Fewer Catheters to Certify a Non-Invasive Heart-Pressure Model

Machine learning models can now estimate pressures inside the heart and lungs from an ultrasound exam instead of a catheter threaded through the veins. Before such a model is trusted at a new hospital, someone has to show it is accurate there. That requires patients who have both the ultrasound and the invasive catheter measurement, so the check itself carries catheter risk. Hospitals differ in scanners and patient populations, so the check must be repeated at every site. Paired cohorts are small, which leaves confidence intervals too wide to prove anything.

We propose to cut the number of catheterized patients needed per site by using the many ultrasound exams at that site that have no catheter measurement. Multimodal models contain partly redundant readouts: a video-based estimate, a Doppler-based estimate, and a metadata-based estimate. How much these readouts disagree on a patient can be computed without a catheter, and it tracks how wrong the model is on that patient. We use this disagreement only as a variance-reducing helper, called a control variate, inside an otherwise standard non-inferiority test. The resulting estimate of site-level error stays unbiased however badly the site differs from the training data, and it becomes much tighter whenever the disagreement signal is informative. The practical payoff is a quantified reduction in how many invasive procedures a hospital must pair with ultrasound before it can accept or reject a locked model.

Core Claim

The claim is narrow. For a frozen multimodal hemodynamic estimator deployed at a shifted site, cross-modal disagreement measured on unlabeled patients predicts per-patient error well enough to reduce the variance of the site-level error estimate substantially. The two reference methods are the pooled development dataset (Development cohort, below) and the naive paired analysis (Baselines).

Let $d_i$ be the per-patient quantity the non-inferiority claim concerns. One choice is the model's absolute error against catheterization minus the expert Doppler reader's absolute error, both on patient $i$. Let $g_i$ be a catheter-free proxy, such as the spread between the modality-specific heads' pressure estimates, computed for all $N$ unlabeled exams and for the $n$ catheterized ones. The estimator of the mean $\theta = \mathbb{E}[d]$ is

$$\hat\theta = \bar d_n - \lambda\,(\bar g_n - \bar g_N),$$

where $\bar d_n$ and $\bar g_n$ are means over the catheterized patients, $\bar g_N$ is the mean over all exams, and $\lambda$ is estimated by cross-fitting. Its variance is approximately

$$\frac{\mathrm{Var}(d)}{n}\bigl(1-\rho^2(1-n/N)\bigr),$$

with $\rho$ the correlation between $d$ and $g$. This is the standard control-variate identity. When $N \gg n$, it acts like an effective sample size of $n/(1-\rho^2)$. A correlation of 0.7 would roughly halve the catheterizations needed for the same interval width.

The key property is that shift can lower $\rho$ but cannot bias $\hat\theta$. The proxy only subtracts a term whose expectation is zero. Validity therefore rests on the catheterized patients being a random sample of the intended-use population. This assumption is load-bearing, and the design must enforce it.

The scientific question underneath is which catheter-free signals keep their correlation with true error under scanner and population shift. Cross-modal disagreement is the candidate. A single model's own confidence is the default it has to beat.

Where this sits in prior work

Several papers motivate the problem. MePH fuses 3D echo video, Doppler spectrograms and metadata against right heart catheterization in 1,237 patients from 12 centers. It reports aggregate error reductions on a 237-patient external cohort but no power analysis or non-inferiority margin. Its multi-stream design is exactly the redundancy we exploit.

Ragnarsdóttir et al. show how badly site and time shift bite: neonatal PH severity F1 falls from 0.84 in cross-validation to 0.63 on held-out patients. Per-site evidence therefore cannot be skipped. The multi-view VAE learns shared and view-specific structure across five echo views. Its view-specific heads are a second natural source of disagreement signals.

Two papers constrain the statistical side. Ostmeier et al. run a formal non-inferiority test of a non-contrast CT model against neuroradiologists on only 32 test patients. That is the template we adopt, and its small n is the problem we attack. Kim et al. and Li et al. show how small cohorts yield non-significant p-values with intervals too wide to claim equivalence.

InputAdapter supplies the regulatory setting: predictors locked as medical devices, with only their inputs adapted. It is complementary rather than competing. Adapting inputs still requires labeled target data and still requires certification afterwards. Our method makes that certification cheaper.

InputAdapter also found that successful adaptation can increase distributional distance between source and target. That finding supports our stance of using unlabeled target data to measure rather than to align. CoReEcho shows that continuous, rank-ordered representations transfer across sites better than brittle regression features. That suggests heads trained this way yield more shift-stable disagreement signals, which we test as one proxy source.

Research Program

The first component is a proxy family. For each frozen model we compute several catheter-free signals:

  • the spread between modality-restricted heads: video-only, Doppler-only, and metadata-only;
  • the gap between the model's mPAP estimate and a classical Doppler-formula estimate;
  • the disagreement across echo views;
  • the model's own predictive variance.

