Single breathhold, three-dimensional measurement of left atrial volume and function using sparse CINE CMR imaging with iterative reconstruction

Pierre Monney, Orestis Vardoulis, Davide Piccini, Amit H. Bermano, Amir Vaxman, Craig Gotsman, Janine Schwitter, Michael O. Zenge, Michaela Schmidt, Mariappan S. Nadar, Matthias Stuber, Nikolaos Stergiopulos, Juerg SchwitterJournal of Cardiovascular Magnetic Resonance 2015 · DOI

Introduction

The left atrium (LA) plays a critical role in cardiac hemodynamics, acting as a reservoir, a conduit, and a pump. Its size and phasic function—encompassing expansion and contraction—are vital prognostic indicators for various cardiovascular diseases. Specifically, LA dilation is often a precursor to poor clinical outcomes. However, accurately measuring LA volume in 3D remains a challenge. Conventional MRI techniques typically require multiple breathholds or long acquisition times, which are burdensome for patients. Alternatively, 2D geometric approximations often fail to capture the complex, asymmetrical shape of the atrium. This study introduces a novel approach to achieve time-efficient and geometrically accurate 3D LA assessment within a single breathhold.

Methodology

The researchers developed a prototype sequence termed sCINE-IR (sparse CINE with Iterative Reconstruction). This method utilizes sparse sampling to acquire only five "cine" slices—two long-axis (LAX) and three short-axis (SAX) views—during a single breathhold. This sparse acquisition achieves a spatial resolution of 1.5 mm and a temporal resolution of 30 ms.

To reconstruct a 3D volume from these sparse 2D slices, a non-model-based iterative reconstruction algorithm was employed. Unlike traditional model-based methods that assume a generic ellipsoidal shape, this approach utilizes the actual contours from the acquired slices to interpolate a patient-specific 3D geometry. This allows for the generation of a precise time-volume curve throughout the cardiac cycle.

Workflow of the sCINE-IR method: from 2D slices to 3D volume reconstruction and time-volume curve generation.

The method was validated through two primary studies:

  1. Phantom Study: Five LA phantoms were carved from potatoes (solanum tuberosum L) to mimic various human atrial shapes. Their true volumes were determined via water displacement.
  2. Patient Study: A pilot study involved three patients with a wide range of LA volumes (62 ml to 395 ml). Results were compared against a "gold standard" self-navigated, high-resolution whole-heart 3D dataset (3D-HR).

Results and Discussion

The sCINE-IR method demonstrated high accuracy and clinical potential. In the phantom study, the calculated volumes showed a strong correlation with the true volumes ($R^2 = 0.94$), despite a slight consistent underestimation of approximately 8.7%. Crucially, the measurements were found to be independent of slice orientation, suggesting robustness in clinical settings where perfect alignment may be difficult to achieve.

In the patient pilot, the correlation with the high-resolution 3D reference was exceptional ($R^2 = 0.99$), with a mean difference of only 2.7 ml. The ability to generate a continuous time-volume curve allowed for the clear identification of the three phases of LA function:

  • Reservoir Phase: Filling during ventricular systole.
  • Passive Emptying: Early diastolic flow into the ventricle.
  • Active Emptying: The "atrial kick" during late diastole.

The primary advantage of this technique is its speed; the entire 3D functional assessment is completed in one breathhold, significantly reducing patient discomfort and total scan time compared to traditional 3D MRI sequences.

Conclusion

This study demonstrates that sparse sampling combined with iterative reconstruction can provide accurate 3D measurements of LA volume and function in a single breathhold. While the results are promising, the study is limited by its small patient sample size. Future research will focus on validating the method in larger, more diverse patient cohorts and automating the contouring process to streamline clinical workflows.

Summary updated

Abstract

Methods A highly accelerated prototype cine sequence with sparse sampling and Iterative Reconstruction (sCINE-IR) was used in phantoms and patients to acquire 5 cine slices (2 long axis, LAX and 3 short axis, SAX) through the LA during a single breathhold yielding a spatial/temporal resolution of 1.5mm/30ms (1.5T Aera, Siemens AG, Germany). The LA volumes were reconstructed from these 5 slices using a non-model based method (Bermano A, ACM trans Graph 2011). As a reference in patients, a self-navigated high-resolution whole-heart 3D dataset (3D-HR) was acquired during mid-diastole, from which the LA volume was segmented. Phantom study. Five LA phantoms made of solanum tuberosum L of known volume (water displacement method) and of different shapes were imaged with both 3D-HR and CS in various slice orientations and the calculated volumes were compared. Patients study. Three patients were scanned with both 3D-HR and sCINE-IR. The volumes obtained with 3D-HR and with sCINE-IR during the corresponding mid-diastolic frame were compared using Bland-Altman method and linear regression.
BibTeX
@article{monney2015single,
  author = {Pierre Monney and Orestis Vardoulis and Davide Piccini and Amit H. Bermano and Amir Vaxman and Craig Gotsman and Janine Schwitter and Michael O. Zenge and Michaela Schmidt and Mariappan S. Nadar and Matthias Stuber and Nikolaos Stergiopulos and Juerg Schwitter},
  title = {{Single breathhold, three-dimensional measurement of left atrial volume and function using sparse CINE CMR imaging with iterative reconstruction}},
  journal = {Journal of Cardiovascular Magnetic Resonance},
  year = {2015},
  doi = {10.1186/1532-429x-17-s1-q35},
}
Figure