Abstract
Accurate quantification of tracer kinetics in dynamic Positron Emission Tomography (dPET) requires a precise Arterial Input Function (AIF). While deep learning models offer a non-invasive alternative to gold-standard blood sampling, they often lack robustness against image artifacts like motion blur. This paper introduces PIDLIF, a physics-informed deep learning framework that embeds biological kinetic modeling into the neural network's loss function. By enforcing physiological consistency, PIDLIF significantly improves AIF estimation accuracy in the presence of severe motion blur, outperforming purely data-driven approaches.
Introduction
In dynamic [18F]FDG PET imaging, researchers track the spatiotemporal distribution of radioactive tracers to quantify metabolic rates. The mathematical modeling of these processes depends heavily on the Arterial Input Function (AIF)—the time-activity curve of the tracer in the blood. Obtaining the AIF typically requires invasive arterial blood sampling, which is technically demanding and often terminal in small-animal studies.
While image-derived input functions (IDIF) and deep learning models have been proposed to automate this process, they are notoriously "brittle." Standard convolutional neural networks (CNNs) rely on sharp spatial features to identify blood vessels. When a subject moves during a scan, the resulting motion blur smears these features, leading to significant errors in AIF estimation. This study addresses this vulnerability by making the estimation process "physics-aware."
Methodology
The PIDLIF Framework
The core innovation of this work is the integration of a Physics-Informed Loss Function ($L_{2TCM}$) into a deep learning architecture (based on FCDLIF). The model uses 3D convolutions to extract spatial features from 4D PET volumes and a temporal extractor to predict the AIF curve.
Kinetic Modeling as a Constraint
Instead of training the model solely to minimize the difference between predicted and ground-truth AIF curves, the researchers introduced a constraint based on the two-tissue compartment model (2TCM). This model describes the exchange of tracer between the blood, free tissue space, and bound intracellular space.
During training, the predicted AIF is plugged into these biological equations to generate theoretical tissue uptake curves for specific regions of interest (the brain and myocardium). The model is then penalized if these predicted tissue curves do not match the actual PET data. This ensures that the estimated AIF is not just a curve that "looks right" but one that is biologically plausible given the observed tissue activity.
Robustness Training
The model was trained on 70 dynamic [18F]FDG PET/CT scans of mice. To simulate real-world movement, the researchers applied 3D blurring kernels (up to 15 voxels) to the images. To maintain stability, the physics-informed loss was gradually "faded in" between training epochs 15 and 30, allowing the model to learn basic features before enforcing complex kinetic constraints.
Key Results
The researchers compared PIDLIF against a baseline data-driven model across varying levels of image degradation:
- Baseline Performance: On clear images, both models performed well ($R^2 \approx 0.94$). However, under severe motion blur (15-voxel kernel), the baseline model's error (RMSE) spiked by 87.2%, and its $R^2$ dropped to 0.79.
- PIDLIF Robustness: Under the same severe blur conditions, the PIDLIF model's error only increased by 25.4%, maintaining a high $R^2$ of 0.89.
- Physiological Consistency: In frames where blood vessels "bled" into surrounding organs due to blur, PIDLIF acted as a "safety rail," preventing the AIF predictions from deviating into biologically impossible territory.
Discussion and Conclusion
The study demonstrates that enforcing physiological consistency is a powerful regularizer for deep learning in medical imaging. While purely data-driven models overfit to specific visual features that disappear under blur, PIDLIF leverages the underlying physics of tracer kinetics to maintain accuracy.
Future work will explore extending this physics-informed approach to voxel-wise modeling and more complex, non-rigid motion patterns. By bridging the gap between deep learning and classical kinetic modeling, PIDLIF provides a more reliable, non-invasive tool for quantitative dynamic PET imaging.
Summary updated