Multi-Channel Hyperspectral Fluorescence Detection Excited by Coupled Plasmon-Waveguide Resonance

Chan Du, Le Liu, Lin Zhang, Jun Guo, Ji-hua Guo, Hui Ma, Yonghong He

TL;DR

This paper introduces a novel biosensor that combines coupled plasmon-waveguide resonance (CPWR) for enhanced fluorescence excitation with hyperspectral imaging for multiplexed detection. This approach achieves a low detection limit of 0.1 nM and can accurately distinguish and quantify spectrally overlapping fluorescent dyes, paving the way for more sophisticated multi-analyte biosensing.

What Problem Does This Paper Address?

Fluorescence spectroscopy is a powerful tool in biological and clinical applications due to its sensitivity and selectivity. However, achieving high throughput and multiplexed detection, especially when dealing with fluorophores that have overlapping emission spectra, remains a challenge. Traditional methods often rely on specific optical filters or limited fluorophore choices. Surface plasmon-enhanced fluorescence (SPEF) offers increased sensitivity by leveraging the enhanced electromagnetic field near a metal surface, but its detection range is limited. To overcome these limitations, researchers are exploring ways to enhance SPEF, such as by incorporating dielectric layers to create coupled plasmon-waveguide resonance (CPWR), which offers a narrower resonance and stronger electromagnetic field. Simultaneously, hyperspectral fluorescence detection, which captures the full emission spectrum for each pixel, has emerged as a promising technique for resolving overlapping spectra and enabling multi-analyte analysis. This paper aims to bridge these advancements by developing a biosensor that integrates the sensitivity enhancement of CPWR with the multiplexing capabilities of hyperspectral imaging.

Methods

The core of this biosensor is a symmetrical optical waveguide structure designed for coupled plasmon-waveguide resonance (CPWR) excitation. This structure consists of three successive layers: MgF₂-Au-MgF₂ deposited on an SF₄ glass substrate. The authors theoretically and experimentally investigated the optimal thicknesses of these layers, particularly the central MgF₂ waveguide layer, to maximize the evanescent field strength and propagation length.

For excitation, a 632.8 nm He-Ne laser is used, p-polarized and focused into a line on the sensor surface. This line illumination, combined with the CPWR effect, creates a dark-field excitation of fluorescence. The emitted fluorescence is then collected and imaged onto the entrance slit of a spectrometer. A spectrometer with a reflective grating provides a high spectral resolution (0.2 nm) over a range of 200 nm. A long-pass filter is employed to remove residual excitation light.

The hyperspectral detection system captures the entire emission spectrum for each point along the illuminated line. This allows for the simultaneous monitoring of multiple channels in a microfluidic system. For data analysis, particularly for distinguishing overlapping spectra, the authors employ multivariate data analysis, specifically linear admixtures, to model the hyperspectral stack and determine the ratios of different fluorescent dyes.

Datasets

Not discussed in the paper. The experiments utilize prepared sensor chips and solutions of Cy5 and Dylight680 dyes at various concentrations.

Key Insights and Ideas

  • Synergy of CPWR and Hyperspectral Imaging: The paper demonstrates that combining CPWR for enhanced fluorescence excitation with hyperspectral imaging for spectral resolution is a powerful strategy for multiplexed biosensing. CPWR boosts the signal, while hyperspectral imaging allows for the deconvolution of complex spectral mixtures.
  • Optimized Symmetrical Waveguide Structure: The symmetrical MgF₂-Au-MgF₂ structure is shown to provide significantly enhanced electromagnetic fields and longer propagation lengths compared to conventional SPR, making it ideal for efficient fluorescence excitation.
  • Resolving Overlapping Spectra: The hyperspectral approach, coupled with multivariate analysis, successfully distinguishes between Cy5 and Dylight680, two dyes with highly overlapping emission spectra, which is a significant challenge for traditional fluorescence detection.
  • Line-Monitoring for Multi-Channel Detection: Illuminating a line on the sensor and capturing spectra along this line enables simultaneous multi-channel detection, greatly enhancing throughput.

Key Results

  • Detection Limit: The biosensor achieved a detection limit of 0.1 nM for Cy5 solutions.
  • Distinguishing Overlapping Dyes: The system successfully distinguished and quantified the contributions of Cy5 and Dylight680 in a mixed solution, demonstrating the effectiveness of hyperspectral analysis and multivariate algorithms.
  • Linearity: A linear relationship was observed between normalized fluorescence intensity and Cy5 concentration over the tested range, indicating good quantitative performance.
  • Enhanced Field: The CPWR structure provides an evanescent field decay distance of approximately 2,000 nm, significantly longer than the ~200 nm of conventional SPR.

Limitations and Future Work

  • Limited Number of Dyes Tested: While the paper demonstrates the ability to distinguish two overlapping dyes, the authors suggest that the system has potential for detecting more than two tags with partially overlapping spectra, implying further investigation is needed for higher multiplexing.
  • Optimization of Multivariate Algorithms: The paper presents a successful multivariate algorithm for spectral deconvolution. Future work could explore more advanced algorithms for improved accuracy and robustness.
  • Application Expansion: The authors highlight the potential for applications in biochemical interactions and DNA microarray technology. Further experimental validation in these specific domains would be beneficial.
  • Sensor Chip Fabrication: While the fabrication process is described, further optimization for mass production and cost-effectiveness could be a future direction.

Summary updated

Abstract

We propose in this paper a biosensor scheme based on coupled plasmon-waveguide resonance (CPWR) excited fluorescence spectroscopy. A symmetrical structure that offers higher surface electric field strengths, longer surface propagation lengths and depths is developed to support guided waveguide modes for the efficient excitation of fluorescence. The optimal parameters for the sensor films are theoretically and experimentally investigated, leading to a detection limit of 0.1 nM (for a Cy5 solution). Multiplex analysis possible with the fluorescence detection is further advanced by employing the hyperspectral fluorescence technique to record the full spectra for every pixel on the sample plane. We demonstrate experimentally that highly overlapping fluorescence (Cy5 and Dylight680) can be distinguished and ratios of different emission sources can be determined accurately. This biosensor shows great potential for multiplex detections of fluorescence analytes.
BibTeX
@inproceedings{du2013multi,
  author = {Chan Du and Le Liu and Lin Zhang and Jun Guo and Ji-hua Guo and Hui Ma and Yonghong He},
  title = {{Multi-Channel Hyperspectral Fluorescence Detection Excited by Coupled Plasmon-Waveguide Resonance}},
  booktitle = {Sensors},
  year = {2013},
  url = {https://openreview.net/forum?id=Rz82PvdyYml},
}
Figure