Statistical modelling and Bayesian inversion for a Compton imaging system: application to radioactive source localization

Cécilia Tarpau, Ming Fang, Konstantinos C. Zygalakis, Marcelo Pereyra, Angela Di Fulvio, Yoann Altmann

TL;DR

This paper introduces a novel statistical forward model for a new type of Compton imager that can simultaneously scatter and absorb photons, enabling a wider field of view. They apply this model within a Bayesian framework to accurately estimate the location and energy of radioactive sources, outperforming traditional methods, especially in low-photon scenarios, and providing crucial uncertainty quantification.

What Problem Does This Paper Address?

Traditional Compton cameras rely on two separate sensor layers: one for scattering and one for absorption. This paper addresses the challenge of imaging with a new "Compton imager" (CI) system that uses a single type of sensor capable of both scattering and absorbing photons. This design offers a wider field of view, making it suitable for applications requiring comprehensive coverage. The core problem is to accurately determine the location and energy of radioactive sources using data from this novel CI, especially when dealing with limited photon counts and inherent measurement noise. Existing methods like back-projection struggle in these low-photon, noisy conditions, and lack robust uncertainty estimation.

Methods

The authors developed a statistical forward model for the Compton imager, which accounts for the unique scattering and absorption capabilities of its sensors. This model is then used to tackle the inverse problem of source localization and energy estimation. They employ a Bayesian framework, which is well-suited for incorporating prior knowledge and quantifying uncertainty.

Specifically:

  • Location Estimation: A Metropolis within Gibbs (MWG) algorithm is used for estimating the source positions. This is a Markov Chain Monte Carlo (MCMC) method that iteratively samples from conditional distributions to explore the posterior distribution of the source locations.
  • Energy Estimation: An Expectation-Maximization (EM) algorithm is used for estimating the energy of the sources and identifying the nature of the photon interactions (Compton scattering vs. absorption). This algorithm iteratively refines estimates of unknown parameters by alternating between estimating the expected values of latent variables (E-step) and maximizing the expected log-likelihood (M-step).
  • Noise and Outlier Handling: The model incorporates latent variables to account for finite spatial and energy resolutions of the sensors, as well as potential outliers (erroneous event pairings). This makes the approach more robust to real-world measurement imperfections.

Datasets

The paper uses simulated data generated using the Monte Carlo N-Particle (MCNP) code [45-48]. These simulations are designed to mimic realistic measurements, including energy depositions and sensor noise. The authors do not introduce a new benchmark dataset but rather use simulated data to evaluate their proposed algorithms.

Key Insights and Ideas

  • Unified Sensor Capability: The CI's ability to both scatter and absorb photons in a single sensor type is a key innovation, enabling wider field-of-view imaging.
  • Bayesian Framework for Robustness: Applying a Bayesian approach with MWG and EM algorithms provides a more accurate and robust estimation of source location and energy, especially in challenging low-photon and noisy environments.
  • Uncertainty Quantification is Crucial: The Bayesian framework naturally provides uncertainty estimates for the source localization, which is vital for practical applications where knowing the confidence in a detection is as important as the detection itself.
  • Handling Real-World Imperfections: The inclusion of latent variables to model sensor noise and outliers significantly improves the algorithm's performance in realistic scenarios.
  • Superiority over Back-Projection: The proposed methods consistently outperform traditional back-projection techniques, particularly in terms of accuracy and consistency, especially when sources are close together.

Key Results

  • Improved Accuracy: The Bayesian approach (Metropolis within Gibbs) demonstrates significantly more accurate source localization compared to standard back-projection (BP) methods, especially in challenging scenarios like localizing multiple or closely spaced sources.
  • Consistent Performance: The Gibbs sampler results are more consistent across simulations, with tighter distributions of estimated source positions and more reliable uncertainty estimates compared to BP.
  • Effective Energy Estimation: The EM algorithm accurately estimates source energies and correctly classifies interaction types (Compton scattering vs. absorption), even with limited data.
  • Robustness to Noise and Outliers: The developed model and algorithms show good performance even with simulated sensor noise and the presence of spurious events.
  • Credible Level Alignment: The observed credible levels from the Gibbs sampler align well with theoretical expectations, indicating good uncertainty quantification.

Limitations and Future Work

  • Computational Cost: The Bayesian approach, particularly the Metropolis within Gibbs sampler, can be computationally expensive. While optimizations like look-up-tables were employed, further improvements in distance calculation methods are needed.
  • Convergence Challenges: In some experiments, especially with two closely spaced sources, the Gibbs sampler chains did not always converge to the true positions. The authors suggest that more iterations might be necessary, but this is limited by computational constraints.
  • Mono-energetic Source Assumption: The current model assumes sources share the same energy. A natural extension would be to handle sources with various energies.
  • Future Work: The authors suggest investigating less computationally expensive algorithms, extending the model to handle sources with different energies, and further exploring the use of additional interactions beyond the first two for improved localization.

Summary updated

BibTeX