Abstract
Measuring vertical profiles of the particle light-absorption coefficient using absorption photometers is often complicated by rapid changes in relative humidity (RH). These instruments determine absorption by measuring light attenuation through a particle-loaded filter. However, filter materials absorb or release water as RH fluctuates, altering their optical properties and creating measurement artifacts. This study investigates these effects using the Single Channel Tricolor Absorption Photometer (STAP) and the microAeth® MA200. Laboratory experiments revealed that the STAP (quartz-fiber filter) and MA200 (PTFE filter) exhibit opposing responses to RH changes. While linear and exponential correction schemes are proposed, the study highlights significant uncertainties due to RH sensor response times and ultimately recommends the use of aerosol dryers for vertical profiling.
Introduction
To accurately model climate change and assess air quality, scientists must quantify how much light is absorbed by atmospheric aerosols, such as black carbon (soot). Filter-based absorption photometers are frequently deployed on mobile platforms like drones, balloons, and aircraft to measure these properties across vertical atmospheric profiles. However, these measurements are prone to "ghost" signals—apparent absorption spikes that occur when the instrument moves through layers of varying humidity, such as when entering or exiting a cloud or crossing a boundary layer.
The Challenge of Humidity
The core of the problem lies in the hygroscopic nature of filter materials. When RH changes rapidly ($dRH/dt$), the filter material (quartz-fiber in the STAP or PTFE in the MA200) interacts with water vapor. This interaction changes the light attenuation through the filter, which the instrument's software incorrectly interprets as a change in particle loading. In clean environments or during rapid vertical ascents, these artifacts can be several times larger than the actual aerosol signal, leading to significant data bias.
Experimental Methodology
To quantify these artifacts, researchers designed a controlled laboratory setup to simulate rapid RH transitions. The experiments tested clean filters, filters loaded with hydrophobic black carbon (BC), and filters loaded with hydrophilic ammonium sulfate.
The setup utilized a 0.5 m³ mixing chamber to prepare aerosol samples, which were then monitored by a suite of instruments including a Multi-Angle Absorption Photometer (MAAP) and a Mobility Particle Size Spectrometer (MPSS) to ensure consistent particle concentrations. The RH was controlled using a humidifier and mass flow controllers (MFC), allowing for precise, stepped changes in humidity delivered to the STAP and MA200 units.
Key Findings: Opposing Responses and Memory Effects
The study revealed a striking difference in how the two instruments react to humidity. The STAP showed a positive linear correlation with RH changes: as humidity increased, the apparent absorption increased. Conversely, the MA200 showed a negative response, where an increase in RH led to a decrease in apparent absorption.
At a wavelength of 624 nm (60 s average), the STAP's sensitivity ranged from 0.14 to 0.29 Mm⁻¹ per % change in RH, depending on the filter loading material. The MA200 was more stable across different materials but showed a consistent negative response of approximately -0.4 Mm⁻¹ per % change in RH.
Furthermore, the MA200 exhibited a "memory effect" characterized by an exponential recovery period after an RH change. Researchers speculate this is due to water forming a thin film on the hydrophobic PTFE fibers, which acts as a refractive index matching layer, temporarily increasing light transmission through the filter.
Proposed Corrections and Limitations
The researchers developed mathematical correction schemes to mitigate these artifacts:
- Linear Correction (STAP): Correlating 1 Hz resolved absorption coefficients with RH change rates. The linear response was estimated at 10.08 Mm⁻¹ s⁻¹ %⁻¹.
- Exponential Recovery Model (MA200): A more complex approach to account for the time-lagged response of the PTFE filter.
However, the authors caution against relying solely on these fixes. The primary bottleneck is the response time of standard RH sensors ($t_{63} < 10s$), which is often slower than the optical response of the filter. If the sensor cannot capture the fastest RH fluctuations, the mathematical correction will remain incomplete.
Recommendations for Future Research
Because of the inherent uncertainties in post-processing corrections, the study concludes with two practical recommendations for the atmospheric science community:
- Hardware Solutions: The most effective way to eliminate RH artifacts is to use an aerosol dryer upstream of the absorption photometer. This ensures the sample air reaches the filter at a constant, low humidity.
- Operational Adjustments: When dryers cannot be used, vertical measurement platforms should ascend or descend slowly through layers with high RH gradients to minimize the $dRH/dt$ term.
While the proposed mathematical corrections point in the right direction and help estimate measurement uncertainties, they should be considered a secondary tool compared to physical humidity control.
Summary updated