Implementation of the HexaPOD™ evo RT system for conventionally fractionated head and neck radiotherapy

Tsung-Yi Ho, R.p. Edwards, Thulani Nyathi, Rachel Kearvell, Caroline ConnellInternational Journal of Radiology & Radiation Therapy 2024 · DOI

Introduction

In radiotherapy for head and neck cancer (HNC), precision is paramount. Treatment volumes are often extensive—frequently exceeding 20 cm—and the anatomy is inherently flexible. Traditional patient setup corrections typically rely on three translational degrees of freedom (up/down, left/right, in/out). However, these methods fail to account for rotational errors (pitch, roll, and yaw).

Due to the "lever effect," even a minor rotational error at a pivot point can lead to significant geographical displacements at the distal ends of a long treatment field. Such mismatches can result in under-dosing the target or over-dosing adjacent critical structures. This study evaluates the clinical feasibility of implementing the HexaPOD™ evo RT system, a 6-degree-of-freedom (6 DOF) robotic couch, to correct these rotations in daily conventional radiotherapy.

Methodology

The study was conducted in two distinct phases at GenesisCare Oncology in Australia:

  1. Retrospective Phase: Researchers analyzed 100 pre-treatment Cone Beam CT (CBCT) images from 10 previous HNC patients. They compared standard 3D translational registrations with 6D registrations (incorporating rotations) using "Maximising Mutual Information" algorithms to determine if adding rotational corrections significantly altered the required shifts.
  2. Prospective Phase: The HexaPOD™ evo RT system was integrated into the daily workflow for 20 new HNC patients (16 male, 4 female; ages 33–98). A "decision-making matrix" was developed to guide therapists on applying shifts, re-imaging, or resetting patients. The iGUIDE® software was used to translate registration data into physical couch movements.

In total, the prospective phase involved 622 partial arc CBCT scans, comparing 6D results against the clinic's previous 3D-only standards.

Results and Discussion

The study confirmed that rotational corrections are essential for maintaining dose distribution in long HNC fields. In the retrospective phase, the difference between 3D and 6D image registrations was found to be statistically significant (p < 0.05), indicating that 3D-only shifts leave residual errors.

Lateral shift variability across ten patients in the retrospective phase.

Key findings from the prospective implementation include:

  • Rotational Prevalence: 97% of all CBCT scans required a rotation of at least ±0.5°.
  • The Impact of Pitch: Pitch (forward/backward head tilt) was the most frequent source of error, with 30% of shifts exceeding 1°.
  • Significant Mismatches: 23% of average rotational shifts were between 1° and 2°. A 2° rotation can cause a 3.5 mm mismatch at the edge of a 20 cm field, often exceeding standard clinical margins.
  • Operational Efficiency: Despite the added complexity, treatment slots were maintained at 12 minutes, demonstrating that 6 DOF technology can be integrated into a busy clinic without compromising throughput.

Conclusion and Future Directions

The implementation of the HexaPOD™ system successfully improved setup accuracy for HNC patients. While the study demonstrated positional benefits, it was limited by the lack of post-treatment scans to assess intrafraction motion and did not explicitly calculate the dosimetric impact of these corrections. Future research should focus on larger multi-center cohorts and direct dosimetric comparisons to further quantify the clinical advantages of 6 DOF robotic couches in conventional head and neck cancer care.

Summary updated

Abstract

Introduction: Head and neck cancer (HNC) treatments often involve long treatment fields whose treatment accuracy can be impacted by rotational and translational inconsistencies. Studies have demonstrated that even small rotational setup errors can result in large displacements at the ends of the target. One method for correcting the treatment position in six dimensions (6D) is to use a robotic couch. This study aimed to determine if using one to facilitate the correction of the treatment position was feasible for HNC patients treated in our centre and to report on the implementation experience. Methods: This two-phase study involved a retrospective review of 100 pre-treatment registration images to establish the magnitude of the rotations followed by a prospective analysis of daily pre-treatment images for 20 patients. Results: The mean difference of the vector shift for the first phase, was statistically significant (p-value &lt; 0.05) between image registrations with 6D compared with those in three dimensions. The post imaging rotational shifts for the second phase, ranged as follows: pitch (-1.1° to 2.0°); roll (-0.9° to 1.7°) and yaw (-1.1° to 1.8°). The pitch reported the greatest average shift (30% &gt;1°). 23% of the average rotational shifts were &gt; 1° and ≤ 2°. Conclusion: Centres that are equipped with the HexaPOD™ system can consider its use in head and neck treatment to improve set up and treatment accuracy.
BibTeX
@article{ho2024implementation,
  author = {Tsung-Yi Ho and R.P. Edwards and Thulani Nyathi and Rachel Kearvell and Caroline Connell},
  title = {{Implementation of the HexaPOD™ evo RT system for conventionally fractionated head and neck radiotherapy}},
  journal = {International Journal of Radiology \& Radiation Therapy},
  year = {2024},
  doi = {10.15406/ijrrt.2024.11.00398},
}
Figure