Ground calibration of Gamma-Ray Detectors of GECAM-C

Chao Zheng, Zhenghua An, Wen-Xi Peng, Dali Zhang, S. L. Xiong, Rui. Qiao, Yan-Qiu Zhang, W. C. Xue, Jia-Cong Liu, Pei-Yi Feng, Ce. Cai, Gao Min, Ke Gong, D. Y. Guo, Dongjie Hou, Gang Li, Xinqiao Li, Yanguo Li, Mao-Shun Li, Xiaohua Liang, Yaqing Liu, Xiao-Jing Liu, L. M. Song, Xilei Sun, Wen-Jun Tan, Chen-Wei Wang, Hui Wang, Jin-Zhou Wang, Xiang-Yang Wen, Shuo Xiao, Yan-Bing Xu, S. Yang, Qi-Bing Yi, Fan Zhang, Peng Zhang, Zhen Zhang, Yi Zhao, Xing ZhouarXiv 2023 · DOI · arXiv

Abstract

As a new member of the GECAM mission, GECAM-C (also named High Energy Burst Searcher, HEBS) was launched onboard the SATech-01 satellite on July 27th, 2022. It is capable of monitoring gamma-ray transients from $\sim$ 6 keV to 6 MeV. The primary instrumentation consists of 12 gamma-ray detectors (GRDs). To verify detector performance and validate Monte Carlo simulations of detector response, comprehensive on-ground calibration experiments were performed using X-ray beams and radioactive sources. These tests assessed energy-channel relations, energy resolution, detection efficiency, SiPM voltage-gain relations, and positional response uniformity. The results demonstrate the excellent performance of the GECAM-C GRD detectors.

Introduction

When a satellite like GECAM-C is launched into orbit to detect Gamma-Ray Bursts (GRBs) or gravitational wave counterparts, it records electronic pulses that must be translated into physical energy values. This translation requires a "Response Matrix"—a complex mathematical map describing how the detector reacts to different energies and incident angles. Because scintillators have non-linear responses and hardware can exhibit subtle spatial inconsistencies, rigorous "pre-flight" calibration is essential to ensure the scientific validity of data sent back from its 500 km sun-synchronous orbit.

Instrument Design and Layout

The GECAM-C instrument utilizes 12 GRDs distributed across two domes (top and bottom) to provide wide-field coverage. These detectors are categorized into two types: NGRDs (based on Sodium Iodide, NaI(Tl)) and LGRDs (based on Lanthanum Bromide, $\text{LaBr}_3$).

Layout of the 12 GRDs on the top and bottom domes of GECAM-C.

Both detector types employ Silicon Photomultiplier (SiPM) arrays for readout, offering a compact and robust alternative to traditional photomultiplier tubes. The internal structure includes a beryllium window for low-energy transparency, the scintillator crystal, and a quartz window coupling the crystal to the SiPM board.

Internal module structure and schematic diagram of a GECAM-C GRD.

Calibration Methodology

The calibration campaign was split into two primary phases:

  1. High-Gain Channel (Low Energy): Conducted at the Hard X-ray Calibration Facility (HXCF) using monochromatic X-ray beams (5–140 keV). This phase established the "Energy–Channel" relation and tested spatial uniformity via a 25-point grid scan.
  2. Low-Gain Channel (High Energy): Performed at IHEP using radioactive sources (e.g., $^{137}\text{Cs}$, $^{60}\text{Co}$) to characterize responses up to several MeV.

Experimental results were used to validate a Geant4 Monte Carlo simulation. Comparison via Kolmogorov-Smirnov (K-S) tests confirmed that the digital model accurately represents the physical detectors, allowing for the generation of reliable response matrices for any scenario encountered in orbit.

Key Results and Insights

  • Energy Resolution and Efficiency: All detectors met design specifications, achieving energy resolution better than 18% at 59.5 keV and intrinsic efficiency exceeding 75% at 15 keV.
  • Non-Proportional Response (nPR): Calibration revealed distinct "dips" in light output near the K-shell binding energies of Iodine, Bromine, and Lanthanum. Accounting for this non-linearity is critical for precise energy reconstruction.
  • SiPM Adaptability: The study confirmed that SiPM gain can be finely tuned via bias voltage, allowing ground control to compensate for thermal drifts or radiation-induced degradation over time.
  • Spatial Uniformity: Surface scans showed a Relative Standard Deviation (RSD) of less than 4%, indicating high-quality crystals with minimal edge degradation.

Conclusion

The ground calibration of GECAM-C has successfully verified the performance of its 12 GRDs across a wide dynamic range (~6 keV to 6 MeV). While ground-based relations provide a baseline, in-flight calibration using characteristic background lines (like the 511 keV positron annihilation line) will be used to maintain precision throughout the mission's lifetime. GECAM-C has already demonstrated its capabilities by detecting significant events, including the record-breaking GRB 221009A.

Summary updated

Abstract

As a new member of GECAM mission, GECAM-C (also named High Energy Burst Searcher, HEBS) was launched onboard the SATech-01 satellite on July 27th, 2022, which is capable to monitor gamma-ray transients from $\sim$ 6 keV to 6 MeV. As the main detector, there are 12 gamma-ray detectors (GRDs) equipped for GECAM-C. In order to verify the GECAM-C GRD detector performance and to validate the Monte Carlo simulations of detector response, comprehensive on-ground calibration experiments have been performed using X-ray beam and radioactive sources, including Energy-Channel relation, energy resolution, detection efficiency, SiPM voltage-gain relation and the non-uniformity of positional response. In this paper, the detailed calibration campaigns and data analysis results for GECAM-C GRDs are presented, demonstrating the excellent performance of GECAM-C GRD detectors.
BibTeX
@misc{zheng2023ground,
  author = {Chao Zheng and Zhenghua An and Wen-Xi Peng and Dali Zhang and S. L. Xiong and Rui. Qiao and Yan-Qiu Zhang and W. C. Xue and Jia-Cong Liu and Pei-Yi Feng and Ce. Cai and Gao Min and Ke Gong and D. Y. Guo and Dongjie Hou and Gang Li and Xinqiao Li and Yanguo Li and Mao-Shun Li and Xiaohua Liang and Yaqing Liu and Xiao-Jing Liu and L. M. Song and Xilei Sun and Wen-Jun Tan and Chen-Wei Wang and Hui Wang and Jin-Zhou Wang and Xiang-Yang Wen and Shuo Xiao and Yan-Bing Xu and S. Yang and Qi-Bing Yi and Fan Zhang and Peng Zhang and Zhen Zhang and Yi Zhao and Xing Zhou},