PuLsE: Accurate and Robust Ultrasound-based Continuous Heart-Rate Monitoring on a Wrist-Worn IoT Device

Marco Giordano, Christoph Leitner, Christian Vogt, Luca Benini, Michele Magno

Abstract

Continuous heart-rate (HR) monitoring is a cornerstone of modern wearable health technology. While Photoplethysmography (PPG) is the current industry standard, it suffers from limitations regarding skin tone sensitivity and shallow tissue penetration. This paper presents PuLsE (Pulse-echo Ultrasound for Low-power Signal Extraction), a novel wrist-worn IoT prototype that leverages low-power ultrasound (US) to achieve ECG-level accuracy. By implementing an analog envelope filter that reduces signal bandwidth by over 5x, the system enables real-time HR extraction on a standard microcontroller with a total power consumption of just 5.8 mW.

Introduction

Most commercial wearables utilize PPG to estimate heart rate by measuring light absorption changes in the skin. However, PPG is often unreliable due to ambient light interference, variations in skin melanin, and its inability to monitor deep arterial flow. Ultrasound offers a robust alternative, capable of penetrating deep into tissue to monitor pulsatile blood flow directly. Historically, the high sampling rates required for US have made it too power-hungry for continuous wearable use. PuLsE addresses these challenges through hardware-software co-design, optimizing for both energy efficiency and high fidelity.

System Design and Signal Processing

The PuLsE system is built around a low-power US pulser and an ARM Cortex-M4 microcontroller. To overcome the data-rate bottleneck, the researchers introduced an innovative analog envelope filter.

Bandwidth Reduction via Enveloping

Standard ultrasound processing requires high-frequency sampling to capture the radio-frequency (RF) carrier. PuLsE employs an analog envelope filter to extract the signal's "shape" before digitization. This reduces the required sampling rate from tens of MHz to just 4 Msps, a 5x reduction that significantly lowers the computational load on the MCU.

Comparison of raw ultrasound pulses and the filtered envelope signal, demonstrating significant bandwidth reduction.

On-Device Processing

The signal processing pipeline operates entirely on-device using fixed-point arithmetic (q1.15 format). The algorithm uses "M-Mode" imaging, stacking echoes over time to create a 2D representation of tissue movement. Time-differentiation isolates the moving arterial walls from static tissue, and a 2D Fast Fourier Transform (FFT) identifies the dominant heart rate frequency.

Experimental Validation

The system was validated using a dataset from 10 healthy adults, totaling approximately 5 hours of data. Recordings were taken at three wrist positions—lateral, central, and medial—using a Polar H10 ECG chest belt as the ground truth. The study identified the lateral position (near the radial artery) as the optimal site for US-based monitoring due to the superficial nature of the artery and the resulting high Signal-to-Noise Ratio (SNR).

Key Results

PuLsE demonstrated exceptional performance, achieving a Pearson correlation coefficient of r = 0.99 (p < 0.001) compared to ECG in the lateral position. The mean error was recorded at a mere 0.69 ± 1.99 bpm.

Accumulated FFT results showing the detected heart rate peak aligning closely with the ECG ground truth.

Beyond accuracy, the system is highly efficient:

  • Power Consumption: 5.8 mW, making it competitive with existing PPG solutions.
  • Memory Footprint: Only 68 kB of RAM.
  • Latency: 71 ms processing time per window.
  • Battery Life: Estimated to exceed 7 days of continuous operation on a standard 309 mAh smartwatch battery.

Conclusion and Future Work

PuLsE proves that high-fidelity ultrasound is a viable, low-power alternative to PPG for continuous heart-rate monitoring. While the current prototype requires US gel for acoustic coupling and was tested in static conditions, future iterations will explore "dry" coupling materials and motion artifact cancellation for active use cases. The researchers have open-sourced the dataset and code to encourage further development in this promising field of wearable medical technology.

Summary updated

Abstract

This work explores the feasibility of employing ultrasound (US) US technology in a wrist-worn IoT device for low-power, high-fidelity heart-rate (HR) extraction. US offers deep tissue penetration and can monitor pulsatile arterial blood flow in large vessels and the surrounding tissue, potentially improving robustness and accuracy compared to PPG. We present an IoT wearable system prototype utilizing a commercial microcontroller MCU employing the onboard ADC to capture high frequency US signals and an innovative low-power US pulser. An envelope filter lowers the bandwidth of the US signal by a factor of &gt;5x, reducing the system's acquisition requirements without compromising accuracy (correlation coefficient between HR extracted from enveloped and raw signals, r(92)=0.99, p&lt;0.001). The full signal processing pipeline is ported to fixed point arithmetic for increased energy efficiency and runs entirely onboard. The system has an average power consumption of 5.8mW, competitive with PPG based systems, and the HR extraction algorithm requires only 68kB of RAM and 71ms of processing time on an ARM Cortex-M4 MCU. The system is estimated to run continuously for more than 7 days on a smartwatch battery. To accurately evaluate the proposed circuit and algorithm and identify the anatomical location on the wrist with the highest accuracy for HR extraction, we collected a dataset from 10 healthy adults at three different wrist positions. The dataset comprises roughly 5 hours of HR data with an average of 80.6+-16.3 bpm. During recording, we synchronized the established ECG gold standard with our US-based method. The comparisons yields a Pearson correlation coefficient of r(92)=0.99, p&lt;0.001 and a mean error of 0.69+-1.99 bpm in the lateral wrist position near the radial artery. The dataset and code have been open-sourced at https://github.com/mgiordy/Ultrasound-Heart-Rate
BibTeX
@article{giordano2025pulse,
  author = {Marco Giordano and Christoph Leitner and Christian Vogt and Luca Benini and Michele Magno},
  title = {{PuLsE: Accurate and Robust Ultrasound-based Continuous Heart-Rate Monitoring on a Wrist-Worn IoT Device}},
  journal = {IEEE Internet of Things Journal},
  year = {2025},
  doi = {10.1109/jiot.2025.3581380},
}
Figure