Estimate the Linear Effect of Inter-Scanner Variability: Insight from Paired Cross-Scanner T1-weighted Images
Abstract
To broaden our knowledge of inter-scanner variability, we established a pipeline to linearly estimate the site effect by incorporating ComBat modeling and superpixel parcellation. We found that the variation prominently manifested in tissue contrast, noise level, and field inhomogeneity with respect to cross-site data. We further used the estimated parameters to harmonize images. The image quality and structural similarity of cross-scanner data can be improved after the harmonization procedure, and the variation in volumetric measures can also be reduced. This study provides further insight for the research focusing on the development of image harmonization methods.
BibTeX
@inproceedings{chen2024estimate,
author = {Chang‐Le Chen and Mahbaneh Eshaghzadeh Torbati and Weiquan Luo and Davneet Minhas and Charles M. Laymon and Seong Jae Hwang and Ciprian Crainiceanu and Pauline Maillard and Evan Fletcher and Charles DeCarli and Howard Aizenstein and Dana Tudorascu},
title = {{Estimate the Linear Effect of Inter-Scanner Variability: Insight from Paired Cross-Scanner T1-weighted Images}},
booktitle = {Proceedings on CD-ROM - International Society for Magnetic Resonance in Medicine. Scientific Meeting and Exhibition/Proceedings of the International Society for Magnetic Resonance in Medicine, Scientific Meeting and Exhibition},
year = {2024},
doi = {10.58530/2023/4611},