Optimization design of thermal condition of Rowland circle spectrometer based on optical, thermal and structural simulations

Lü Yi, Ren Wanjie, Guo Guojian, You Xinran, Hu Guoxing, Lijun Wu, Xun Sun, Sun Xun, 山东大学 机械工程学院 高效洁净制造教育部重点实验室,山东 济南 250061Journal of Applied Optics 2023 · DOI

Abstract

This study addresses the critical need for thermal optimization in ultraviolet (UV) spectrometers to maintain high spectral resolution. A Rowland circle spectrometer with a detection range of 200 nm to 450 nm and a resolution of at least 0.2 nm was designed. Through coupled optical, thermal, and structural simulations, the optical chamber's thermal conditions were optimized. Results indicate that without active management, thermal equilibrium is difficult to achieve. By optimizing the air inlet speed to 0.8 m/s and implementing a "top-layer" heating method, the overall temperature was stabilized and thermal deformation of optical components was reduced by approximately 40% (from 0.203 mm to 0.122 mm). Optical simulations confirm that the optimized design maintains the required 0.2 nm resolution even under operational thermal stress.

Introduction

Precision optical instruments, particularly ultraviolet (UV) spectrometers, are highly sensitive to temperature fluctuations. Internal heat sources, primarily from Charge-Coupled Device (CCD) sensors, cause thermal expansion and structural shifts in the optical bench. Even microscopic deformations can lead to spectral drift or blurring, compromising the instrument's ability to distinguish between closely spaced wavelengths. While thermal analysis is common for spaceborne instruments, terrestrial UV spectrometers requiring high resolution (0.2 nm) across a broad range (200 nm–450 nm) lack comprehensive thermal optimization research. This paper presents a multi-physics strategy to achieve rapid thermal equilibrium and minimize structural warping.

Methodology

Optical Configuration

The researchers utilized a Rowland circle configuration, which is ideal for the UV spectrum. This design employs a concave grating to both disperse and focus light, minimizing the number of reflective surfaces and reducing the overall error budget associated with multiple moving parts.

Optical structure of the Rowland circle spectrometer

Multi-physics Coupling

A coupled simulation workflow was employed across three domains:

  1. Thermal Simulation: Finite element analysis modeled heat flow from a 4W CCD source. Two primary variables were tested: inlet wind speed and various heating methods (bottom-layer, surrounding, and top-layer).
  2. Structural Analysis: Thermal gradients were converted into physical displacement data to map the deformation of the optical intercept.
  3. Optical Verification: The resulting "warped" geometry was imported into Zemax to verify if the spectrometer could still resolve a 0.2 nm wavelength difference.

Results and Optimization

Thermal Management Strategies

Initial simulations showed that without active management, the base temperature and internal gradients continued to rise, preventing thermal equilibrium. Optimization of the air inlet identified 0.8 m/s as the ideal speed, balancing effective cooling with mechanical stability. This adjustment dropped the overall temperature range to 36.1°C–39.9°C.

Furthermore, the study found that "top-layer heating" was the most effective active method. By maintaining the top of the chamber at a constant 40°C, it acted as a thermal buffer, reducing the maximum thermal deformation of the optical intercept from 0.203 mm to 0.122 mm—a reduction of approximately 39.9%.

Resolution Performance

Post-deformation optical simulations confirmed that the optimized design successfully meets the resolution requirements. As shown in the simulation results, the system can clearly distinguish between three beams with a wavelength difference of 0.2 nm (e.g., 253.4 nm, 253.6 nm, and 253.8 nm) across various slit widths.

Simulated spectral resolution showing the ability to distinguish 0.2 nm wavelength differences

Conclusion

By coupling optical, thermal, and structural simulations, this research demonstrates an effective optimization strategy for high-resolution UV spectrometers. The combination of an optimized 0.8 m/s air inlet and top-layer heating significantly reduces thermal deformation, ensuring that the Rowland circle spectrometer maintains its 0.2 nm resolution under operational heat loads. Future work will focus on experimental validation with physical prototypes and detailed vibration analysis.

Summary updated

Abstract

针对目前紫外光谱仪热工况优化研究缺乏的问题,该文设计出一款探测范围为200 nm~450 nm、全波段分辨率不低于0.2 nm的罗兰圆光谱仪,并耦合光、热、结构模拟对其光室热工况开展优化研究。热仿真结果表明:在无加热、无入口风速下光谱仪底座温度、温差随时间不断增加,难以达到热平衡;优化光室入口风速,发现当入口风速为0.8 m/s时,整体温度降至36.103 ℃~39.859 ℃;基于光学器件间的热变形量的计算,光学器件截距总热变形量为0.203 mm;优化加热方式,发现顶层式加热方式最佳,整体温度降至34.241 ℃~36.139 ℃,光学器件截距总热变形量降至0.122 mm。对此进行光学仿真,结果表明,优化工况后的罗兰圆光谱仪工作热变形后可清晰分辨出两束波长相差0.2 nm的光束。
BibTeX
@article{yi2023optimization,
  author = {Lü Yi and REN Wanjie and GUO Guojian and YOU Xinran and HU Guoxing and Lijun Wu and Xun Sun and SUN Xun and 山东大学 机械工程学院 高效洁净制造教育部重点实验室,山东 济南 250061},
  title = {{Optimization design of thermal condition of Rowland circle spectrometer based on optical, thermal and structural simulations}},
  journal = {Journal of Applied Optics},
  year = {2023},
  doi = {10.5768/jao202344.0101006},
}
Figure