Abstract
Trichophyton is the most pathogenic type of fungal skin infection. It often invades and grows in a keratin-rich matrix, and lesions include human skin, hair, and fingernails (toenails). We designed LAMP primers for Trichophyton and developed a LAMP-Microfluidic chip detection system for Trichophyton. This system detects six common species of Trichophyton in the genus Trichophyton, including Trichophyton rubrum, Trichophyton mentagrophyte, Trichophyton violaceum, Trichophyton tonsurans, Trichophyton verrucosum, and Trichophyton schoenleinii. The specificity reached 100%, and the sensitivity could reach about 1 × 10^2 copies/μl. The entire detection process can be completed within 60 min and does not cross-react with other dermatophytes. The established LAMP-Microfluidic chip detection system has the advantages of simple operation, high specificity, and high sensitivity, and has the potential for clinical application.
Introduction
Superficial fungal infections caused by the genus Trichophyton represent a significant global health burden, affecting nearly 25% of the population. These pathogens cause common conditions such as tinea capitis (ringworm of the scalp) and onychomycosis (nail fungus) by invading keratin-rich tissues like skin, hair, and nails.
Current diagnostic methods face significant limitations. Microscopy is rapid but subjective, with false-negative rates between 15–30%. Fungal culture, the gold standard, can take up to two weeks, often delaying appropriate treatment. While PCR offers high accuracy, it requires expensive thermal cycling equipment and specialized environments. There is a pressing need for a rapid, specific, and simple Point-of-Care (POC) diagnostic tool.
Methods
The researchers developed a "lab-on-a-chip" platform integrating Loop-mediated Isothermal Amplification (LAMP) with centrifugal microfluidics.
LAMP Integration
Unlike PCR, which requires cycling through different temperatures, LAMP operates at a constant temperature (63°C). This significantly simplifies hardware requirements. The team screened multiple primer sets, selecting a set that provided the highest specificity for the Trichophyton genome.
Centrifugal Microfluidic Chip
The system utilizes a disk-shaped microfluidic chip designed for automation. The chip contains 8 sample feeding slots, each connected to 4 reaction chambers, totaling 32 wells.
By applying centrifugal force through rotation, reagents are moved through microchannels into the reaction chambers. This closed system automates mixing and minimizes the risk of aerosol contamination. The entire workflow—including a 15-minute nucleic acid extraction and a 40-minute amplification phase—is completed within an hour.
Results and Validation
The system was validated using a library of 111 fungal and bacterial strains, including 66 target Trichophyton strains and 45 non-target control strains (e.g., Microsporum, Candida, and Aspergillus).
Specificity and Sensitivity
The detection system achieved 100% specificity, showing no cross-reactivity with non-target dermatophytes or bacteria. The sensitivity reached approximately $1 \times 10^2$ copies/µl, which is sufficient for clinical diagnostic requirements.
Speed and Reliability
Positive samples consistently showed amplification within 40 minutes, while negative controls remained flat even after 60 cycles. This rapid turnaround time offers a significant improvement over traditional culture methods.
Discussion
The LAMP-microfluidic chip system provides a portable, energy-efficient alternative to PCR. By utilizing centrifugal force for fluid management, the design avoids complex internal pumps or valves, making it more suitable for clinical settings.
Future work will focus on differentiating between specific Trichophyton species within a single run and optimizing the extraction process for complex clinical matrices like nail clippings. Despite these remaining steps, the system demonstrates high potential as a rapid POC diagnostic tool for fungal infections.