From Bench to Bedside: The First Live Robotic Surgery on the dVRK to Enable Remote Telesurgery with Motion Scaling

Florian Richter, Emily K. Funk, Won Seo Park, Ryan K. Orosco, Michael C. Yip

TL;DR

This paper reports the first-ever live animal surgery using the da Vinci Research Kit (dVRK), successfully demonstrating the translation of motion scaling techniques for remote telesurgery from the lab to a real-world scenario. This work addresses the critical gap between robotic surgery research and clinical application by tackling the challenges of live tissue manipulation and communication latency.

What Problem Does This Paper Address?

Surgical robotics research has seen significant advancements, but a major hurdle is translating these innovations from controlled lab environments to live surgeries. This "bench-to-bedside" gap is substantial, as live surgeries introduce complexities like surgical staff coordination, intricate procedures, and the unpredictable nature of living tissue. The da Vinci Research Kit (dVRK), an open-source platform widely used by researchers, has been available for a decade, yet no live surgeries have been performed with it. This paper aims to bridge this gap by demonstrating how motion scaling techniques, previously validated in lab settings, can be successfully applied in a live remote telesurgery scenario using the dVRK. The authors highlight the need to consider factors often overlooked in early research, such as surgical staff, procedure specifics, and the unique challenges of working with live tissue.

Methods

The core of this paper's methodology lies in the translation of motion scaling techniques for remote telesurgery. The authors adapted their previously developed motion scaling solutions, which aim to mitigate the effects of communication latency in teleoperation. These solutions modify the telecommands sent from the remote administrator (surgeon) to the bedside manipulator (robot).

Specifically, they implemented a system where telecommands are sent over a network with an introduced artificial delay to simulate remote telesurgery conditions. The motion scaling algorithm dynamically adjusts the scaling factor based on the operator's movement, decreasing it for precise motions and increasing it for faster, less precise movements. This is achieved through a combination of constant scaling ($\gamma_c$) and velocity scaling ($\gamma_v$). The paper details the kinematic calculations and Jacobian used to derive the telecommands and their transmission via ROS (Robot Operating System).

For the live experiment, a porcine model was chosen for a thyroidectomy procedure, a common surgery with anatomical similarities to humans. The surgical team, comprising otolaryngologists, adapted the procedure to maximize data collection per animal. Two cadaveric surgeries were performed beforehand to refine the procedure and familiarize the team with the dVRK, which differs from the commercial da Vinci systems they typically use.

Datasets

Not discussed in the paper. The study relies on data collected during the live animal surgery, including kinematic and endoscopic camera data.

Key Insights and Ideas

  • Bridging the Bench-to-Bedside Gap: This work is a significant step in demonstrating that research-level robotic surgery platforms like the dVRK can be successfully translated to live animal procedures, a critical milestone for broader clinical adoption.
  • Motion Scaling in Live Environments: The paper provides empirical evidence that motion scaling, particularly velocity scaling, can improve control and efficiency in live telesurgery under delay, addressing a key challenge in remote operations.
  • dVRK's Potential for Live Surgery: The successful execution of a live surgery on the dVRK opens doors for other research groups to conduct similar in-vivo experiments, fostering a more robust translation of surgical robotics research.
  • Importance of Pre-Clinical Preparation: The authors emphasize the necessity of extensive pre-clinical work, including cadaveric studies and careful procedure adaptation, to ensure a smooth and successful live animal experiment.
  • Challenges of dVRK vs. Commercial Systems: The surgeons noted differences in the dVRK's joint limits, mobility, and gravity compensation compared to commercial da Vinci systems, highlighting the need for adaptation and practice.

Key Results

The study demonstrated that motion scaling, particularly when reducing constant scaling and incorporating velocity scaling, led to more controlled motions compared to normal constant scaling under delay.

  • Time Efficiency: On average, reducing constant scaling and using velocity scaling modified the time to complete surgical tasks under a 500ms round-trip delay by 175% and 88% respectively, compared to normal constant scaling. Velocity scaling resulted in the lowest times.
  • Distance Traveled: Reducing constant scaling and using velocity scaling modified the total distance traveled by the bedside manipulators by 156% and 61% respectively, compared to normal constant scaling. Velocity scaling also resulted in the shortest distances.
  • Qualitative Improvement: The bedside assistant noted a significant reduction in uncontrolled motions when using reduced constant and velocity scaling. The expert surgeon found operating under delay challenging but acknowledged the improved control with the tested scaling methods.

Limitations and Future Work

Limitations: * The study involved a limited number of experimental trials (two pigs), providing a proof of concept rather than extensive validation. * The authors acknowledge that the reduced constant scaling, while improving control, made motions frustratingly slow. * The dVRK's limitations (joint limits, mobility, gravity compensation) compared to commercial systems required adaptation from the surgical team.

Future Work: * Conduct a larger number of randomized trials to further validate motion scaling for remote telesurgery. * Develop a scoring metric for expert surgeons to quantitatively assess the performance of surgical procedures from a clinical perspective. * Explore more advanced motion scaling strategies and their impact on surgical outcomes.

Summary updated

Abstract

Innovations from surgical robotic research rarely translates to live surgery due to the significant difference between the lab and a live environment. Live environments require considerations that are often overlooked during early stages of research such as surgical staff, surgical procedure, and the challenges of working with live tissue. One such example is the da Vinci Research Kit (dVRK) which is used by over 40 robotics research groups and represents an open-sourced version of the da Vinci ® Surgical System. Despite dVRK being available for nearly a decade and the ideal candidate for translating research to practice on over 5,000 da Vinci ® Systems used in hospitals around the world, not one live surgery has been conducted with it. In this paper, we address the challenges, considerations, and solutions for translating surgical robotic research from bench-to-bedside. This is explained from the perspective of a remote telesurgery scenario where motion scaling solutions previously experimented in a lab setting are translated to a live pig surgery. This study presents results from the first ever use of a dVRK in a live animal and discusses how the surgical robotics community can approach translating their research to practice.
BibTeX
@inproceedings{richter2021from,
  author = {Florian Richter and Emily K. Funk and Won Seo Park and Ryan K. Orosco and Michael C. Yip},
  title = {{From Bench to Bedside: The First Live Robotic Surgery on the dVRK to Enable Remote Telesurgery with Motion Scaling}},
  booktitle = {ISMR},
  year = {2021},
  url = {https://openreview.net/forum?id=6tyQbw014N},
}
Figure