Introduction
In radiotherapy for head and neck cancer (HNC), precision is paramount. Treatment volumes are often extensive—frequently exceeding 20 cm—and the anatomy is inherently flexible. Traditional patient setup corrections typically rely on three translational degrees of freedom (up/down, left/right, in/out). However, these methods fail to account for rotational errors (pitch, roll, and yaw).
Due to the "lever effect," even a minor rotational error at a pivot point can lead to significant geographical displacements at the distal ends of a long treatment field. Such mismatches can result in under-dosing the target or over-dosing adjacent critical structures. This study evaluates the clinical feasibility of implementing the HexaPOD™ evo RT system, a 6-degree-of-freedom (6 DOF) robotic couch, to correct these rotations in daily conventional radiotherapy.
Methodology
The study was conducted in two distinct phases at GenesisCare Oncology in Australia:
- Retrospective Phase: Researchers analyzed 100 pre-treatment Cone Beam CT (CBCT) images from 10 previous HNC patients. They compared standard 3D translational registrations with 6D registrations (incorporating rotations) using "Maximising Mutual Information" algorithms to determine if adding rotational corrections significantly altered the required shifts.
- Prospective Phase: The HexaPOD™ evo RT system was integrated into the daily workflow for 20 new HNC patients (16 male, 4 female; ages 33–98). A "decision-making matrix" was developed to guide therapists on applying shifts, re-imaging, or resetting patients. The iGUIDE® software was used to translate registration data into physical couch movements.
In total, the prospective phase involved 622 partial arc CBCT scans, comparing 6D results against the clinic's previous 3D-only standards.
Results and Discussion
The study confirmed that rotational corrections are essential for maintaining dose distribution in long HNC fields. In the retrospective phase, the difference between 3D and 6D image registrations was found to be statistically significant (p < 0.05), indicating that 3D-only shifts leave residual errors.
Key findings from the prospective implementation include:
- Rotational Prevalence: 97% of all CBCT scans required a rotation of at least ±0.5°.
- The Impact of Pitch: Pitch (forward/backward head tilt) was the most frequent source of error, with 30% of shifts exceeding 1°.
- Significant Mismatches: 23% of average rotational shifts were between 1° and 2°. A 2° rotation can cause a 3.5 mm mismatch at the edge of a 20 cm field, often exceeding standard clinical margins.
- Operational Efficiency: Despite the added complexity, treatment slots were maintained at 12 minutes, demonstrating that 6 DOF technology can be integrated into a busy clinic without compromising throughput.
Conclusion and Future Directions
The implementation of the HexaPOD™ system successfully improved setup accuracy for HNC patients. While the study demonstrated positional benefits, it was limited by the lack of post-treatment scans to assess intrafraction motion and did not explicitly calculate the dosimetric impact of these corrections. Future research should focus on larger multi-center cohorts and direct dosimetric comparisons to further quantify the clinical advantages of 6 DOF robotic couches in conventional head and neck cancer care.