Counting Reprocessing Cycles: A Dose–Response Safety Study of Reused Ablation Catheters

Hospitals throw away most of the catheters used in heart-rhythm procedures after one patient. Regulated reprocessing means cleaning, re-sterilizing, testing and reusing these "single-use" devices. It is promoted as one of the main ways to cut the carbon cost of cardiology. A single atrial fibrillation ablation, a procedure that burns away tissue causing an irregular heartbeat, emits a mean of 76.9 kg CO2-equivalent, driven mainly by disposable catheters, sensors and patches (Carbon footprint of atrial fibrillation catheter ablation). Clinical acceptance stalls on one question: is a reused catheter as safe as a new one? Nobody in the literature we found has measured this on real patients.

We propose to answer a sharper version of that question. Catheters are multi-material objects: metal electrodes glued to polymer shafts, and braided layers laminated together. Engineers studying concrete have shown that the thin boundary layer between two bonded materials is where cracks start, and that these weak zones compound when material is recycled and survive decades of service. If the same holds for catheters, the risk of reuse should not be flat. It should grow with the number of times a device has been reprocessed. We will measure that growth by combining bench tests of the bonded joints in retired catheters with a multicenter retrospective patient cohort, meaning a study built from existing hospital records. The cohort compares procedures that used the same catheter model at different cycle counts. A Delphi expert panel will first define what a "device-caused failure" looks like in a medical record. The output is a cycle-specific safety curve that can be traded against carbon savings to set an evidence-based reuse limit.

Core Claim

The useful safety question is how the hazard of device-caused failure changes with each additional reprocessing cycle. Comparing "reprocessed versus new" is less useful, and it is also the more confounded comparison. Two reasons make the cycle-count design better than the obvious one.

The first reason is physical. The research on concrete transition zones gives a concrete mechanism to test. These are the porous, crack-prone boundary layers between aggregate and cement paste. Their weakness arises from poor packing and excess water at the interface (A Review on Multi-Scale Toughening and Regulating Methods for Modern Concrete). It doubles when recycled aggregate brings an old interface into a new composite (A Review: The Modification Methods of Concrete Recycled Coarse Aggregates). It persists after 50 years of service (Research on the material properties of concrete in subway stations after in service for 50 years). We use this literature as a source of hypotheses and measurement methods, not as evidence that catheters behave like concrete. Polymer–metal adhesive joints are a different material system. Whether repeated heat, chemical sterilization and handling accumulate damage at those joints is exactly what the bench arm will test.

The second reason is epidemiological. Whether a patient gets a reprocessed or a new catheter is a clinical choice. Operators may save new devices for harder cases, so a direct comparison suffers from confounding by indication. Confounding by indication means that the factors driving treatment choice also drive outcomes. Among procedures that receive a reprocessed device, the specific cycle count is largely set by which unit happens to come off the inventory shelf. That makes cycle count closer to as-good-as-random, conditional on center and date, than the reprocessed-versus-new decision. We will test this assumption with balance checks rather than assume it.

Research Program

Delphi definitions. The program starts with a Delphi panel. The Delphi method is an anonymous, multi-round survey in which experts rate items, see the group's medians and anonymous rationales, and revise their ratings until agreement stabilizes. Byrne et al. showed that this controlled feedback brings outliers to converge or to document their reasons (An international, Delphi consensus study to identify priorities for methodological research in behavioral trials in health research). Content validity indices are item-level agreement scores used to drop weak items, as Wake et al. did for trauma follow-up instruments (Follow-up after major traumatic injury: a survey of services in Australian and New Zealand public hospitals). We will apply these indices to two products. One is a taxonomy of device-attributable failure events: electrode detachment, loss of tip contact-force signal, shaft kinking or delamination, and unplanned intra-procedural catheter exchange. The other is the record signatures that identify each event in operative reports, device logs and coded complications.

The panel will deliberately mix electrophysiologists, biomedical engineers and reprocessing technicians. This is the step the French society's position paper did not take. That paper endorsed reprocessing as expert opinion, without a formal consensus process or outcome data (Position paper on sustainability in cardiac pacing and electrophysiology). The Delphi stage is necessary because hospital records do not code "the glued electrode let go." Without agreed definitions the cohort would have no measurable outcome.

