Self-Rotation of Cells in an Irrotational AC E-Field in an Opto-Electrokinetics Chip

Long-Ho Chau, Wenfeng Liang, Florence W. K. Cheung, Wing Keung Liu, Wen J. Li, Shih‐Chi Chen, Gwo‐Bin LeePLoS 2013 · DOI

Abstract

The use of optical dielectrophoresis (ODEP) to manipulate microparticles and biological cells has become increasingly popular due to its tremendous flexibility in providing reconfigurable electrode patterns and flow channels. ODEP enables the parallel and free manipulation of small particles on a photoconductive surface on which light is projected, thus eliminating the need for complex electrode design and fabrication processes. In this paper, we demonstrate that mouse cells comprising melan-a cells, RAW 267.4 macrophage cells, peripheral white blood cells and lymphocytes, can be manipulated in an opto-electrokinetics (OEK) device with appropriate DEP parameters. Our OEK device generates a non-rotating electric field and exerts a localized DEP force on optical electrodes. Hitherto, we are the first group to report that among all the cells investigated, melan-a cells, lymphocytes and white blood cells were found to undergo self-rotation in the device in the presence of a DEP force. The rotational speed of the cells depended on the voltage and frequency applied and the cells' distance from the optical center. We discuss a possible mechanism for explaining this new observation of induced self-rotation based on the physical properties of cells. We believe that this rotation phenomenon can be used to identify cell type and to elucidate the dielectric and physical properties of cells.

Introduction

Single-cell manipulation is a cornerstone of modern biomedical research, essential for everything from drug development to cancer diagnostics. Traditional methods like microfluidics often rely on fixed, physical channels that lack flexibility. Dielectrophoresis (DEP)—using electric fields to move particles—is more versatile but usually requires complex, pre-fabricated metal electrodes.

While Optical Dielectrophoresis (ODEP) solved the flexibility issue by using light to create "virtual electrodes," it was primarily used to move cells from point A to point B (translation). To make a cell rotate (electrorotation), scientists typically had to engineer complex, rotating electric fields. This research addresses the challenge of inducing and measuring cell rotation in a much simpler, "irrotational" AC field, providing a window into the cell's internal dielectric and structural makeup without the need for specialized electrode geometries.

Experimental Methodology

The team utilized an Opto-Electrokinetics (OEK) chip, which consists of a liquid medium sandwiched between a top layer of conductive glass (ITO) and a bottom layer of photoconductive amorphous silicon (a-Si:H).

Schematic of the OEK device structure showing the light-patterned virtual electrodes.
  1. Virtual Electrodes: Instead of metal wires, a commercial digital projector was used to shine light patterns onto the chip. Where the light hits, the silicon becomes conductive, creating a localized electric field.
  2. Cell Samples: Four types of mouse cells were tested: melan-a (pigment cells), RAW 264.7 macrophages, peripheral white blood cells (WBCs), and lymphocytes. Control experiments were conducted using 10 $\mu$m polystyrene beads.
  3. Signal Control: An AC signal (10 kHz to 3 MHz) was applied across the chip.
  4. Rotation Tracking: To quantify the spinning, the researchers developed custom software that used a "similarity index" algorithm. By comparing video frames of a rotating cell against a reference frame, they could precisely calculate rotational speed (RPM).

The Mechanism of Induced Self-Rotation

The most striking finding is that cells rotate in a non-rotating field. This suggests that the torque isn't coming from the field's geometry, but from the interaction between the field and the cell's internal structure. The authors hypothesize that rotation occurs because of an uneven distribution of mass within the cell (e.g., a large, off-center nucleus in a lymphocyte or heavy melanin granules in a melan-a cell). When the center of the induced dipole doesn't align with the cell's center of gravity, the DEP force generates a torque.

Microscopic images showing the manipulation and rotation of white blood cells compared to inert polystyrene beads.

Not all cells behave the same. Macrophages did not rotate at any frequency, while melan-a cells and lymphocytes spun vigorously. This "spinning signature" could act as a mechanical biomarker. Furthermore, rotational speed is highly dependent on frequency. For example, lymphocytes reached their peak speed at 120 kHz, mirroring the "imaginary part" of the Clausius-Mossotti factor.

Key Results

  • High-Speed Rotation: Melan-a cells reached speeds of 400 RPM at 20 Vpp (40 kHz). Lymphocytes were even faster, peaking at 1350 RPM at 120 kHz.
  • Voltage Correlation: Rotational speed followed a quadratic relationship with the applied voltage—the higher the voltage, the faster the spin.
  • Sorting Potential: At specific frequencies (like 100 kHz), the researchers achieved a mix of positive and negative DEP forces. This allowed them to attract some white blood cells into the light spot while others were repelled, all while observing different rotation rates.
  • Label-Free Identification: Because the rotation depends on internal dielectric properties, the system can distinguish between cell types (e.g., lymphocytes vs. macrophages) without using any chemical stains or fluorescent labels.

Conclusion and Future Outlook

While the results are compelling, a complete theoretical model for cell rotation in ODEP is still under development. The current explanation regarding mass imbalance is a strong hypothesis but requires more rigorous mathematical validation.

Additionally, cell-cell interactions can influence rotation; if two cells are too close, their local dipoles interfere with each other. Future work will likely focus on using computer vision and AI to automatically identify cell types and even cell-cycle phases (like mitosis) based on these rotational signatures, potentially turning the OEK chip into a high-throughput diagnostic tool.

Summary updated

Abstract

The use of optical dielectrophoresis (ODEP) to manipulate microparticles and biological cells has become increasingly popular due to its tremendous flexibility in providing reconfigurable electrode patterns and flow channels. ODEP enables the parallel and free manipulation of small particles on a photoconductive surface on which light is projected, thus eliminating the need for complex electrode design and fabrication processes. In this paper, we demonstrate that mouse cells comprising melan-a cells, RAW 267.4 macrophage cells, peripheral white blood cells and lymphocytes, can be manipulated in an opto-electrokinetics (OEK) device with appropriate DEP parameters. Our OEK device generates a non-rotating electric field and exerts a localized DEP force on optical electrodes. Hitherto, we are the first group to report that among all the cells investigated, melan-a cells, lymphocytes and white blood cells were found to undergo self-rotation in the device in the presence of a DEP force. The rotational speed of the cells depended on the voltage and frequency applied and the cells' distance from the optical center. We discuss a possible mechanism for explaining this new observation of induced self-rotation based on the physical properties of cells. We believe that this rotation phenomenon can be used to identify cell type and to elucidate the dielectric and physical properties of cells.