High‐Precision, Low‐Threshold Neuromodulation With Ultraflexible Electrode Arrays for Brain‐to‐Brain Interfaces

Yifei Ye, Ye Tian, Haifeng Liu, Jiaxuan Liu, Cunkai Zhou, Cheng‐Jian Xu, Ting Zhou, Yanyan Nie, Yu Wu, Lunming Qin, Zhitao Zhou, Xiaoling Wei, Jianlong Zhao, Zhenyu Wang, Meng Li, Tiger H. Tao, Liuyang SunExploration 2025 · DOI

TL;DR

Researchers have developed an ultraflexible neural electrode array that enables precise, low-current neuromodulation in mice, significantly improving upon traditional rigid electrodes. This technology has been successfully applied to create a brain-to-brain interface, allowing humans to control mouse turning behavior with high accuracy.

What Problem Does This Paper Address?

Traditional methods for neuromodulation, which is crucial for understanding the brain and treating neurological disorders, often rely on rigid electrodes. These rigid implants face several challenges: they can cause tissue damage due to mechanical mismatch with the soft brain, require high stimulating currents leading to imprecise targeting, and can be dislodged by natural brain movements. This limits their effectiveness and safety for long-term use. The paper addresses the need for a more stable, precise, and less invasive neuromodulation tool that can also be integrated into advanced applications like brain-to-brain interfaces.

Methods

The core of this research is the development of an ultraflexible neural electrode array. Key aspects of their methodology include:

  • Fabrication: A multi-step microfabrication process was used to create the flexible probes, incorporating a PtIr coating on the electrodes to enhance electrical properties and reduce impedance.
  • Neuromodulation: The array allows for both neural recording and electrical stimulation. They demonstrated precise control of mouse turning behavior by stimulating specific regions of the secondary motor cortex (M2).
  • Low-Threshold Stimulation: They found that currents as low as ~5 µA were sufficient to induce turning behavior, a significant reduction compared to conventional rigid electrodes.
  • High Spatial Resolution: The densely packed microelectrodes allowed for targeted stimulation of specific brain areas.
  • Durability and Biocompatibility: In vitro fatigue tests (10 million pulses) and in vivo implantation studies (6 months) were conducted to assess the electrodes' long-term stability and biocompatibility, comparing them to traditional rigid microwires.
  • Brain-to-Brain Interface (B2BI): They integrated the flexible electrode array with human brain signal detection (SSVEP paradigm) and advanced deep neural networks (DNNs) for decoding. This allowed a human user to issue commands that controlled the turning behavior of two mice simultaneously.

Datasets

The paper does not explicitly mention pre-existing public datasets used for training or evaluation. Instead, the authors generated their own datasets through experiments involving freely moving mice and human participants for the B2BI system.

  • Mouse Behavior Data: Collected from mice undergoing electrical stimulation, including movement velocity, neuronal firing patterns, and angular/linear displacements.
  • Human EEG Data: Collected from human participants using SSVEP paradigms, with signals recorded from the occipital region.

Key Insights and Ideas

  • Flexibility is Key: The ultraflexible nature of the electrodes is crucial for achieving stable, long-term neural interfaces that minimize tissue damage and maintain consistent electrical contact.
  • Low Current, High Precision: The research demonstrates that effective neuromodulation, even for complex behaviors like turning, can be achieved with remarkably low currents, enhancing precision and reducing collateral effects.
  • Targeted Stimulation Works: By precisely targeting specific brain regions (like M2), specific behavioral outputs can be reliably elicited, opening doors for more nuanced brain control.
  • B2BI is Feasible and Scalable: The successful implementation of a human-to-two-mice B2BI system shows the potential for controlling multiple animals with a single human operator, a significant step towards scalable brain interfaces.
  • DNNs are Essential for B2BI: Advanced deep neural networks are critical for decoding complex brain signals (like SSVEPs) and translating them into precise commands for neuromodulation.

Key Results

  • Low-Threshold Stimulation: Mouse turning behavior was effectively induced with approximately 5 µA of stimulating current, which is 1-2 orders of magnitude lower than conventional rigid electrodes.
  • Precise Turning Control: Stimulation of the right M2 reliably induced leftward turning, and stimulation of the left M2 induced rightward turning.
  • Durability: Electrodes maintained stable impedance after 10 million biphasic pulse stimulations and 4 weeks of PBS immersion.
  • Biocompatibility: After 6 months of implantation, the flexible probes showed significantly less tissue damage and glial scarring compared to traditional rigid microwires.
  • B2BI Accuracy: The B2BI system achieved high accuracy in decoding human SSVEP signals (up to 98.1% for 8 targets) and controlling mouse turning behavior (over 95% accuracy for controlling two mice simultaneously).

Limitations and Future Work

  • Limitations:
    • The paper notes that for precise directional turning, stimulating broader areas was more effective than narrow subregions, suggesting that the precise neural circuits involved are not yet fully understood.
  • While the B2BI system demonstrated impressive accuracy, the current implementation involves specific visual stimuli (SSVEP) for human input.
  • Future Work:
    • The authors suggest that this technology opens new avenues for interspecies communication and potential neuromodulation applications.
  • Further research could explore controlling more complex behaviors or integrating other forms of human brain signal detection.
  • The scalability of the 1:N brain-to-brain interface is highlighted as a significant future direction.

Summary updated

Abstract

Neuromodulation is crucial for advancing neuroscience and treating neurological disorders. However, traditional methods using rigid electrodes have been limited by large stimulating currents, low precision, and the risk of tissue damage. In this work, we developed a biocompatible ultraflexible electrode array that allows for both neural recording of spike firings and low-threshold, high-precision stimulation for neuromodulation. Specifically, mouse turning behavior can be effectively induced with approximately five microamperes of stimulating current, which is significantly lower than that required by conventional rigid electrodes. The array's densely packed microelectrodes enable highly selective stimulation, allowing precise targeting of specific brain areas critical for turning behavior. This low-current, targeted stimulation approach helps maintain the health of both neurons and electrodes, as evidenced by stable neural recordings after extended stimulations. Systematic validations have confirmed the durability and biocompatibility of the electrodes. Moreover, we extended the flexible electrode array to a brain-to-brain interface system that allows human brain signals to directly control mouse behavior. Using advanced decoding methods, a single individual can issue eight commands to simultaneously control the behaviors of two mice. This study underscores the effectiveness of the flexible electrode array in neuromodulation, opening new avenues for interspecies communication and potential neuromodulation applications.
BibTeX
@inproceedings{ye2025highprecision,
  author = {Yifei Ye and Ye Tian and Haifeng Liu and Jiaxuan Liu and Cunkai Zhou and Cheng‐Jian Xu and Ting Zhou and Yanyan Nie and Yu Wu and Lunming Qin and Zhitao Zhou and Xiaoling Wei and Jianlong Zhao and Zhenyu Wang and Meng Li and Tiger H. Tao and Liuyang Sun},
  title = {{High‐Precision, Low‐Threshold Neuromodulation With Ultraflexible Electrode Arrays for Brain‐to‐Brain Interfaces}},
  booktitle = {Exploration},
  year = {2025},
  doi = {10.1002/exp.70040},
}
Figure