Bench measurements. In parallel, we will collect retired catheters of one or two widely reprocessed models at known cycle counts, using the reprocessor's tracking labels. At each bonded joint we will measure three things. The first is dynamic stiffness, using laser-induced vibration, which treats the interface as an elastic spring; Ibaraki et al. used this method to quantify bond recovery in carbonated recycled aggregate (Applicability of Laser-Induced Vibration Method for Quality Modification and Evaluation of Carbonated Recycled Aggregate). The second is void and crack volume at the adhesive interface, measured by micro-CT. The third is destructive tip pull-off force. The implant-coating literature motivates using interface-specific pull-out tests rather than bulk strength: there, loosening was a boundary-layer failure, and push-out force rose 161% when the interface was engineered (Peptide-based inflammation-responsive implant coating sequentially regulates bone regeneration to enhance interfacial osseointegration). The bench arm yields a degradation curve per cycle. That curve sets the expected shape of the clinical dose–response, and the clinical curve can later confirm or contradict it.

Retrospective cohort. The third component is a multicenter retrospective cohort assembled from centers that already reprocess ablation catheters and log device serial numbers and cycle counts. The analysis follows the multicenter safety-cohort template of Ding et al. (Ding et al.). It also borrows the registry-linkage approach used with the burns registry in Belinda Gabbe's work, where a structured registry was linked to standardized outcome definitions. Exposure is cycle count, treated as an ordinal variable and as a spline, with new devices as cycle zero. The primary outcome is a Delphi-defined device-attributable failure. Secondary outcomes are 30-day complications (pericardial effusion, infection), procedure duration, and arrhythmia recurrence at one year.

Carbon trade-off. Finally, we will combine the hazard curve with per-cycle carbon savings into a curve of carbon saved per additional expected failure. Hospitals could use this curve to set a cycle cap.

Baselines

The naive reference point is the status quo: every device is used once. Against this reference, every hazard ratio is computed, every carbon saving is credited, and every cycle cap is judged. Its failure rate comes from the cycle-zero procedures in our own cohort. We found no reported number on Lacuna for failure rates of new versus reprocessed ablation catheters, or for any reprocessed-device cohort.

The strongest existing analytic method is the propensity-score-matched comparison of reprocessed versus new devices. Following Tewari et al. (Tewari et al.), each procedure gets a probability of receiving a reprocessed device given baseline covariates, and procedures are matched across groups. It is the strongest competitor because it is the analysis a cardiology journal would expect and the one the field's gap analysis calls for. We will run it on exactly the same cohort, covariate set and outcome definitions as our cycle-count model. Both analyses will use the same pre-registered specification, and neither will be tuned. No reported result was found on Lacuna for PSM applied to reprocessed medical devices, so the comparison is internal.

"Beating" this baseline does not mean a larger effect. It means two things. First, our estimate should pass the falsification checks described below, such as the negative-control outcome, where the PSM estimate fails or is uninformative. Second, our design should resolve the question PSM cannot answer: whether risk at cycles one to two differs from risk at cycles four and above.

Two ablations isolate the key ingredients. The first replaces the Delphi outcome definition with a plain diagnosis-code definition, which tests whether the consensus step changes event capture and effect estimates. The second collapses cycle count to a binary reprocessed indicator inside our model. If the binary model fits equally well, the dose–response framing adds nothing.

Why not just…?

Why not just run the propensity-matched reprocessed-versus-new comparison and stop? This is a real threat. If the cycle curve is flat, the PSM answer is nearly as informative and much simpler. Its weakness is that matching can only balance what is recorded, and the reason an operator reaches for a new catheter on a hard case is often undocumented. Real-world allocation in the breast cancer cohort of Chen et al. followed age and tumor grade rather than the formal criteria. That is the kind of practice-driven allocation that leaves room for unrecorded drivers. The within-reprocessed cycle comparison avoids that choice. We will report both analyses side by side. If they agree and the cycle curve is flat, we will say the simpler analysis sufficed.

