Robotic Force Stabilization for Beating Heart Intracardiac Surgery

Shelten G. Yuen, Michael C. Yip, Nikolay V. Vasilyev, Douglas P. Perrin, Pedro J. del Nido, Robert D. HoweMICCAI 2009 · OpenReview

Abstract

The manipulation of fast moving, delicate tissues in beating heart procedures presents a considerable challenge to surgeons. We present a new robotic force stabilization system that assists surgeons by maintaining a constant contact force with the beating heart. The system incorporates a novel, miniature uniaxial force sensor that is mounted to surgical instrumentation to measure contact forces during surgical manipulation. Using this sensor in conjunction with real-time tissue motion information derived from 3D ultrasound, we show that a force controller with feed-forward motion terms can provide safe and accurate force stabilization in an in vivo contact task against the beating mitral valve annulus. This confers a 50% reduction in force fluctuations when compared to a standard force controller and a 75% reduction in fluctuations when compared to manual attempts to maintain the same force.
BibTeX
@inproceedings{yuen2009robotic,
  author = {Shelten G. Yuen and Michael C. Yip and Nikolay V. Vasilyev and Douglas P. Perrin and Pedro J. del Nido and Robert D. Howe},
  title = {{Robotic Force Stabilization for Beating Heart Intracardiac Surgery}},
  booktitle = {International Conference on Medical Image Computing and Computer-Assisted Intervention},
  year = {2009},
  url = {https://openreview.net/forum?id=lswLoBTyfk1},
}