Three-dimensional position reconstruction of orthogonal-strip planar high-purity germanium detectors using maximum likelihood estimation

Qiuli Zhang, Peng Zhang, W. H. Dai, Mingxin Yang, Yang Tian, Ming Zeng, Hao Ma, Zhi Zeng

Abstract: Orthogonal-strip planar high-purity germanium (HPGe) detectors can reconstruct three-dimensional (3D) positions of photon interactions through analysis of parameters extracted from multiple charge signals. The conventional method independently reconstructs positions in each dimension using amplitude-based parameters, leading to noise sensitivity and systematic biases. In this study, we propose a multi-parameter-joint reconstruction method based on maximum likelihood estimation (MLE) which establishes a mapping between pulse shape parameters and corresponding 3D positions. To mitigate the effects of electronic noise, we employ integral-based parameters. The reconstruction performance was evaluated using pulse shape simulations. For 100 keV photons under 1 keV root-mean-square (RMS) electronic noise, the maximum Z reconstruction bias was reduced from 0.4 mm to 0.02 mm in the central region and from 2 mm to 0.15 mm near the electrodes. The maximum reconstruction bias in the X/Y directions was reduced from 0.4 mm to 0.016 mm. Furthermore, the use of integral-based parameters mitigated the rapid degradation of resolution under high-noise conditions. The achieved position resolution ranged from 0.07 mm to 0.16 mm in the Z directions and from 0.07 mm to 0.44 mm in the X/Y direction. This method offers a promising approach to 3D position reconstruction with HPGe detectors for applications such as medical imaging and gamma-ray astronomy.

Introduction

Orthogonal-strip planar High-Purity Germanium (HPGe) detectors are prized in medical imaging and gamma-ray astronomy for their exceptional energy resolution and spatial sensitivity. By utilizing a grid of orthogonal strips on the cathode and anode, these detectors can pinpoint the 3D coordinates of photon interactions. However, traditional reconstruction algorithms face two major hurdles: cross-dimensional interference and high sensitivity to electronic noise. Standard methods often treat X, Y, and Z dimensions as independent, ignoring the physical reality that lateral signals influence depth measurements, resulting in significant systematic spatial warping.

Detector Geometry and Simulation

The study focuses on a cylindrical HPGe detector with a 30 mm diameter and 14 mm thickness. The device features eight orthogonal strips on both the cathode and anode surfaces, separated by 0.5 mm gaps with a 3 mm pitch.

Schematic of the cylindrical orthogonal-strip planar HPGe detector.

To evaluate the proposed method, the researchers utilized the SolidStateDetectors.jl package to simulate electric fields and charge carrier drift. Interactions were modeled at 100 keV with Gaussian electronic noise ranging from 1 keV to 3 keV RMS added to the resulting pulses.

Methodology: Multi-Parameter Joint Reconstruction

The proposed solution moves away from independent-dimension calculations in favor of a joint Maximum Likelihood Estimation (MLE) approach. This method establishes a statistical mapping between a set of pulse shape parameters and the most probable 3D coordinate $(x, y, z)$.

Key to this approach is the use of integral-based parameters instead of traditional peak amplitudes. By integrating the net area of charge pulses over a 300 ns window, the algorithm effectively applies a low-pass filter, making it far more robust against high-frequency electronic noise. The MLE model utilizes four specific parameters ($S_C, S_I, S_{AX}, S_{AY}$) derived from charge collection and image signals.

Detailed view of charge collection and image signals used for parameter extraction.

By modeling these signals as a multivariate normal distribution, the MLE framework can "correct" for the interference between dimensions. For example, it accounts for how the relationship between collection time and depth varies based on the lateral (X/Y) position of the interaction.

Key Results

The multi-parameter joint reconstruction achieved a nearly 20-fold reduction in reconstruction bias compared to conventional methods under 1 keV RMS noise:

  • Z-Direction Bias: Reduced from 0.4 mm to 0.02 mm in the central region, and from 2 mm to 0.15 mm near the electrodes.
  • X/Y-Direction Bias: Slashed from 0.4 mm to 0.016 mm.
  • Spatial Resolution: Achieved resolutions of 0.07–0.16 mm in Z and 0.07–0.44 mm in X/Y.
  • Noise Robustness: While conventional amplitude-based methods saw error rates explode exponentially as noise increased to 3 keV, the MLE method's resolution degraded linearly and predictably.

Conclusion

This MLE-based reconstruction method significantly improves the spatial accuracy of HPGe detectors, particularly in high-noise environments or near electrode edges where signal behavior is non-linear. While currently validated through high-fidelity simulations, future work will focus on experimental validation with physical hardware to account for real-world variables such as crystal lattice anisotropy and localized impurities.

Summary updated

Abstract

Orthogonal-strip planar high-purity germanium (HPGe) detectors can reconstruct three-dimensional (3D) positions of photon interactions through analysis of parameters extracted from multiple charge signals. The conventional method independently reconstructs positions in each dimension using amplitude-based parameters, leading to noise sensitivity and systematic biases. In this study, we propose a multi-parameter-joint reconstruction method based on maximum likelihood estimation (MLE) which establishes a mapping between pulse shape parameters and corresponding 3D positions. To mitigate the effects of electronic noise, we employ integral-based parameters. The reconstruction performance was evaluated using pulse shape simulations. For 100 keV photons under 1 keV root-mean-square (RMS) electronic noise, the maximum Z reconstruction bias was reduced from 0.4 mm to 0.02 mm in the central region and from 2 mm to 0.15 mm near the electrodes. The maximum reconstruction bias in the X/Y directions was reduced from 0.4 mm to 0.016 mm. Furthermore, the use of integral-based parameters mitigated the rapid degradation of resolution under high-noise conditions. The achieved position resolution ranged from 0.07 mm to 0.16 mm in the Z directions and from 0.07 mm to 0.44 mm in the X/Y direction. This method offers a promising approach to 3D position reconstruction with HPGe detectors for applications such as medical imaging and gamma-ray astronomy.
BibTeX
@article{zhang2025three,
  author = {Qiuli Zhang and Peng Zhang and W. H. Dai and Mingxin Yang and Yang Tian and Ming Zeng and Hao Ma and Zhi Zeng},
  title = {{Three-dimensional position reconstruction of orthogonal-strip planar high-purity germanium detectors using maximum likelihood estimation}},
  journal = {Journal of Instrumentation},
  year = {2025},
  doi = {10.1088/1748-0221/20/10/p10028},
}

Loading related content...

Figure