Abstract
Continuous heart-rate (HR) monitoring is a cornerstone of modern wearable health technology. While Photoplethysmography (PPG) is the current industry standard, it suffers from limitations regarding skin tone sensitivity and shallow tissue penetration. This paper presents PuLsE (Pulse-echo Ultrasound for Low-power Signal Extraction), a novel wrist-worn IoT prototype that leverages low-power ultrasound (US) to achieve ECG-level accuracy. By implementing an analog envelope filter that reduces signal bandwidth by over 5x, the system enables real-time HR extraction on a standard microcontroller with a total power consumption of just 5.8 mW.
Introduction
Most commercial wearables utilize PPG to estimate heart rate by measuring light absorption changes in the skin. However, PPG is often unreliable due to ambient light interference, variations in skin melanin, and its inability to monitor deep arterial flow. Ultrasound offers a robust alternative, capable of penetrating deep into tissue to monitor pulsatile blood flow directly. Historically, the high sampling rates required for US have made it too power-hungry for continuous wearable use. PuLsE addresses these challenges through hardware-software co-design, optimizing for both energy efficiency and high fidelity.
System Design and Signal Processing
The PuLsE system is built around a low-power US pulser and an ARM Cortex-M4 microcontroller. To overcome the data-rate bottleneck, the researchers introduced an innovative analog envelope filter.
Bandwidth Reduction via Enveloping
Standard ultrasound processing requires high-frequency sampling to capture the radio-frequency (RF) carrier. PuLsE employs an analog envelope filter to extract the signal's "shape" before digitization. This reduces the required sampling rate from tens of MHz to just 4 Msps, a 5x reduction that significantly lowers the computational load on the MCU.
On-Device Processing
The signal processing pipeline operates entirely on-device using fixed-point arithmetic (q1.15 format). The algorithm uses "M-Mode" imaging, stacking echoes over time to create a 2D representation of tissue movement. Time-differentiation isolates the moving arterial walls from static tissue, and a 2D Fast Fourier Transform (FFT) identifies the dominant heart rate frequency.
Experimental Validation
The system was validated using a dataset from 10 healthy adults, totaling approximately 5 hours of data. Recordings were taken at three wrist positions—lateral, central, and medial—using a Polar H10 ECG chest belt as the ground truth. The study identified the lateral position (near the radial artery) as the optimal site for US-based monitoring due to the superficial nature of the artery and the resulting high Signal-to-Noise Ratio (SNR).
Key Results
PuLsE demonstrated exceptional performance, achieving a Pearson correlation coefficient of r = 0.99 (p < 0.001) compared to ECG in the lateral position. The mean error was recorded at a mere 0.69 ± 1.99 bpm.
Beyond accuracy, the system is highly efficient:
- Power Consumption: 5.8 mW, making it competitive with existing PPG solutions.
- Memory Footprint: Only 68 kB of RAM.
- Latency: 71 ms processing time per window.
- Battery Life: Estimated to exceed 7 days of continuous operation on a standard 309 mAh smartwatch battery.
Conclusion and Future Work
PuLsE proves that high-fidelity ultrasound is a viable, low-power alternative to PPG for continuous heart-rate monitoring. While the current prototype requires US gel for acoustic coupling and was tested in static conditions, future iterations will explore "dry" coupling materials and motion artifact cancellation for active use cases. The researchers have open-sourced the dataset and code to encourage further development in this promising field of wearable medical technology.