We also train a small error regressor on these features using source-site data only, so the target estimate is not contaminated.

The second component is the test. A two-sided confidence interval for $\theta$ is built from $\hat\theta$ and compared with a pre-specified margin $\delta$, following Ostmeier et al. A group-sequential version lets a site stop pairing new patients once the interval clears or crosses $\delta$.

The third component handles retrospective settings, where catheterized patients were selected by clinical indication rather than at random. For these we add an inverse-propensity-weighted variant. We report explicitly how its guarantees weaken. The prospective protocol we recommend randomizes which referral-population patients receive research pairing.

Baselines

The naive reference is the standard paired non-inferiority analysis on catheterized patients alone. It uses a t-interval or bootstrap on $d_i$, as in Ostmeier et al., who established non-inferiority at $p<0.05$ with 32 patients. No reported result found on Lacuna for interval widths or catheter counts needed for non-inferiority of a hemodynamic estimator. The closest figure is MePH's 49.73% mPAP error reduction relative to experts on 237 external patients, reported without intervals. This baseline is the strongest competitor on validity because it is assumption-light. It receives the same revealed patients, and beating it means narrower intervals at matched coverage.

The second baseline is hierarchical pooling across sites. A random-effects model shrinks the new site's error toward the average error of other sites. It is the strongest competitor on interval width. It gets identical labeled data from all other sites. Beating it means matching its width while keeping nominal coverage on held-out sites that are genuinely different, where shrinkage should under-cover.

The third baseline is pseudo-labeling, which treats the Doppler-formula or expert estimate as truth on unlabeled patients. It represents the "use the unlabeled data directly" instinct.

Two ablations isolate the key ingredient. The first replaces cross-modal disagreement with the model's own confidence only. The second uses a proxy fitted on target labels without cross-fitting, to show the resulting invalidity. Every method receives identical frozen models, identical revealed-patient subsets over 1,000 random draws, and the same margin $\delta$.

Why not just…?

Why not just catheterize more patients per site? Every additional pairing is either an indicated procedure that must be scheduled and timed within 24 hours of the echo, or unethical to obtain at all. The data-desert point in the file applies directly: paired cohorts sit in the hundreds. Our experiments report how many pairings the proxy saves at fixed width. If $\rho$ is small the savings vanish, and we would report that plainly.

Why not just treat the Doppler or expert estimate as ground truth on unlabeled patients? Doppler-derived pressures carry large errors; MePH's expert baseline is substantially worse than the model. Under shift, those errors change in unknown ways. Pseudo-labels therefore give tight intervals centered in the wrong place. The coverage experiment below measures this directly, whereas our estimator uses the same signals only as a mean-zero correction.

Why not just pool evidence across sites with a hierarchical model? This is the real threat, because when sites are similar it will produce narrower intervals than ours. Its weakness is that sites are least similar exactly where certification matters, as the 0.84-to-0.63 drop in Ragnarsdóttir et al. illustrates. Leave-one-site-out coverage, stratified by scanner vendor, will show whether shrinkage under-covers on shifted sites. If it does not, a pooled model plus our control variate is the natural combination, and we will test that too.

Why not just adapt the model with InputAdapter and skip the statistics? Adaptation changes the model's error at the new site, and it still requires labeled target patients. Afterwards, someone must still show the adapted pipeline is non-inferior. The two methods compose, and we evaluate certification both before and after adaptation on the ICU benchmark.

Proposed Experiments

The primary testbed is the MePH cohort, subject to data access. If access is denied, we reproduce the pipeline on a comparable paired echo–catheterization cohort. Development centers define the frozen model and the source-trained proxy regressor. For each external site, we reveal catheter values for $n \in \{10, 20, 40, 80\}$ randomly chosen patients and hide the rest. Because every patient in this cohort has catheter data, the full-site $\theta$ is known exactly. This lets us measure interval coverage, width, and the $n$ needed to decide non-inferiority at a margin $\delta$ set before analysis. We also report $\rho$ per proxy per scanner vendor, which answers the underlying question of which signals survive shift.

The neonatal dataset of Ragnarsdóttir et al. and the multi-view VAE provides a second, harder check. There the error is misclassification of severity, the proxy is cross-view disagreement, and the 78-neonate test year is the shifted site.

Validity is verified at scale on MIMIC-IV to eICU with InputAdapter's frozen predictors. Labs-only and vitals-only heads supply the disagreement signal, and the large labeled target set lets us check coverage over many draws.

The decisive outcomes are two. Coverage must stay at nominal level on every shifted site, which is the go/no-go criterion. Cross-modal proxies must reach $\rho$ high enough to save at least a third of pairings at matched width, where self-confidence proxies do not. If cross-modal disagreement correlates no better with error than single-model confidence does, the multimodal argument fails. The contribution would then shrink to a generic variance-reduction recipe.