The calibration of DD neutron indium activation diagnostic for Shenguang-III facility

Zifeng Song, Jiabin Chen, Zhongjie Liu, Xiayu Zhan, Qi TangarXiv 2014 · DOI · arXiv

Abstract

The indium activation diagnostic was calibrated on an accelerator neutron source to diagnose deuterium-deuterium (DD) neutron yields for implosion experiments at the Shenguang-III facility. To improve accuracy, the scattered neutron background was measured using a polypropylene shield, with the optimal shield geometry determined via Monte Carlo simulations. The calibration achieved a linear range up to 100 cps net count rate and identified a room-scattered background of approximately 9%. The resulting calibration factor for the indium samples is $4.52 \times 10^{-7}$ counts/n with a low uncertainty of 4.3%.

Introduction

In Inertial Confinement Fusion (ICF) research, accurately measuring the neutron yield is critical for calculating the Experimental Ignition Threshold Factor (ITFx), a key metric for fusion progress. For high-yield experiments exceeding $10^{10}$ neutrons, indium activation is a preferred diagnostic due to its robustness against detector saturation. Neutrons activate $^{115}\text{In}$ into the metastable state $^{115m}\text{In}$, which subsequently decays by emitting 336 keV gamma rays. However, precise calibration of the activation-to-yield ratio is often hampered by room-scattered neutrons and isotopic interference. This study describes a rigorous calibration process designed to minimize these systematic errors for the Shenguang-III facility.

Methodology

The calibration utilized a titanium-deuterium (TiD) target on an accelerator to produce 2.45 MeV neutrons, simulating the DD fusion environment. A primary challenge in such environments is "room scatter," where neutrons reflecting off surfaces inflate the measured activation.

Background Subtraction and Shielding

To isolate the direct neutron signal, researchers employed a 25-cm long conical polypropylene shield placed between the source and the indium sample. This shield effectively blocked direct neutrons, allowing for a direct measurement of the background noise contributed by scattered neutrons. This physical subtraction method was essential for refining the final calibration factor.

Figure 5: (a) Schematic of the conical polypropylene shield used for background measurement. (b) Simulated neutron energy distribution at the sample position and the $^{115}\text{In}(n,n')^{115m}\text{In}$ reaction cross-section.

Simulation and Optimization

Monte Carlo simulations using the Geant4 toolkit were performed to optimize the diagnostic geometry. Specifically, the team modeled the trade-off between increased neutron capture in thicker samples and the subsequent internal absorption of the emitted gamma rays.

Figure 2: Normalized gamma counts as a function of indium sample thickness. The simulation identifies 1 cm as the optimal thickness for maximizing detectable signal by balancing activation against self-absorption.

The simulations confirmed that a 1-cm thickness provides the best balance for DD neutron diagnostics. Furthermore, the team verified the isotopic purity of the signal by tracking the sample's decay curve over 48 hours, confirming a half-life consistent with the 4.486-hour decay of $^{115m}\text{In}$, proving that interference from other isotopes was negligible.

Results and Discussion

The calibration established a linear response range for the diagnostic up to a net count rate of 100 counts per second (cps). Beyond this threshold, detector "dead time" begins to introduce non-linearities. The room-scattered background was found to contribute approximately 9% to the total activation signal during calibration.

By accounting for these factors, the team derived a calibrated factor ($F$) of $4.52 \times 10^{-7}$ counts per neutron. The total uncertainty was reduced to 4.3%, a significant improvement over previous benchmarks which typically ranged from 7.5% to 8%.

Conclusion

The successful calibration of the indium activation diagnostic provides a high-precision tool for measuring DD neutron yields at the Shenguang-III facility. By combining physical shielding with advanced Monte Carlo modeling, the researchers achieved a 4.3% uncertainty level. This diagnostic is capable of measuring yields up to $5 \times 10^{12}$ neutrons and can be scaled for even higher-yield experiments by adjusting sample geometry and placement to manage detector saturation.

Summary updated

Abstract

The indium activation diagnostic was calibrated on an accelerator neutron source in order to diagnose deuterium-deuterium (DD) neutron yields of implosion experiments on Shenguang-III facility. The scattered neutron background of the accelerator room was measured by placing a polypropylene shield in front of indium sample, in order to correct the calibrated factor of this activation diagnostic. The proper size of this shield was given by Monte Carlo simulation software. The affect from some other activated nuclei on the calibration was verified by judging whether the measured curve obeys exponential decay and contrasting the half life of the activated sample. The calibration results showed that the linear range reached up to 100 cps net count rate in the full energy peak of interest, the scattered neutron background of accelerator room was about 9% of the total neutrons and the possible interferences mixed scarcely in the sample. Subtracting the portion induced by neutron background, the calibrated factor of our sample was 4.52*10^-7 counts/n with uncertainty of 4.3%.
BibTeX
@misc{song2014calibration,
  author = {Zifeng Song and Jiabin Chen and Zhongjie Liu and Xiayu Zhan and Qi Tang},
  title = {{The calibration of DD neutron indium activation diagnostic for Shenguang-III facility}},
  year = {2014},
  eprint = {1405.2520},
  archivePrefix = {arXiv},
  url = {https://arxiv.org/abs/1405.2520},
}