Endoscopic measurement of nasal septum perforations

Jean-Claude Rosenthal, Eric L. Wisotzky, Carsten Matuschek, Melanie Hobl, Anna Hilsmann, Peter Eisert, Florian C. Uecker

Abstract

Based on the very good measurements, we outline a therapeutic procedure which should enable the production of patient-specific NSP implants based on endoscopic data only.

Introduction

A nasal septum perforation (NSP) is a defect in the wall dividing the nostrils, often resulting from trauma, surgery, or inflammation. These perforations cause significant symptoms, including restricted breathing, chronic nosebleeds, and sleep disturbances. While "septal buttons" are commonly used for treatment, standard sizes often fit poorly, leading to tissue necrosis. Creating custom-fit implants currently requires either invasive silicone impressions or radiation-intensive CT scans. This research presents a non-contact, radiation-free alternative using 3D stereo endoscopes and computer vision to measure NSPs with sub-millimeter precision.

Anatomical location of the nasal septum (a) and the corresponding endoscopic view of a perforation (b).

Methodology

The researchers developed a computer vision pipeline to transform 2D stereo video frames into metric 3D models. The system is first calibrated using a reference body to determine the endoscope's intrinsic parameters, such as focal length and stereo base.

The computer vision (CV) pipeline involves several key steps:

  1. Stereo Rectification: Aligning the left and right images so corresponding points lie on the same horizontal line.
  2. Disparity Estimation: Calculating the depth of every pixel by comparing the shift between the two views.
  3. Metric Survey: Converting pixel-depths into real-world millimeters using the calibration data.
3D point cloud generation and disparity estimation on the endoscopic image for metric measurement.

Results and Validation

The study involved a clinical evaluation of six patients, comparing endoscopic measurements against CT scans, the current "gold standard." The results demonstrated high precision, with a relative measurement deviation of less than 5%. The mean absolute difference between the endoscope and CT was only 0.28 mm for the horizontal axis and 0.16 mm for the vertical axis. Notably, the endoscopic method directly visualizes the mucosal edges, which can be difficult to define accurately on CT scans.

Discussion and Future Work

While highly accurate, the current method faces challenges such as a limited field of view for large perforations and environmental factors like lens fogging. Future work will focus on integrating 3D Mosaicking and SLAM (Simultaneous Localization and Mapping) to allow surgeons to stitch multiple images together into a global 3D model, effectively overcoming current field-of-view limitations.

Conclusion

The high precision of endoscopic 3D measurement enables a fully digital workflow for the production of patient-specific NSP implants, offering a gentler and more accurate alternative to traditional molding and imaging techniques.

Summary updated

Abstract

Based on the very good measurements, we outline a therapeutic procedure which should enable the production of patient-specific NSP implants based on endoscopic data only.
BibTeX
@inproceedings{rosenthal2021endoscopic,
  author = {Jean-Claude Rosenthal and Eric L. Wisotzky and Carsten Matuschek and Melanie Hobl and Anna Hilsmann and Peter Eisert and Florian C. Uecker},
  title = {{Endoscopic measurement of nasal septum perforations}},
  booktitle = {HNO},
  year = {2021},
  doi = {10.1007/s00106-021-01102-4},
}