Abstract: One of the major reasons that totally implantable cochlear microphones are not readily available is the lack of good implantable microphones. An implantable microphone has the potential to provide a range of benefits over external microphones for cochlear implant users including the filtering ability of the outer ear, cosmetics, and usability in all situations. This paper presents results from experiments in human cadaveric ears of a piezofilm microphone concept under development as a possible component of a future implantable microphone system for use with cochlear implants. This microphone is referred to here as a drum microphone (DrumMic) that senses the robust and predictable motion of the umbo, the tip of the malleus. The performance was measured of five DrumMics inserted in four different human cadaveric temporal bones. Sensitivity, linearity, bandwidth, and equivalent input noise were measured during these experiments using a sound stimulus and measurement setup. The results demonstrate that the DrumMic behaves robustly across ears and fabrication, with performance approaching that of commercial hearing aid microphones.
Introduction
Cochlear implants (CIs) have revolutionized hearing for over a million people, yet they remain dependent on external hardware. External microphones are aesthetically conspicuous, susceptible to wind noise, and unusable in water or during sleep. Perhaps most importantly, they bypass the natural "pinna effect"—the way the outer ear and ear canal naturally filter and amplify sound to provide cues for localization and speech clarity.
The "holy grail" of CI research is a totally implantable system. However, the primary bottleneck has been the microphone. Subcutaneous sensors often suffer from poor signal-to-noise ratios due to skin attenuation. This study explores a middle-ear approach using a "DrumMic" designed to sense the mechanical vibrations of the ossicular chain directly.
Methodology and Design
The DrumMic concept is based on a piezoelectric polymer called polyvinylidene fluoride (PVDF). The device consists of a thin PVDF membrane stretched over a 3mm brass ring.
As sound vibrates the eardrum and the ossicular chain, the umbo (the tip of the malleus bone) moves, stretching the PVDF membrane. This mechanical strain generates an electrical charge processed by a custom low-noise charge amplifier.
The researchers tested five custom-fabricated DrumMics across four human cadaveric temporal bones. To ensure optimal mechanical contact, tiny plastic shims (0.65 mm thick) were used to tension the microphone between the cochlear promontory and the umbo. Laser Doppler Vibrometry (LDV) was used to measure the actual velocity of the umbo to benchmark the microphone's electrical output.
Key Results
Experimental testing demonstrated that the DrumMic provides a robust and predictable signal:
- Linearity: The device behaved linearly across a massive dynamic range, from 46 dB SPL to at least 100 dB SPL.
- Bandwidth: It maintained a strong signal-to-noise ratio up to 8 kHz, covering the essential frequencies for human speech.
- Sensitivity: The output was tightly clustered around 25 fC/Pa across different ears and devices, despite anatomical variations.
- Noise Floor: The measured Equivalent Input Noise (EIN) was 54 dB SPL. However, when accounting for the natural 10–20 dB pressure gain provided by the ear canal and pinna, the effective EIN was estimated at 46 dB SPL.
Discussion
A critical finding of this study is the "canceling effect" that makes the DrumMic robust to surgical placement. While a tighter fit (using more shims) increases the tension and sensitivity of the PVDF film, it also stiffens the ear's mechanics, which reduces the umbo's motion. These two factors effectively balance each other, resulting in consistent performance across different anatomical structures.
Furthermore, by placing the sensor behind the eardrum, the system leverages the natural anatomy of the ear. The pressure gain from the ear canal helps offset the intrinsic noise of the sensor, bringing its performance closer to the ~34 dB SPL floor of top-tier external hearing aid microphones.
Conclusion and Future Work
The DrumMic demonstrates linear, wide-bandwidth performance that is remarkably stable across different subjects. To transition this concept into a human-implantable prototype, future research must focus on hermetic encapsulation to protect the device from the moist environment of the middle ear, biocompatibility, and long-term mechanical stability. Additionally, further optimization of the charge amplifier and electrical shielding will be necessary to further reduce the noise floor.