Fully Automatic Catheter Localization in C-Arm Images Using ℓ1-Sparse Coding
Lists
Only you see this
Abstract
We propose a method to perform automatic detection and tracking of electrophysiology (EP) catheters in C-arm fluoroscopy sequences. Our approach does not require any initialization, is completely automatic, and can concurrently track an arbitrary number of overlapping catheters. After a pre-processing step, we employ sparse coding to first detect candidate catheter tips, and subsequently detect and track the catheters. The proposed technique is validated on 2835 C-arm images, which include 39,690 manually selected ground-truth catheter electrodes. Results demonstrated sub-millimeter detection accuracy and real-time tracking performances.
BibTeX
@inproceedings{milletari2014fully,
author = {Fausto Milletari and Vasileios Belagiannis and Nassir Navab and Pascal Fallavollita},
title = {{Fully Automatic Catheter Localization in C-Arm Images Using ℓ1-Sparse Coding}},
booktitle = {International Conference on Medical Image Computing and Computer-Assisted Intervention},
year = {2014},
doi = {10.1007/978-3-319-10470-6\_71},
}Loading related content...