Extending bioelectric navigation for displacement and direction detection

Heiko Maier, Heribert Schunkert, Nassir NavabInt. J. Comput. Assist. Radiol. Surg 2023 · DOI

Introduction

Endovascular procedures, such as stent placement, currently rely on fluoroscopy—real-time X-ray imaging. While effective, this gold standard exposes patients and medical staff to ionizing radiation and requires contrast agents that can be nephrotoxic. Bioelectric navigation has emerged as a promising radiation-free alternative, utilizing local electric fields to track catheter positions. However, existing methods struggle with continuous tracking in uniform vessel segments and lack directional awareness. This study, published in the International Journal of Computer Assisted Radiology and Surgery, proposes a novel extension to bioelectric navigation to overcome these limitations.

The Challenges of Bioelectric Navigation

Current bioelectric systems primarily rely on "feature-matching," identifying anatomical landmarks like bifurcations or stenoses. This approach faces two major hurdles:

  1. The "Straightaway" Problem: Accuracy drops significantly in long, uniform vessel segments where distinct anatomical features are absent.
  2. Directionality: Standard models often assume unidirectional forward movement, failing to account for the complex "push-pull" maneuvers surgeons perform during navigation.

Proposed Methodology: Continuous Displacement Tracking

The researchers introduced a stationary electrode, typically integrated into the vascular access sheath at the entry point (e.g., the groin). By establishing a low-frequency AC current between the catheter tip and this stationary reference, a measurable electric field is generated along the vessel.

Catheter electrode configurations and stationary reference.

Two mathematical models were evaluated:

  • Approach I: A basic model assuming a static electric field, calculating displacement via voltage drops between sensing electrodes.
  • Approach II: An advanced model designed to compensate for "parallel conductance"—the electrical interference caused by surrounding conductive tissues (organs, fat) that warps the electric field.

Experimental Validation

The team validated their approach using a combination of computational simulations and physical experiments.

Validation environments: 3D models and physical phantom.

Datasets and Models

  • Simulated Vasculature: A 3D model of an abdominal aortic aneurysm (AAA) derived from patient CT scans was used for Finite Element Method (FEM) simulations.
  • Physical Phantom: A 3D-printed vascular tree provided a controlled environment to validate the physics of displacement sensing before introducing biological complexity.

Key Results

The study demonstrated high precision in ideal conditions and identified the impact of tissue interference:

  • Simulation: In non-conductive environments, the system achieved an absolute error of less than 0.89 mm over an 89 mm path.
  • Conductive Interference: Introducing conductive surrounding tissue increased errors significantly (60.27 mm for Approach I). However, the refined Approach II reduced this error to 33.96 mm, highlighting the importance of compensating for parallel conductance.
  • Physical Testing: Experiments with the 3D-printed phantom yielded a mean absolute error of 6.3 mm with high reproducibility (SD ≤ 1.1 mm).

Discussion and Future Outlook

While the current results are promising, clinical application requires further refinement. Sub-5mm accuracy is often necessary for complex procedures, and current models must be adapted to account for pulsatile blood flow and varying catheter positions within the vessel. Future work will focus on testing these models in more complex biological environments, such as animal studies, to better handle real-world electrical interference. Integrating this displacement tracking with existing map-matching algorithms could eventually provide a robust, radiation-free "GPS" for endovascular surgery.

Summary updated

BibTeX
@article{maier2023extending,
  author = {Heiko Maier and Heribert Schunkert and Nassir Navab},
  title = {{Extending bioelectric navigation for displacement and direction detection}},
  journal = {Int. J. Comput. Assist. Radiol. Surg.},
  year = {2023},
  doi = {10.1007/s11548-023-02927-w},
}
Figure