Precise Pose Recovery of Distal Locking Holes From Single Calibrated Fluoroscopic Image via a Novel Variable Decomposition Approach

Xuan Zhang, Guoyan ZhengISBI 2007 · OpenReview

Abstract

This paper presents a novel variable decomposition approach for pose recovery of the distal locking holes using single calibrated fluoroscopic image. The problem is formulated as a model-based optimal fitting process, where the control variables are decomposed into two sets: (a) the angle between the nail axis and its projection on the imaging plane, and (b) the translation and rotation of the geometrical model of the distal locking hole around the nail axis. By using an iterative algorithm to find the optimal values of the latter set of variables for any given value of the former variable, we reduce the multiple-dimensional model-based optimal fitting problem to a one-dimensional search along a finite interval. We report the results of our in vitro experiments, which demonstrate that the accuracy of our approach is adequate for successful distal locking of intramedullary nails.
BibTeX
@inproceedings{zhang2007precise,
  author = {Xuan Zhang and Guoyan Zheng},
  title = {{Precise Pose Recovery of Distal Locking Holes From Single Calibrated Fluoroscopic Image via a Novel Variable Decomposition Approach}},
  booktitle = {ISBI},
  year = {2007},
  url = {https://openreview.net/forum?id=AgnL9kZKqH},
}

Loading related content...

Figure