Abstract
The indium activation diagnostic was calibrated on an accelerator neutron source to diagnose deuterium-deuterium (DD) neutron yields for implosion experiments at the Shenguang-III facility. To improve accuracy, the scattered neutron background was measured using a polypropylene shield, with the optimal shield geometry determined via Monte Carlo simulations. The calibration achieved a linear range up to 100 cps net count rate and identified a room-scattered background of approximately 9%. The resulting calibration factor for the indium samples is $4.52 \times 10^{-7}$ counts/n with a low uncertainty of 4.3%.
Introduction
In Inertial Confinement Fusion (ICF) research, accurately measuring the neutron yield is critical for calculating the Experimental Ignition Threshold Factor (ITFx), a key metric for fusion progress. For high-yield experiments exceeding $10^{10}$ neutrons, indium activation is a preferred diagnostic due to its robustness against detector saturation. Neutrons activate $^{115}\text{In}$ into the metastable state $^{115m}\text{In}$, which subsequently decays by emitting 336 keV gamma rays. However, precise calibration of the activation-to-yield ratio is often hampered by room-scattered neutrons and isotopic interference. This study describes a rigorous calibration process designed to minimize these systematic errors for the Shenguang-III facility.
Methodology
The calibration utilized a titanium-deuterium (TiD) target on an accelerator to produce 2.45 MeV neutrons, simulating the DD fusion environment. A primary challenge in such environments is "room scatter," where neutrons reflecting off surfaces inflate the measured activation.
Background Subtraction and Shielding
To isolate the direct neutron signal, researchers employed a 25-cm long conical polypropylene shield placed between the source and the indium sample. This shield effectively blocked direct neutrons, allowing for a direct measurement of the background noise contributed by scattered neutrons. This physical subtraction method was essential for refining the final calibration factor.
Simulation and Optimization
Monte Carlo simulations using the Geant4 toolkit were performed to optimize the diagnostic geometry. Specifically, the team modeled the trade-off between increased neutron capture in thicker samples and the subsequent internal absorption of the emitted gamma rays.
The simulations confirmed that a 1-cm thickness provides the best balance for DD neutron diagnostics. Furthermore, the team verified the isotopic purity of the signal by tracking the sample's decay curve over 48 hours, confirming a half-life consistent with the 4.486-hour decay of $^{115m}\text{In}$, proving that interference from other isotopes was negligible.
Results and Discussion
The calibration established a linear response range for the diagnostic up to a net count rate of 100 counts per second (cps). Beyond this threshold, detector "dead time" begins to introduce non-linearities. The room-scattered background was found to contribute approximately 9% to the total activation signal during calibration.
By accounting for these factors, the team derived a calibrated factor ($F$) of $4.52 \times 10^{-7}$ counts per neutron. The total uncertainty was reduced to 4.3%, a significant improvement over previous benchmarks which typically ranged from 7.5% to 8%.
Conclusion
The successful calibration of the indium activation diagnostic provides a high-precision tool for measuring DD neutron yields at the Shenguang-III facility. By combining physical shielding with advanced Monte Carlo modeling, the researchers achieved a 4.3% uncertainty level. This diagnostic is capable of measuring yields up to $5 \times 10^{12}$ neutrons and can be scaled for even higher-yield experiments by adjusting sample geometry and placement to manage detector saturation.