Efficient Joint Estimation of Tracer Distribution and Background Signals in Magnetic Particle Imaging Using a Dictionary Approach

Tobias Knopp, Mirco Grosser, Matthias Gräser, Timo Gerkmann, Martin Möddel

Abstract

Background signals are a primary source of artifacts in magnetic particle imaging (MPI) and limit the sensitivity of the method since background signals are often not precisely known and vary over time. The state-of-the-art method for handling background signals uses one or several background calibration measurements with an empty scanner bore and subtracts a linear combination of these background measurements from the actual particle measurement. This approach yields satisfying results in case that the background measurements are taken in close proximity to the particle measurement and when the background signal drifts linearly. In this work, we propose a joint estimation of particle distribution and background signal based on a dictionary that is capable of representing typical background signals and allows for precise estimation of the background even when the latter is drifting non-linearly over time. Using a singular-value decomposition, the dictionary is derived from a large number of background calibration scans that do not need to be recorded in close proximity to the particle measurement. The dictionary is sufficiently expressive and represented by its principle components. The proposed joint estimation of particle distribution and background signal is expressed as a linear Tikhonov-regularized least squares problem, which can be efficiently solved. In phantom experiments it is shown that the method strongly suppresses background artifacts and even allows to estimate and remove the direct feed-through of the excitation field.

Introduction

Magnetic Particle Imaging (MPI) is a high-contrast tomographic technique that tracks super-paramagnetic iron-oxide (SPIO) nanoparticles. Despite its potential, MPI sensitivity is frequently compromised by background signals—interference stemming from thermal drifts in the scanner hardware or direct feed-through from the excitation field.

The conventional "gold standard" for noise mitigation involves subtracting a "blank" measurement (an empty scanner scan) from the particle data. While effective for static or linearly drifting backgrounds, this method fails during long experiments or in clinical settings where noise drifts non-linearly. Previous solutions often required complex multi-patch sequences that sacrificed temporal resolution or utilized non-linear solvers that were computationally prohibitive. This paper introduces a dictionary-based joint estimation framework designed to be both computationally efficient and compatible with standard MPI hardware.

Methods: The Dictionary Approach

The core innovation lies in treating background subtraction not as a pre-processing step, but as part of a joint estimation framework. The goal is to solve for the tracer distribution ($c$) and the background signal ($b$) simultaneously.

1. Dictionary Construction

The researchers compiled a library of background scans from an empty scanner bore. By applying Singular Value Decomposition (SVD) to this dataset, they extracted principal components that define a mathematical subspace representing typical background noise. This "dictionary" ($\Phi$) allows the model to represent complex, non-linear noise patterns.

2. Mathematical Formulation

The reconstruction is formulated as a Tikhonov-regularized least squares problem, where the algorithm minimizes the difference between measured data and the sum of the predicted tracer signal plus the dictionary-based background estimate. By maintaining a linear structure, the problem can be solved using the iterative Kaczmarz method, the standard high-speed solver in MPI. This ensures that the joint estimation does not significantly increase reconstruction time, maintaining the feasibility of real-time imaging.

Experimental Setup

The method was validated using experimental data from a custom human-sized brain MPI scanner. The study utilized two primary phantoms:

  • Static Dot Phantom: A 31.25 µg iron sample measured over 140 frames (65 seconds) to evaluate long-term background drift suppression.
  • Dynamic Bolus Phantom: A perfusion simulation using tubes where an 850 µg iron bolus is circulated and measured over 490 frames (247 seconds).
Dynamic bolus phantom setup used for perfusion simulation.

Key Results and Insights

The dictionary approach demonstrated several significant advantages over traditional subtraction methods:

  • Superior Artifact Suppression: In static phantom tests, the method almost entirely eliminated background artifacts that typically emerge as scanner hardware drifts over time.
  • Quantitative Stability: Unlike standard methods, which often show a systematic drift in estimated iron content as noise is mistaken for tracer, the joint estimation provided a stable and accurate estimate of iron mass across long time series.
  • Excitation Frequency Recovery: Perhaps most notably, the method effectively suppressed the massive direct feed-through of the excitation field. This suggests that the excitation frequency might no longer need to be filtered out, potentially recovering valuable low-frequency tracer data previously thought lost to noise.
  • Computational Viability: The algorithmic complexity remains nearly as fast as standard reconstruction, making it suitable for clinical real-time applications.

Conclusion and Future Directions

The proposed dictionary approach offers a robust solution to the persistent problem of background noise in MPI. By leveraging the separability of tracer signals and background interference in mathematical subspaces, it enhances image quality without complicating scanning protocols. Future work will focus on hyperparameter optimization and investigating the generalizability of these dictionaries across different imaging sequences.

Summary updated

Abstract

Background signals are a primary source of artifacts in magnetic particle imaging and limit the sensitivity of the method since background signals are often not precisely known and vary over time. The state-of-the art method for handling background signals uses one or several background calibration measurements with an empty scanner bore and subtracts a linear combination of these background measurements from the actual particle measurement. This approach yields satisfying results in case that the background measurements are taken in close proximity to the particle measurement and when the background signal drifts linearly. In this work, we propose a joint estimation of particle distribution and background signal based on a dictionary that is capable of representing typical background signals and allows for precise estimation of the background even when the latter is drifting non-linearly over time. Using a singular-value decomposition, the dictionary is derived from a large number of background calibration scans that do not need to be recorded in close proximity to the particle measurement. The dictionary is sufficiently expressive and represented by its principle components. The proposed joint estimation of particle distribution and background signal is expressed as a linear Tikhonov-regularized least squares problem, which can be efficiently solved. In phantom experiments it is shown that the method strongly suppresses background artifacts and even allows to estimate and remove the direct feed-through of the excitation field.
BibTeX
@article{knopp2021efficient,
  author = {Tobias Knopp and Mirco Grosser and Matthias Gräser and Timo Gerkmann and Martin Möddel},
  title = {{Efficient Joint Estimation of Tracer Distribution and Background Signals in Magnetic Particle Imaging Using a Dictionary Approach}},
  journal = {IEEE Trans. Medical Imaging},
  year = {2021},
  url = {https://openreview.net/forum?id=6zPSkY_SQh},
}