GRAVITY: the calibration unit

Nicolas Blind, F. Eisenhauer, M. Haug, S. Gillessen, M. Lippa, L. Burtscher, O. Hans, F. Haußmann, Stefan Huber, A. Janssen, S. Kellner, Y. Kok, Thomas Ott, O. Pfuhl, E. Sturm, Johannes Weber, E. Wieprecht, A. Amorim, W. Brandner, G. Perrin, K. Perraut, C. StraubmeierProceedings of SPIE, the International Society for Optical Engineering/Proceedings of SPIE 2014 · DOI · arXiv

Abstract

We present the design and characterisation of the Calibration Unit (CU), a new sub-system of the VLTI 2nd generation instrument GRAVITY. The CU provides all functions to test and calibrate the beam combiner instrument: it creates two artificial stars on four beams and includes four delay lines with internal metrology. It also features artificial stars for tip-tilt and pupil guiding, metrology pick-up diodes, reference targets for alignment, and safety shutters. We present the characterisation and validation results of these sub-units.

Introduction

The GRAVITY instrument at the Very Large Telescope Interferometer (VLTI) is designed for narrow-angle astrometry with an unprecedented precision of 10 micro-arcseconds. Achieving such sensitivity—comparable to observing a tennis ball on the Moon from Earth—requires meticulous alignment and calibration of all optical sub-systems. Because atmospheric turbulence and telescope availability preclude the use of real stars for initial laboratory testing and daily health checks, a dedicated "star simulator" is required. This paper describes the Calibration Unit (CU), which provides the controlled "ground truth" necessary to validate GRAVITY's performance.

System Design and Integration

The Calibration Unit is a complex optomechanical assembly mounted directly to the entrance of the GRAVITY cryostat. This placement ensures that the calibration light follows nearly the same optical path as starlight from the VLTI, minimizing systematic errors.

The Calibration Unit (CU) mounted on the GRAVITY beam combiner instrument cryostat.

The CU's structure is primarily composed of a stainless steel frame and support structure, housing aluminum opto-mechanics. It includes motorized linear stages for precise positioning and adjustable mirror mounts. To maintain high fringe contrast, the design prioritizes symmetry across the four beams to limit differential birefringence and ensure polarization stability.

Optical Injection and Simulation

The heart of the CU is its injection system. It utilizes a "V-groove" assembly containing three optical fibers: two simulating stars (separated by 1.5 arcseconds on the sky) and one providing a tip-tilt reference.

Schematic of the injection optics and launch telescope, illustrating the path from the fiber V-groove through the pellicle beamsplitter to the launch telescope.

As shown in the schematic, the light from these fibers is collimated by a parabolic launch telescope to create four beams. The CU also simulates realistic observing conditions using rotating phase screens to mimic atmospheric "seeing" and linear polarizers to test the instrument's response to different polarization states. Additionally, it includes 4x4 pupil tracking reference sources that mimic the light sources mounted on the telescope secondary mirror support structures ("spiders") at the Paranal Observatory.

Performance Characterization

Characterization at the Max Planck Institute for Extraterrestrial Physics (MPE) confirmed that the CU meets GRAVITY's stringent requirements:

  • Optical Quality: The system achieved a Strehl ratio of >95% in the K-band (2.0–2.5 µm) and >90% in the H-band.
  • Wavefront Precision: Total wavefront error was measured at approximately 34 nm RMS, well within the limits for high-fidelity simulation.
  • Polarization Stability: Differential retardance remained below 5°, limiting fringe contrast loss to less than 1%.
  • Mechanical Performance: Motorized delay lines showed linearity better than 0.1% over a 14 mm stroke.

Conclusion and Future Outlook

The Calibration Unit has proven to be an essential tool for the integration and testing of GRAVITY. While initial tests were successful, future improvements include replacing bare gold mirrors with protected gold versions to prevent degradation and re-optimizing the pupil tracker mask to resolve minor alignment discrepancies. The unit is now transitioning from laboratory testing to final deployment and alignment with the VLTI in Paranal, Chile.

Reference: Blind, N., et al. "GRAVITY: the calibration unit." Proceedings of SPIE, 2014. (arXiv:1407.6660v1)

Summary updated

Abstract

We present in this paper the design and characterisation of a new sub-system of the VLTI 2<sup><i>nd</i></sup> generation instrument GRAVITY: the Calibration Unit. The Calibration Unit provides all functions to test and calibrate the beam combiner instrument: it creates two artificial stars on four beams, and dispose of four delay lines with an internal metrology. It also includes artificial stars for the tip-tilt and pupil guiding systems, as well as four metrology pick-up diodes, for tests and calibration of the corresponding sub-systems. The calibration unit also hosts the reference targets to align GRAVITY to the VLTI, and the safety shutters to avoid the metrology light to propagate in the VLTI-lab. We present the results of the characterisation and validtion of these differrent sub-units.
BibTeX
@inproceedings{blind2014gravity,
  author = {Nicolas Blind and F. Eisenhauer and M. Haug and S. Gillessen and M. Lippa and L. Burtscher and O. Hans and F. Haußmann and Stefan Huber and A. Janssen and S. Kellner and Y. Kok and Thomas Ott and O. Pfuhl and E. Sturm and Johannes Weber and E. Wieprecht and A. Amorim and W. Brandner and G. Perrin and K. Perraut and C. Straubmeier},
  title = {{GRAVITY: the calibration unit}},
  booktitle = {Proceedings of SPIE, the International Society for Optical Engineering/Proceedings of SPIE},
  year = {2014},
  doi = {10.1117/12.2055542},
}
Figure