Novel Reconstruction Method to Improve the Sharpness of Free Breathing Cardiac MR Late Gadolinium Enhancement Images on a Commercial 0.55T System
Abstract
Low-field cardiac MR imaging has intrinsically low SNR that degrades the image quality, especially for LGE imaging. Using multiple acquisitions and motion-corrected averaging is a well-established method to improve SNR. However, the ineffectiveness of MOCO at low field leads to blurring. We proposed a novel method that integrates motion correction into compressed sensing and reconstructs a single motion-corrected image. Quantitative sharpness measurements demonstrate that the proposed method improves boundary sharpness, independent of differences in SNR.
BibTeX
@inproceedings{ding2024novel,
author = {Yu Ding and Yingmin Liu and Chong Chen and Ning Jin and Rizwan Ahmad and Orlando P. Simonetti},
title = {{Novel Reconstruction Method to Improve the Sharpness of Free Breathing Cardiac MR Late Gadolinium Enhancement Images on a Commercial 0.55T System}},
booktitle = {Proceedings on CD-ROM - International Society for Magnetic Resonance in Medicine. Scientific Meeting and Exhibition/Proceedings of the International Society for Magnetic Resonance in Medicine, Scientific Meeting and Exhibition},
year = {2024},
doi = {10.58530/2023/5131},