Abstract
Background: With the development of laser-assisted platforms, the outcomes of cataract surgery have been improved by automating several procedures. The cataract-extraction step continues to be manually performed, but due to deficiencies in sensing capabilities, surgical complications such as posterior capsule rupture and incomplete cataract removal remain. Methods: An optical coherence tomography (OCT) system is integrated into our intraocular robotic interventional surgical system (IRISS) robot. The OCT images are used for preoperative planning and intraoperative intervention in a series of automated procedures. Real-time intervention allows surgeons to evaluate the progress and override the operation. Results: The developed system was validated by performing lens extraction on 30 postmortem pig eyes. Complete lens extraction was achieved on 25 eyes, and “almost complete” extraction was achieved on the remainder due to an inability to image small lens particles behind the iris. No capsule rupture was found. Conclusion: The IRISS successfully demonstrated semiautomated OCT-guided lens removal with real-time supervision and intervention.
Introduction
Cataract surgery is the most frequently performed surgical procedure in the United States. While laser-assisted platforms have successfully automated initial incisions, the critical "extraction" phase—breaking up and aspirating the clouded lens—remains a high-stakes manual task. This stage requires surgeons to operate within millimeters of the posterior capsule, a membrane only a few microns thick.
Accidental puncture of the posterior capsule (rupture) can lead to permanent vision loss, while incomplete removal of lens fragments can cause significant post-operative inflammation. The physiological limits of human hand stability and visual depth perception present inherent risks that robotic intervention aims to mitigate through superior sensing and precision.
System Design and Methodology
The researchers at UCLA integrated the Intraocular Robotic Interventional Surgical System (IRISS) with high-resolution Spectral-Domain Optical Coherence Tomography (SD-OCT). The system operates on four primary pillars:
- Automated Alignment: Using OCT scans, the system identifies the corneal incision and aligns the robot’s Remote Center of Motion (RCM). This ensures the surgical tool pivots precisely at the entry point, minimizing mechanical stress on ocular tissue.
- 3D Anatomical Modeling: Prior to extraction, volume scans create a 3D map of the cornea, pupil, and lens. This map defines "safe zones" and "no-go zones" to protect the posterior capsule.
- Flower-Shaped Trajectory: A novel tool-tip path was designed to mobilize lens material toward the center for aspiration. This pattern includes a "scooping" component (adjusting the tool's angle, $\theta_4$) that creates intraocular turbulence, helping to detach stubborn lens pieces.
- Semi-Automated Control: The robot follows the pre-planned path while the surgeon monitors the procedure via a real-time interface, retaining the ability to pause or manually override the system.
A key technical insight was the use of a 1060 nm light source for the OCT. This wavelength provides an optimal balance between image resolution and the depth penetration required to visualize dense lens structures in pig models.
Experimental Results
The IRISS was evaluated using 30 post-mortem pig eyes, which possess a lens structure similar to humans but with greater thickness. The accuracy of the 3D modeling algorithm was validated against human-labeled ground truths from 10 randomly sampled OCT images.
The system demonstrated exceptional safety and precision:
- Safety: Zero posterior capsule ruptures occurred across all 30 trials.
- Efficacy: Complete lens removal was achieved in 25 out of 30 eyes. The remaining five cases were classified as "almost complete," with minor particles remaining in "shadow zones" behind the iris where OCT line-of-sight is obstructed.
- Accuracy Metrics: The system achieved a registration accuracy of 0.19 mm, modeling accuracy of 79.6 µm, and tracking accuracy of 71 µm.
- Efficiency: The average procedure time was 277 seconds, outperforming previous teleoperated robotic benchmarks.
Discussion and Future Work
While the IRISS successfully demonstrated the feasibility of semi-automated cataract removal, imaging "blind spots" behind the iris remain a challenge. Future iterations may incorporate ultrasound biometry or multi-angle imaging to address these areas.
Transitioning to live subjects will require the integration of robust eye-tracking algorithms to compensate for involuntary microsaccades and the use of suction-based fixation devices. Additionally, the team aims to refine tissue-detection algorithms to automatically distinguish between lens material and intraocular air bubbles in real-time.