Assessing Hemodynamic Impact of Tissue-Engineered Vascular Graft Displacement: Combining Postoperative in vivo Results and Computational Modeling to Improve Surgical Planning.

Seda Aslan, Enze Chen, Miya Mese-Jones, Jacqueline Contento, Hidenori Hayashi, Keigo Kawaji, Joey Huddle, Jed Johnson, Yue‐Hin Loke, Mark Fuge, Laura Olivieri, Thao D. Nguyen, Narutoshi Hibino, Axel Krieger

TL;DR

This study quantifies how surgical imprecision—specifically the displacement and sizing of tissue-engineered vascular grafts (TEVGs)—alters blood flow dynamics. By combining porcine in vivo data with high-fidelity computational simulations, the researchers developed a "tolerance map" that defines the allowable margins for graft placement to ensure long-term surgical success.

What Problem Does This Paper Address?

For children born with aortic anomalies like coarctation (narrowing of the aorta), traditional synthetic grafts are a poor fit because they cannot grow with the patient. Tissue-engineered vascular grafts (TEVGs) offer a solution: they act as biodegradable scaffolds that the body eventually replaces with native tissue.

However, there is a major "translation gap." While engineers can design a perfectly optimized graft in CAD software, the reality of the operating room is messy. Surgeons must work within tight anatomical spaces, and even a few millimeters of misalignment or a slight rotation during suturing can significantly alter hemodynamics (blood flow patterns). Until now, we didn't know exactly how much "surgical error" a graft could handle before the risk of complications like thrombosis (clotting) or stenosis (re-narrowing) became critical.

Methods

The researchers employed a two-part approach:

  1. In Vivo Validation: They implanted customized TEVGs into four porcine models. They used Magnetic Resonance Angiography (MRA) to compare the intended preoperative design with the actual postoperative geometry.
  2. Virtual Implantation Framework: To systematically study the impact of placement errors, they developed a simulation pipeline:
  3. Finite Element Modeling (FEM): Unlike simple geometric overlays, they used FEM to simulate "virtual anastomosis"—the physical act of suturing the graft to the native vessel. This accounted for how the soft tissues deform and pull on each other.
  4. Computational Fluid Dynamics (CFD): They ran blood flow simulations on these deformed geometries to measure two key metrics: Peak Systolic Pressure Drop (PSPD) and Time-Averaged Wall Shear Stress (TAWSS).
  5. Sensitivity Analysis: They incrementally varied the graft’s translation (up to 6.9 mm), rotation (up to 38°), and size to see which factors most affected blood flow.

Datasets

The study utilized a primary dataset of four porcine models (n=4), weighing 20-30 kg.

  • Imaging: Cardiac MRA was used for both preoperative planning and postoperative analysis.
  • Benchmarks: The study did not use external public datasets; instead, it established its own benchmark by comparing "as-designed" CAD models against "as-implanted" postoperative geometries.

Key Insights and Ideas

  • The "Virtual Anastomosis" Advantage: Simply placing a CAD model into a scan isn't accurate. By using FEM to simulate the mechanical stress of suturing, the researchers improved the accuracy of their aortic shape predictions by 27.5%.
  • Direction Matters: Hemodynamics are not equally sensitive to all movements. The system was most sensitive to inferior-superior (vertical) translation and anterior-posterior rotation. A small shift up or down is much more dangerous than a shift left or right.
  • The Danger of Undersizing: The study found that undersized grafts (smaller than the design) produced much larger deviations in pressure and shear stress than oversized grafts. This suggests that if a surgeon is between sizes, "erring on the side of larger" may be hemodynamically safer.
  • Tolerance Mapping: The authors produced a spatial "safe zone" map. For these models, the allowable translational deviation was roughly 1–5 mm depending on the direction, providing a concrete target for surgical precision.

Key Results

  • Real-world Displacement: In the animal study, grafts experienced up to 6.9 mm of displacement and 38° of rotation compared to the plan.
  • Hemodynamic Impact: These real-world deviations led to a 50% increase in pressure drop (2.5 mmHg) and a 72% increase in wall shear stress (7.2 Pa) over predicted values.
  • Clinical Thresholds: Despite the shifts, the virtual study showed that PSPD generally stayed below the 20 mmHg clinical threshold for intervention, meaning Wall Shear Stress (TAWSS) is actually the more restrictive and critical metric for surgeons to watch.

Limitations and Future Work

  • Small Cohort: With only four pigs, the statistical power is limited, though the trends were consistent.
  • Rigid Wall Assumption: The CFD simulations assumed the vessel walls were rigid to save computational time. Future work should use Fluid-Structure Interaction (FSI) to account for the pulsing of the aorta.
  • Short-term Scope: The study focused on immediate postoperative results. TEVGs are designed to remodel over months; long-term studies are needed to see if the body "corrects" for initial placement errors or if those errors worsen over time.
  • Clinical Integration: The next step is moving this "tolerance mapping" into a real-time intraoperative guidance tool for surgeons.

Summary updated

Abstract

Tissue-engineered vascular grafts (TEVG) have shown promise in advancing vascular reconstructions. However, precise in vivo implantation is challenging, and it is unclear how deviations in location and size affect hemodynamics. This study aims to compare preoperative designs and postoperative anatomies of TEVG in an in vivo study to evaluate discrepancies and investigate the impact of graft displacement and size on hemodynamics by virtually simulating implantation scenarios that are informed by in vivo postoperative results. Designed and postoperative geometries of four porcine aortas were compared to measure the mismatch in implantation location and graft shape. These results informed a virtual TEVG implantation study. TEVG location, orientation, and size were varied to investigate the effects on the final aorta shape and hemodynamics. Anastomosis of TEVG was simulated using finite element modeling. Key hemodynamic metrics were obtained from virtual implantations and actual postoperative anatomies using computational fluid dynamics. Our in vivo study showed that TEVGs can experience up to 6.9 mm displacement and a 38 degree rotational shift post-implantation, leading to discrepancies in pressure drop (2.5 mmHg, 50%) and time-averaged wall shear stress (7.2 Pa, 72%) compared to predictions. Virtual anastomosis simulations improved aortic shape predictions by 27.5%. Our results highlight the sensitivity of key hemodynamic metrics to graft implantation location and size mismatch. By quantifying displacement ranges and their impacts during surgery, surgeons can make informed decisions.