Abstract
Transport of neonatal patients between critical care units can expose patients to whole-body vibrations which may pose a risk to the vulnerable patients' health. This study characterizes and mitigates vibrations transmitted by the Neonatal Patient Transport System (NPTS) used in Ontario. A 3-degree-of-freedom planar model in the roll plane was developed using the Newton-Euler method. Simulation results for vertical motion closely match road test data, while the roll direction requires further calibration. This model serves as a first step toward a full 3D simulation for optimizing patient safety during transport.
Introduction
Transporting critically ill neonates is a high-stakes operation where physical vibrations from the ambulance floor pose a clinical risk, including potential intracranial hemorrhage. To mitigate these risks, researchers are developing high-fidelity dynamic simulations to understand how vibrations propagate through the stretcher and incubator to the infant. Building such a model requires identifying unknown system parameters like stiffness and damping, which is best achieved through simplified planar models before moving to complex 3D simulations.
Methodology
The researchers developed a 3-Degree-of-Freedom (3-DOF) planar roll model to represent the NPTS. This model simplifies the system into a rigid body (stretcher and incubator) and a point mass (the patient) moving in the roll plane.
Key technical aspects include:
- Newton-Euler Formulation: Equations of motion were derived by balancing forces and moments and cross-validated against Lagrange’s equations.
- Kinematic Modeling: The model tracks vertical displacement ($q_1$), angular roll ($q_2$), and the relative displacement of the patient on the mattress ($q_3$).
- Parameter Tuning: The simulation was tuned using real-world vibration data from the Ottawa Paramedic Service (OPS), recorded during a high-speed arterial road transport using a Stryker Power-PRO stretcher and a 2.5 kg manikin.
Results and Discussion
The study successfully identified critical vibration modes and established baseline mechanical parameters.
- Vertical Accuracy: The simulated vertical acceleration Power Spectral Density (PSD) closely matched road test data, particularly capturing a significant vibration mode around 9 Hz.
- Natural Frequencies: Identified system natural frequencies were 4.11 Hz, 8.33 Hz, and 9.26 Hz.
- Tuned Parameters: Baseline values were established for stretcher stiffness ($K = 330,000$ N/m) and mattress stiffness ($k = 8,000$ N/m).
- Roll Deviation: While vertical motion was accurate, the roll (angular) response deviated from road data, showing a high-magnitude peak at a lower frequency than expected. This suggests the "sliding pin joint" model for the floor-to-stretcher interface requires further refinement.
Conclusion and Future Work
While the vertical model is highly accurate, the roll plane discrepancy indicates a need for more complex modeling of the stretcher's mechanical interface. Future work will focus on refining angular parameters, integrating pitch models, and eventually scaling to a full 7-degree-of-freedom 3D model to simulate multi-axis transport conditions in both ground and air ambulances.