Iterative Reconstruction for Low-Dose CT using Deep Gradient Priors of Generative Model

Yi Zhang, Zhuonan He, Yikun Zhang 0001, Yu Guan, Shanzhou Niu, Yi Zhang 0018, Yang Chen, Qiegen Liu

Abstract

Low-dose computed tomography (LDCT) is essential for reducing radiation risks but often suffers from significant noise and artifacts. While supervised deep learning methods have shown promise, they frequently struggle with generalization and require extensive paired datasets. This paper presents EASEL (dEep grAdient priorS of genErative modeL), an unsupervised framework that integrates score-based generative models into an iterative reconstruction pipeline. By learning the gradient of the data distribution (the "score") from normal-dose images, EASEL provides a robust prior that, when combined with a physical data-fidelity term via Separable Quadratic Surrogates (SQS), achieves state-of-the-art reconstruction quality. Experimental results on the AAPM Grand Challenge dataset demonstrate that EASEL outperforms existing supervised and iterative methods in both noise suppression and fine detail preservation, while exhibiting superior robustness to varying noise levels.

Introduction

Computed Tomography (CT) is a cornerstone of modern medical diagnostics. However, the ionizing radiation associated with standard-dose CT scans poses potential health risks, including carcinogenesis. Reducing the X-ray dose (LDCT) is a primary clinical goal, but it leads to degraded image quality characterized by severe noise and streaking artifacts.

Traditional solutions rely on Iterative Reconstruction (IR) with hand-crafted priors, such as Total Variation (TV). While mathematically grounded, these often produce "blocky" artifacts and lose subtle textures. Conversely, supervised deep learning (DL) models learn direct mappings from low-dose to normal-dose images. Despite their high performance, these models are often "black boxes" that require large paired datasets and lack robustness when the test data's noise characteristics deviate from the training set.

Taxonomy of deep generative models, highlighting the position of score-based models.

To address these limitations, the authors propose EASEL, which leverages score-based generative models. As shown in the taxonomy above, these models fall under explicit generative models that learn the gradient of the log-probability density ($\nabla_x \log p(x)$), providing a powerful unsupervised prior for iterative solvers.

Methodology: The EASEL Framework

The EASEL framework treats CT reconstruction as a posterior sampling problem, combining a learned image prior with a physical data-fidelity term.

Score-based Generative Prior

Instead of estimating the probability density $p(x)$ directly, EASEL learns the score function—the vector field pointing toward regions of higher data density. This is achieved through Denoising Score Matching (DSM), where a network is trained to estimate the score of a noise-perturbed version of normal-dose images. This approach is unsupervised, requiring only high-quality images for training.

Annealed Langevin Dynamics and Data Consistency

Reconstruction is performed using Annealed Langevin Dynamics. Starting from a noisy initial estimate, the algorithm iteratively updates the image by following the learned score function. To ensure the result remains faithful to the actual X-ray measurements, a data-fidelity term is integrated using the Separable Quadratic Surrogates (SQS) algorithm.

The EASEL training and reconstruction pipeline, illustrating the integration of score-matching and iterative SQS updates.

As illustrated in the pipeline diagram, the process alternates between the generative prior (moving the image toward the manifold of clean CT scans) and the SQS step (ensuring consistency with the raw sinogram data $y$). A momentum update is further applied to accelerate convergence and stabilize the trajectory.

Efficient Architecture: Channel-Copy Guided RefineNet

To handle the high dimensionality of CT data efficiently, the authors employ a modified RefineNet. A "channel-copy" strategy is used at the input to expand the single-channel CT image into multiple feature channels without the computational cost of high-dimensional initial convolutions.

Experimental Results

The framework was evaluated on the AAPM Low Dose CT Grand Challenge dataset and anthropomorphic CIRS Phantom data.

Quantitative Performance

EASEL achieved a Peak Signal-to-Noise Ratio (PSNR) of 41.58 dB on the AAPM dataset, surpassing leading supervised models like DP-ResNet (41.03 dB) and RED-CNN (39.37 dB). Notably, EASEL maintained high performance even when the noise level in the test data differed from the training assumptions, whereas supervised models experienced significant degradation.

Qualitative Analysis

Visual comparison revealed that EASEL effectively suppresses noise while preserving sharp edges and fine anatomical structures. Unlike TV-based methods that often "over-smooth" images, EASEL maintains the natural texture of the tissue, which is critical for accurate clinical diagnosis.

Conclusion

EASEL represents a significant step forward in LDCT reconstruction by bridging the gap between unsupervised generative modeling and physics-based iterative optimization. By utilizing deep gradient priors, it offers a robust, high-fidelity solution that generalizes better than traditional supervised networks. Future work will focus on improving convergence speed and extending the framework to 3D volumetric reconstruction.

Summary updated

Abstract

Dose reduction in computed tomography (CT) is essential for decreasing radiation risk in clinical applications. Iterative reconstruction is one of the most promising ways to compensate for the increased noise due to the reduction of photon flux. Rather than most existing prior-driven algorithms that benefit from manually designed prior functions or supervised learning schemes, in this work, we integrate the data consistency as a conditional term into the iterative generative model for low-dose CT. At the stage of prior learning, the gradient of data density is directly learned from normal-dose CT images as a prior. Then, at the iterative reconstruction stage, the stochastic gradient descent is employed to update the trained prior with annealed and conditional schemes. The distance between the reconstructed image and the manifold is minimized along with data fidelity during reconstruction. Experimental comparisons demonstrated the noise reduction and detail preservation abilities of the proposed method.
BibTeX
@article{zhang2022iterative,
  author = {Yi Zhang and Zhuonan He and Yikun Zhang and Yu Guan and Shanzhou Niu and Yi Zhang and Yang Chen and Qiegen Liu},
  title = {{Iterative Reconstruction for Low-Dose CT using Deep Gradient Priors of Generative Model}},
  journal = {IEEE Transactions on Radiation and Plasma Medical Sciences},
  year = {2022},
  doi = {10.1109/trpms.2022.3148373},
}
Figure