Abstract
Subretinal injection is a high-precision ophthalmic procedure requiring the delivery of therapeutic agents into the potential space between the retina and the retinal pigment epithelium (RPE). The extreme delicacy of this task exceeds human manual stability, necessitating robotic assistance. This paper presents a framework for autonomous robotic navigation using intraoperative Optical Coherence Tomography (iOCT). By implementing a novel "intelligent virtual B-scan" slicing method, the system reduces complex 3D volumetric data into manageable 2D tasks. Experimental results in ex-vivo porcine eyes demonstrate micrometer-level navigation precision (32 ± 4 μm) and real-time performance, marking a significant step toward fully automated subretinal surgery.
Introduction
Subretinal injections are increasingly critical for delivering gene and stem cell therapies. However, the target zone—the space between the photoreceptors and the RPE-Bruch’s membrane—is incredibly thin, making manual needle insertion prone to error and tissue trauma. Intraoperative OCT (iOCT) provides the high-resolution 3D visualization necessary to see these layers, but it introduces a "data deluge" problem. Processing full 3D volumes in real-time for robotic guidance is computationally prohibitive and often too slow for the dynamic environment of an operating room.
Furthermore, surgical workflows are complicated by the need for dynamic registration. As surgeons move the microscope's field of view, the coordinate system between the robot and the imaging window shifts. To address these challenges, this research proposes a "surgeon-in-the-loop" framework that leverages dimensionality reduction and deep learning to achieve real-time, autonomous needle navigation.
Methodology
The core innovation of this framework is the simplification of 3D navigation into two consecutive 2D segmentation tasks, ensuring high speed without losing critical depth information.
1. Axial Projection and XY Localization
The system first flattens the 3D iOCT volume into a 2D "top-down" axial projection. A Convolutional Neural Network (CNN)—specifically a U-Net with a ResNet18 backbone—segments the needle footprint. This step determines the needle’s X-Y position and its yaw (θz) relative to the retinal surface.
2. Intelligent Virtual B-Scan Slicing
Rather than processing the entire volume, the system extracts a single, optimized 2D slice called a "virtual B-scan." This slice is automatically aligned with the needle's detected path. This "intelligent" slicing ensures that the needle tip and the underlying retinal layers are always captured in the same frame for depth analysis.
3. Dual-Task Segmentation and Depth Control
A second CNN processes the virtual B-scan to identify the needle tip’s depth (Z-axis) and pitch (θy). Simultaneously, it segments critical retinal layers, including the Inner Limiting Membrane (ILM) and the RPE. To ensure accuracy, the system applies refractive index corrections, accounting for the speed of light variations through air, vitreous, and tissue that can otherwise warp OCT images.
4. Dynamic Registration and Trajectory Planning
The system performs online registration between the robot and the iOCT volume, allowing the robot to maintain its orientation even if the microscope is moved. The trajectory is planned in two phases: first, aligning the needle with the target, and second, performing a straight-line insertion to minimize lateral tissue stress.
Experimental Setup
The models were trained using a combination of synthetic and real-world data:
- ds_axial: 100 iOCT volumes for needle footprint detection.
- ds_bscan: 300 high-resolution images of NanoFil needles and micropipettes in ex-vivo porcine retinas.
- Synthetic Data: Pre-training on synthetic iOCT data improved model robustness before fine-tuning on biological tissue.
Ground truth was established by two biomedical engineering experts who manually labeled tool tip locations across the datasets.
Results
The framework achieved impressive benchmarks in both speed and accuracy:
- Temporal Efficiency: The system reached an inference time of ~0.23 seconds for pose estimation and ~0.19 seconds for trajectory planning, meeting the requirements for real-time surgical intervention.
- Mechanical Precision: The robotic hardware demonstrated a base accuracy of 5 ± 2 μm.
- Navigation Accuracy: In air (no tissue contact), the mean error was 24 ± 5 μm. When tested inside ex-vivo porcine retinas, the navigation error was 32 ± 4 μm.
- Consistency: Kernel Density Estimation (KDE) plots of the error distribution confirmed high repeatability, a vital metric for clinical safety and predictable surgical outcomes.
Discussion and Limitations
While the results are promising, several hurdles remain for clinical adoption. The current system assumes a straight needle geometry; however, micro-surgical tools can often bend under pressure. Future work will involve extracting multiple cross-sectional slices to track non-linear tool shapes.
Additionally, the study focused on translational movements. Real-world surgery requires the robot to respect a "Remote Center of Motion" (RCM) at the scleral entry point to prevent ocular wall damage. Finally, while the algorithms are fast, the physical transmission of high-bandwidth iOCT data from the microscope to the processing unit remains a hardware bottleneck that needs optimization.
Conclusion
This research demonstrates that "intelligent" dimensionality reduction of iOCT data enables autonomous, micrometer-scale robotic navigation. By focusing on virtual B-scans rather than full volumes, the system achieves the real-time performance necessary for delicate subretinal procedures, providing a robust foundation for the next generation of automated ophthalmic surgery.