Abstract
Minimally invasive treatment of vascular disease demands dynamic navigation through complex blood vessel pathways and accurate placement of an interventional device, which has resulted in increased reliance on fluoroscopic guidance and commensurate radiation exposure to the patient and staff. Here we introduce a guidance system inspired by electric fish that incorporates measurements from a newly designed electrogenic sensory catheter with preoperative imaging to provide continuous feedback to guide vascular procedures without additional contrast injection, radiation, image registration, or external tracking. Electrodes near the catheter tip simultaneously create a weak electric field and measure the impedance, which changes with the internal geometry of the vessel as the catheter advances through the vasculature. The impedance time series is then mapped to a preoperative vessel model to determine the relative position of the catheter within the vessel tree. We present navigation in a synthetic vessel tree based on our mapping technique. Experiments in a porcine model demonstrated the sensor's ability to detect cross-sectional area variation in vivo. These initial results demonstrate the capability and potential of this novel bioimpedance-based navigation technology as a non-fluoroscopic technique to augment existing imaging methods.
Introduction: The Challenges of Modern Fluoroscopy
Modern minimally invasive vascular procedures—such as the placement of stents or the clearing of arterial blockages—rely heavily on fluoroscopy, or live X-ray imaging. While this gold standard is effective, it presents significant drawbacks for both patients and medical staff. Continuous exposure to ionizing radiation accumulates over time, and the use of nephrotoxic contrast dyes to visualize "invisible" vessels can be taxing on patients with compromised kidney function. Furthermore, surgeons must navigate complex, three-dimensional vascular trees using two-dimensional X-ray projections, often resulting in a repetitive "guess and check" workflow.
While electromagnetic tracking systems exist for specific cardiac procedures, they are often impractical for peripheral vasculature due to vessel deformation, patient movement, and the requirement for bulky external hardware. This research introduces a "self-contained" navigation method that allows a catheter to determine its own position within the body's internal "map."
Bioelectric Navigation: Learning from Electric Fish
The proposed Bioelectric Navigation (BSN) system is inspired by the active electrolocation used by weakly electric fish. These organisms navigate murky waters by generating a weak electric field and sensing how nearby objects distort it.
The Hardware and Physics
The system utilizes a sensory catheter equipped with ring electrodes near the tip. These electrodes emit a low-frequency AC signal and measure the resulting impedance. Because the impedance of blood is significantly lower than that of vessel walls, the measured impedance is inversely proportional to the vessel's cross-sectional area (CSA). As the catheter traverses narrowings (stenoses) or widenings (bifurcations), it generates a unique "geometric barcode" or signal fingerprint.
The Algorithm
To translate these live signals into a precise location, the researchers employed Open-ended Dynamic Time Warping (OE-DTW). This algorithm compares the live, incomplete impedance stream from the catheter against a pre-calculated dictionary of reference signals derived from a pre-operative CT Angiography (CTA) scan. OE-DTW is particularly robust as it accounts for variations in the speed at which the surgeon advances the catheter.
Experimental Validation and Results
The BSN system was validated through three distinct stages: computational modeling, synthetic anatomical phantoms, and an in vivo porcine model.
Key Findings
- High Classification Accuracy: In trials using a 3D-printed acrylic phantom modeled after human femoral and coronary arteries, the algorithm correctly identified the catheter's path in 21 out of 24 trials. In the three instances of failure, the correct path was identified as the second-best match, typically occurring when two paths remained geometrically identical until the very end of the trajectory.
- In Vivo Success: In a live porcine model, the system successfully detected the renal artery ostia (the openings to the kidney arteries) and accurately distinguished between the left and right renal arteries despite the presence of blood flow and physiological movement.
- Signal Correlation: The study confirmed a strong correlation between the inverse of the vessel's CSA (from CT scans) and the measured voltage magnitude, validating the underlying physical model for clinical application.
Discussion and Future Directions
The core insight of this work is that the internal geometry of the vascular tree is sufficiently unique to serve as a map. By using active sensing—carrying its own "field" rather than relying on external cameras—the catheter becomes less susceptible to errors caused by patient movement or bed shifting.
However, several steps remain before clinical adoption. The current signal matching was performed post-hoc; real-time implementation is necessary to provide a live "GPS" for surgeons. Additionally, the 6F (2mm) prototype must be miniaturized to fit standard surgical guidewires for reaching smaller vessels, such as those in the brain. Future work will also need to address vessel deformation caused by the catheter itself, which can introduce mismatches between live signals and static pre-operative maps. Despite these challenges, BSN represents a promising path toward reducing radiation and contrast reliance in vascular interventions.