Image-Based Methods for Phase Estimation, Gating, and Temporal Superresolution of Cardiac Ultrasound

Deepak R. Chittajallu, Matthew McCormick, Samuel Gerber, Tomasz J. Czernuszewicz, Ryan C. Gessner, Monte S. Willis, Marc Niethammer, Roland Kwitt, Stephen Aylward

Abstract: Our methods can enable tracking of cardiorespiratory phases without additional hardware and reconstruction of respiration-free single cardiac-cycle videos at a much higher temporal resolution.

Introduction

In preclinical cardiac imaging, particularly involving small animals like mice, researchers face significant technical hurdles. Mice exhibit extremely high heart rates (up to 840 BPM) and rapid respiratory cycles. Tracking these phases typically requires bulky, expensive external hardware such as Electrocardiograms (ECG). Furthermore, standard commercial ultrasound transducers often lack the temporal resolution to capture rapid events like peak systole. This paper presents a sophisticated image-processing pipeline that extracts cardiac and respiratory information directly from ultrasound videos, eliminating the need for external sensors and enabling high-resolution temporal reconstruction.

Methodology

The proposed retrospective pipeline consists of three primary stages:

  1. Phase Estimation: The video is converted into a univariate time series by calculating inter-frame similarity using normalized correlation. To separate the respiratory signal from the heartbeat, the authors employ a Hodrick-Prescott (HP) filter, treating respiration as a low-frequency trend and the heartbeat as a high-frequency residual. A Hilbert transform is then applied to these decoupled signals to determine the instantaneous phase of every frame.
  2. Robust Gating: To eliminate frames blurred or displaced by respiration, a two-step gating process is used. After an initial cut based on respiratory phase, LOWESS (Locally Weighted Regression) models the relationship between cardiac phase and image similarity, allowing for the identification and removal of outliers.
  3. Temporal Super-resolution: High-frame-rate videos are synthesized using Nadaraya-Watson (NW) kernel regression. This technique reconstructs images for arbitrary cardiac phases by taking a weighted average of nearby frames. A bivariate kernel weights frames based on both phase proximity and similarity to a "clean" heartbeat model, effectively suppressing respiratory artifacts.
Figure 2: Decoupling of respiration and heartbeat signals using similarity matrices, HP filtering, and Hilbert transforms for phase estimation.

Key Results

The method was validated using 2D Echocardiography videos from anesthetized mice, with simultaneous ECG recordings as the gold standard. Key findings include:

  • High Accuracy: Cardiac phase estimation showed a mean error of only 3%–6% compared to ECG.
  • Superior Peak Detection: The method achieved a mean-frame-error of 0.73–1.36 frames in locating R-wave peaks, significantly outperforming existing image-based techniques.
  • Visual Clarity: M-mode visualizations demonstrated the successful removal of "jagged" discontinuities caused by respiratory motion, resulting in a smooth, continuous representation of the cardiac cycle.
  • Reconstruction Quality: The kernel-regression model achieved a mean-normalized correlation of 0.81–0.85 against ground truth in cross-validation.
Figure 3: M-mode sequences comparing raw data (d) with sequences ordered by ECG (e), the proposed cardiac phase (f), and the final respiratory-gated result (g).

Discussion and Future Work

By explicitly decoupling respiration as a trend and the heartbeat as a periodic residual, this method proves more robust than implicit manifold learning approaches. The use of the Hilbert transform further stabilizes phase estimation against minor fluctuations.

However, the current model assumes a relatively steady heart rhythm and may struggle with severe cardiac arrhythmia. Additionally, the use of global similarity measures might overlook subtle, localized wall motion abnormalities. Future work may involve moving toward "manifold kernel regression" using diffeomorphic registration to preserve sharp edges and reduce blurring at very high temporal magnifications.

Summary updated

Abstract

Our methods can enable tracking of cardiorespiratory phases without additional hardware and reconstruction of respiration-free single cardiac-cycle videos at a much higher temporal resolution.
BibTeX
@article{chittajallu2018image,
  author = {Deepak R. Chittajallu and Matthew McCormick and Samuel Gerber and Tomasz J. Czernuszewicz and Ryan C. Gessner and Monte S. Willis and Marc Niethammer and Roland Kwitt and Stephen Aylward},
  title = {{Image-Based Methods for Phase Estimation, Gating, and Temporal Superresolution of Cardiac Ultrasound}},
  journal = {IEEE Transactions on Biomedical Engineering},
  year = {2018},
  doi = {10.1109/tbme.2018.2823279},
}
Figure