.
  • The "Rate-Limiting Step": The authors identify that the most difficult part of advanced resuscitation isn't the surgery itself, but the initial safe cannulation of the artery. Solving this one step unlocks a suite of other advanced treatments.
  • Anatomical Challenges: They found that porcine anatomy is actually more curved and difficult than human anatomy, suggesting that success in these animal models is a strong indicator of potential success in humans.

Key Results

  • 100% Success in Phantoms: Three users with no medical training were able to successfully insert the guidewire into the phantom artery across multiple trials. Average times ranged from roughly 4 to 6 minutes.
  • First In-Vivo Success: The system successfully performed arterial catheterization in three live pigs. This included the needle puncture, guidewire placement, and catheter threading.
  • Safety: No complications, such as lacerations or hematomas, were observed during the porcine trials, even though the operators were engineers rather than surgeons.

Limitations and Future Work

  • Manual Guidewire Insertion: While the needle puncture is robot-assisted, the operator still has to manually feed the guidewire through the needle. Future versions aim to automate the entire Seldinger sequence.
  • Computer Vision: Currently, the user must manually identify the vessel. The team plans to integrate deep learning for autonomous vessel segmentation and tracking.
  • Puncture Detection: In trauma patients with very low blood pressure, the traditional "flash" of blood in the needle might not happen. The authors suggest adding force sensors or bio-impedance sensors to "feel" when the needle has entered the vessel lumen.
  • 3D Modeling: To move toward full autonomy, the system will need to reconstruct a 3D model of the vessel and skin surface rather than relying on 2D ultrasound slices.

Summary updated

Abstract

Advanced resuscitative technologies, such as Extra Corporeal Membrane Oxygenation (ECMO) cannulation or Resuscitative Endovascular Balloon Occlusion of the Aorta (REBOA), are technically difficult even for skilled medical personnel. This paper describes the core technologies that comprise a teleoperated system capable of granting femoral vascular access, an essential step in these procedures, and a significant roadblock in their broader use in the field. These technologies include a kinematic manipulator, various sensing modalities, and a user interface. In addition, we evaluate our system on a surgical phantom as well as in-vivo porcine experiments. To the best of our knowledge, these resulted in the first robot-assisted arterial catheterizations, a significant step towards our eventual goal of automatic catheter insertion through the Seldinger technique.