Repix: reliable, reusable, versatile chronic Neuropixels implants using minimal components

Mattias Horan, Daniel Regester, Cristina Mazuski, Thomas Jahans-Price, Shanice Bailey, Emmett Thompson, Zuzanna Slonina, Viktor Plattner, Elena Menichini, Irmak Toksöz, Sandra Romero Pinto, Mark Burrell, Isabella Varsavsky, Henry Dalgleish, Célian Bimbard, Anna Lebedeva, Marius Bauža, Francesca Cacucci, Thomas J. Wills, Athena Akrami, Julija Krupic, Marcus Stephenson‐Jones, Caswell Barry, Neil Burgess, John O’Keefe, Yoh IsogaibioRxiv 2024 · DOI

TL;DR

This paper introduces Repix, a simple, robust, and reusable chronic implantation system for Neuropixels probes. Repix enables high-density neural recordings in unrestrained animals across months, even up to a year, facilitating the study of natural behaviors, including social interactions, and significantly lowering the barrier to entry for researchers.

What Problem Does This Paper Address?

High-density electrophysiology using Neuropixels probes has revolutionized neuroscience, allowing simultaneous recording from hundreds of neurons. However, these advancements have primarily benefited head-restrained animals with acute recordings. Chronic implantation in unrestrained animals, which is crucial for studying natural behaviors, learning, and long-term neural dynamics, has remained challenging. Key hurdles include the complexity of implantation, the need for specialized expertise, and the difficulty in reusing expensive Neuropixels probes. This paper addresses the need for a straightforward, reliable, and versatile system that democratizes chronic Neuropixels recordings for a broader range of research questions and users.

Methods

The core of the Repix system is its minimalistic design, comprising only three main components: a probe cassette (payload module), posts (docking module), and a cover. These components are machined from aluminum for durability, particularly important for social behavior experiments. The system utilizes a payload-and-docking principle, allowing probes to be permanently mounted to a cassette, which then slots into posts secured to the animal's skull. This design facilitates both stable chronic implantation and straightforward explantation for probe reuse. The authors also provide detailed surgical protocols and CAD files, emphasizing ease of adoption and customization.

Datasets

The paper does not describe datasets used for training machine learning models. Instead, it presents an evaluation of the Repix system using electrophysiological data collected from 40 animals (mice and rats) across 10 different laboratories. This real-world deployment and user feedback serve as the primary evaluation dataset, demonstrating the system's performance and transferability across various users and experimental paradigms.

Key Insights and Ideas

  • Simplicity as a Virtue: The Repix system's minimal component count and straightforward design are highlighted as key advantages, significantly reducing the learning curve for new users.
  • Robustness for Social Behavior: The use of aluminum components and a secure post-and-cassette mechanism makes Repix resilient enough for demanding social interaction paradigms in mice.
  • Long-Term Stability: Repix enables stable recordings of hundreds of neurons for extended periods (up to a year), crucial for studying neural plasticity and learning.
  • Democratization of Chronic Recordings: By sharing the design and protocol, Repix aims to make advanced chronic electrophysiology accessible to a wider research community, regardless of prior surgical experience.
  • Probe Reusability: The system's design facilitates the reuse of expensive Neuropixels probes, significantly reducing experimental costs.

Key Results

  • High Success Rates: Across 175 Neuropixels implantations and 127 explantations performed by 16 researchers in 10 labs, Repix achieved implantation success rates of 97%, data collection success of 89%, and explantation success of 83%.
  • Extended Recording Durations: Implants were successfully maintained for up to 364 days, demonstrating long-term stability.
  • Stable Unit Yield: Recordings showed stable unit yields over weeks and months, with some brain regions exhibiting minimal initial attenuation.
  • Probe Reuse: 30% of implantations were performed with reused probes, with no significant detriment to initial yield or waveform amplitude observed.
  • Rapid Proficiency: Users became proficient in the implantation procedure after an average of five attempts, regardless of prior surgical experience.
  • Versatility: Repix was successfully used with various Neuropixels probe types (1.0, 2.0 single and multi-shank) and adapted for Neuronexus probes, and demonstrated efficacy in both mice and rats across diverse behavioral tasks.

Limitations and Future Work

  • Weight: At 2.4g, Repix is slightly heavier than some 3D-printed plastic alternatives, which could be a challenge for very small or young mice, especially when combined with a headplate. The authors suggest using heavier mouse strains or implanting without the headstage initially.
  • Dual-Probe Implantation: The current Repix design does not natively support simultaneous dual-probe implantation in mice, though prototypes exist and multiple Repix modules have been used in rats.
  • Anterior Brain Regions: Two users experienced failures when targeting very anterior brain areas, suggesting a need for refinement in such cases.
  • Future Directions: The authors envision future implant development incorporating features directly into the probes for simpler attachment, further miniaturization, and standardized rigs for implantation and explantation. They also highlight the potential for further customization using the provided CAD files.

Summary updated

Abstract

Neuropixels probes represent the state-of-the-art for high-yield electrophysiology in neuroscience: the simultaneous recording of hundreds of neurons is now routinely carried out in head-restrained animals. In contrast, neural recording in unrestrained animals, as well as recording and tracking neurons over days, remains challenging, though it is possible using chronic implants. A major challenge is the availability of simple methods that can be implemented with limited or no prior experience with Neuropixels probes, while achieving reliable, reusable, versatile high-density electrophysiology. Here we developed, deployed, and evaluated the real-world performance of Repix, a chronic implantation system that permits the repeated re-use of Neuropixels probes. The lightweight system allows implanted animals to express a full range of natural behaviors, including social behaviors. We show that Repix allows the recording of hundreds of neurons across many months, up to a year, with implants across cortical and subcortical brain regions. Probes can be reused repeatedly with stable yield. Repix has been used by 16 researchers in 10 laboratories to date, and we evaluated the real-world performance of Repix in a variety of chronic recording paradigms in both mice and rats with a combined 209 implantations. We found that the key advantage of Repix is robustness and simplicity. Adopters of Repix became proficient at five procedures on average, regardless of prior experience with in vivo electrophysiology. With the companion protocol alongside this article, the performance and user-friendliness of Repix should facilitate a wide uptake of chronic Neuropixels recordings.
BibTeX
@misc{horan2024repix,
  author = {Mattias Horan and Daniel Regester and Cristina Mazuski and Thomas Jahans-Price and Shanice Bailey and Emmett Thompson and Zuzanna Slonina and Viktor Plattner and Elena Menichini and Irmak Toksöz and Sandra Romero Pinto and Mark Burrell and Isabella Varsavsky and Henry Dalgleish and Célian Bimbard and Anna Lebedeva and Marius Bauža and Francesca Cacucci and Thomas J. Wills and Athena Akrami and Julija Krupic and Marcus Stephenson‐Jones and Caswell Barry and Neil Burgess and John O’Keefe and Yoh Isogai},
  title = {{Repix: reliable, reusable, versatile chronic Neuropixels implants using minimal components}},
  year = {2024},
  doi = {10.1101/2024.04.25.591118},
  howpublished = {bioRxiv},
}
Figure