Abstract
As a new member of the GECAM mission, GECAM-C (also named High Energy Burst Searcher, HEBS) was launched onboard the SATech-01 satellite on July 27th, 2022. It is capable of monitoring gamma-ray transients from $\sim$ 6 keV to 6 MeV. The primary instrumentation consists of 12 gamma-ray detectors (GRDs). To verify detector performance and validate Monte Carlo simulations of detector response, comprehensive on-ground calibration experiments were performed using X-ray beams and radioactive sources. These tests assessed energy-channel relations, energy resolution, detection efficiency, SiPM voltage-gain relations, and positional response uniformity. The results demonstrate the excellent performance of the GECAM-C GRD detectors.
Introduction
When a satellite like GECAM-C is launched into orbit to detect Gamma-Ray Bursts (GRBs) or gravitational wave counterparts, it records electronic pulses that must be translated into physical energy values. This translation requires a "Response Matrix"—a complex mathematical map describing how the detector reacts to different energies and incident angles. Because scintillators have non-linear responses and hardware can exhibit subtle spatial inconsistencies, rigorous "pre-flight" calibration is essential to ensure the scientific validity of data sent back from its 500 km sun-synchronous orbit.
Instrument Design and Layout
The GECAM-C instrument utilizes 12 GRDs distributed across two domes (top and bottom) to provide wide-field coverage. These detectors are categorized into two types: NGRDs (based on Sodium Iodide, NaI(Tl)) and LGRDs (based on Lanthanum Bromide, $\text{LaBr}_3$).
Both detector types employ Silicon Photomultiplier (SiPM) arrays for readout, offering a compact and robust alternative to traditional photomultiplier tubes. The internal structure includes a beryllium window for low-energy transparency, the scintillator crystal, and a quartz window coupling the crystal to the SiPM board.
Calibration Methodology
The calibration campaign was split into two primary phases:
- High-Gain Channel (Low Energy): Conducted at the Hard X-ray Calibration Facility (HXCF) using monochromatic X-ray beams (5–140 keV). This phase established the "Energy–Channel" relation and tested spatial uniformity via a 25-point grid scan.
- Low-Gain Channel (High Energy): Performed at IHEP using radioactive sources (e.g., $^{137}\text{Cs}$, $^{60}\text{Co}$) to characterize responses up to several MeV.
Experimental results were used to validate a Geant4 Monte Carlo simulation. Comparison via Kolmogorov-Smirnov (K-S) tests confirmed that the digital model accurately represents the physical detectors, allowing for the generation of reliable response matrices for any scenario encountered in orbit.
Key Results and Insights
- Energy Resolution and Efficiency: All detectors met design specifications, achieving energy resolution better than 18% at 59.5 keV and intrinsic efficiency exceeding 75% at 15 keV.
- Non-Proportional Response (nPR): Calibration revealed distinct "dips" in light output near the K-shell binding energies of Iodine, Bromine, and Lanthanum. Accounting for this non-linearity is critical for precise energy reconstruction.
- SiPM Adaptability: The study confirmed that SiPM gain can be finely tuned via bias voltage, allowing ground control to compensate for thermal drifts or radiation-induced degradation over time.
- Spatial Uniformity: Surface scans showed a Relative Standard Deviation (RSD) of less than 4%, indicating high-quality crystals with minimal edge degradation.
Conclusion
The ground calibration of GECAM-C has successfully verified the performance of its 12 GRDs across a wide dynamic range (~6 keV to 6 MeV). While ground-based relations provide a baseline, in-flight calibration using characteristic background lines (like the 511 keV positron annihilation line) will be used to maintain precision throughout the mission's lifetime. GECAM-C has already demonstrated its capabilities by detecting significant events, including the record-breaking GRB 221009A.