Why not just rely on manufacturer and reprocessor bench validation? Bench validation tests whether a device meets specification after reprocessing. It does not test whether small interfacial losses translate into patient events under real handling and anatomy. The subway concrete study shows that interfacial microcracks can persist while bulk properties look acceptable (Research on the material properties of concrete in subway stations after in service for 50 years). That is a reason to doubt pass/fail bulk tests, though in a different material system. Our design tests directly whether the bench degradation curve predicts the clinical curve.

Why not just mine spontaneous adverse-event reporting systems? Databases such as the FDA reporting system used in Cardiac adverse events associated with dual immune checkpoint inhibitors have no denominator of procedures and no cycle counts. Failures of reprocessed devices may also be reported more or less often than failures of new ones. Such databases can raise signals but cannot estimate a per-cycle hazard.

Why not skip the Delphi panel and use diagnosis codes? This is plausible if codes capture device failures well. The first ablation tests exactly that. If code-based and consensus-based outcomes give the same estimates and the same event counts, the panel was unnecessary for analysis, though still useful for adoption.

Proposed Experiments and Decisive Checks

We will pre-register three checks.

The first is a balance check of the as-good-as-random assumption. Patient covariates (age, ablation type, redo status, operator) should not predict cycle count within center and month. If they do, we fall back to adjusting for them and weaken our claim accordingly.

The second is a negative-control outcome: complications at the femoral access site caused by sheaths, which are never reprocessed in the cohort. These should show no gradient with ablation-catheter cycle count. A gradient would reveal residual confounding.

The third is a mechanism-consistency test. The bench per-cycle decline in joint stiffness and pull-off force should share its shape with the clinical hazard curve, including any threshold cycle. A clinical gradient without bench degradation would point to sterility or handling rather than interfaces, and we would report it that way.

Power is the honest limitation. Device failures are rare, so the cohort must pool several centers over many years, in the spirit of the decade-long multicenter window in 10 years trends and hospitalization outcomes of non-neonatal tetanus. We will state up front the smallest per-cycle hazard ratio detectable at our sample size, and we will interpret a null result as "no gradient larger than X," not as proof of safety.

The decisive outcome comes in three forms. A flat curve with intact bench interfaces supports raising cycle caps. A rising curve matched by interfacial degradation supports a mechanism-based cap. A rising clinical curve with intact interfaces redirects attention to sterilization.

References

  1. Carbon footprint of atrial fibrillation catheter ablation
  2. A Review on Multi-Scale Toughening and Regulating Methods for Modern Concrete: From Toughening Theory to Practical Engineering Application
  3. A Review: The Modification Methods of Concrete Recycled Coarse Aggregates
  4. Research on the material properties of concrete in subway stations after in service for 50 years
  5. An international, Delphi consensus study to identify priorities for methodological research in behavioral trials in health research
  6. Follow-up after major traumatic injury: a survey of services in Australian and New Zealand public hospitals
  7. Position paper on sustainability in cardiac pacing and electrophysiology from the Working Group of Cardiac Pacing and Electrophysiology of the French Society of Cardiology
  8. Applicability of Laser-Induced Vibration Method for Quality Modification and Evaluation of Carbonated Recycled Aggregate
  9. Peptide-based inflammation-responsive implant coating sequentially regulates bone regeneration to enhance interfacial osseointegration
  10. Occurrence rates and risk factors of in-hospital venous thromboembolism, major bleeding, and death in patients receiving fondaparinux after orthopedic surgery or trauma surgery
  11. Long-Term Survival in Men With High Grade Prostate Cancer: A Comparison Between Conservative Treatment, Radiation Therapy and Radical Prostatectomy—A Propensity Scoring Approach
  12. Practice patterns and outcomes for patients with node-negative hormone receptor-positive breast cancer and intermediate 21-gene Recurrence Scores
  13. Cardiac adverse events associated with dual immune checkpoint inhibitors: a pharmacovigilance analysis from the FDA adverse event reporting system
  14. 10 years trends and hospitalization outcomes of non-neonatal tetanus: a large-scale multicenter retrospective study in China