DopUS-Net: Quality-Aware Robotic Ultrasound Imaging based on Doppler Signal

Zhongliang Jiang, Felix Duelmer, Nassir Navab

Introduction

Peripheral Arterial Disease (PAD) affects approximately 20% of individuals over the age of 55. Despite its prevalence, it remains under-diagnosed due to the operator-dependent nature of manual ultrasound screenings. Robotic ultrasound systems (RUSS) offer a path toward standardization, but segmenting small, deep-seated vessels like the radial or ulnar arteries is challenging. Standard B-mode ultrasound is often obscured by speckle noise and artifacts. DopUS-Net addresses this by integrating Doppler flow signals into a quality-aware robotic control loop.

Methods

DopUS-Net utilizes a dual-encoder deep learning architecture to fuse anatomical (B-mode) and functional (Doppler) information.

System overview showing (a) robotic setup, (b-c) input ultrasound modalities, and (d-e) 3D vessel reconstruction.
  1. Dual-Encoder Architecture: One encoder processes duplex images (B-mode + Doppler) for spatial context, while a second encoder focuses solely on Doppler images to act as a region proposal module.
  2. Temporal Memory: Convolutional Gated Recurrent Units (convGRU) are employed at the bottleneck to maintain temporal consistency across frames, ensuring smooth segmentation of tubular structures.
  3. Quality-Aware Robotic Loop: The system monitors Doppler signal stability in real-time. If the signal degrades, the robot initiates an Artery Re-Identification procedure, automatically tilting the probe (±10°) to recover the optimal signal.
  4. Hardware: The framework is implemented on a KUKA LBR iiwa 7-DoF robot arm using impedance control for safe patient contact.

Key Insights

  • Doppler as a Prior: Doppler signals provide high-confidence "hints" about vessel location, allowing the network to ignore B-mode artifacts.
  • Active Sensing: Unlike passive models, DopUS-Net recognizes poor input quality and commands the robot to adjust the probe pose.
  • Tubular Consistency: The use of RNNs (convGRUs) prevents segmentation "flickering" by leveraging the continuous nature of blood vessels.

Key Results

The integration of Doppler signals and robotic feedback significantly improved performance:

  • Accuracy: The Dice score increased from 0.44 (standard U-Net) to 0.86 with the full DopUS-Net system.
  • Robustness: Intersection over Union (IoU) improved from 0.47 to 0.78.
  • Real-time Performance: Inference takes only 20ms, enabling real-time control.
  • Robotic Success: The re-identification module successfully recovered lost signals in 86% to 92% of physical tests.
3D reconstruction of the vessel path (e) demonstrating the continuity achieved by the system.

Limitations and Future Work

While promising, the system currently lacks compensation for large patient movements and is limited to one-dimensional probe "fanning" during re-identification. Future work will explore 3D search patterns and testing on clinical populations with PAD, where calcifications and irregular flow patterns present additional challenges.

Reference: Jiang, Z., Duelmer, F., & Navab, N. (2023). DopUS-Net: Quality-Aware Robotic Ultrasound Imaging based on Doppler Signal. arXiv:2305.08938v1.

Summary updated

BibTeX
@article{jiang2024dopus,
  author = {Zhongliang Jiang and Felix Duelmer and Nassir Navab},
  title = {{DopUS-Net: Quality-Aware Robotic Ultrasound Imaging based on Doppler Signal}},
  journal = {IEEE Trans Autom. Sci. Eng.},
  year = {2024},
  doi = {10.1109/tase.2023.3277331},
}