Direct Estimation of Pharmacokinetic Parameters from DCE-MRI Using Deep CNN with Forward Physical Model Loss

Cagdas Ulas, Giles Tetteh, Michael J. Thrippleton, Paul A. Armitage, Stephen D. Makin, Joanna M. Wardlaw, Mike E. Davies, Bjoern H. Menze

Abstract

Dynamic Contrast-Enhanced (DCE) MRI is essential for estimating pharmacokinetic (PK) parameters like vascular permeability, which are critical for diagnosing various pathologies. However, the trade-off between spatial and temporal resolution often requires significant undersampling, making parameter estimation computationally expensive and slow. This paper introduces a deep learning framework designed to directly estimate PK parameters from highly undersampled DCE-MRI data. By incorporating a forward physical model into the network's loss function and employing residual learning, the proposed Convolutional Neural Network (CNN) achieves a massive 4,000x speedup compared to traditional iterative model-based reconstruction methods while maintaining or improving accuracy.

Introduction

Dynamic Contrast-Enhanced (DCE) MRI involves the rapid acquisition of MRI scans following the injection of a contrast agent. Analyzing the resulting image-time series allows clinicians to estimate pharmacokinetic (PK) parameters, such as $K^{trans}$ (vascular permeability) and $v_p$ (plasma volume fraction). These metrics are vital for managing brain tumors, strokes, and multiple sclerosis.

The primary challenge in DCE-MRI is the severe trade-off between spatial resolution, temporal resolution, and volume coverage. To capture rapid contrast dynamics, scanners must undersample k-space data. Reconstructing accurate PK parameters from this incomplete data typically relies on "indirect" methods (reconstruction followed by fitting) or "direct" model-based iterative methods. While direct methods are more accurate, they involve solving massive optimization problems that can take hours per patient, limiting their clinical utility.

Methods

The authors propose a 3D CNN architecture that maps undersampled, corrupted DCE-MRI time-series directly to PK parameter maps. The framework incorporates three key innovations:

  1. Residual Learning: The network is trained to predict "residual maps"—the difference between the true parameters and the corrupted parameters derived from undersampled data. Residual maps are typically sparser and topologically simpler, facilitating easier learning.
  2. Forward Physical Model Loss: This core contribution embeds the physics of contrast agent kinetics into the training process. The loss function penalizes the discrepancy between the measured MRI signal and a simulated signal generated by passing the network's predicted PK parameters through the Patlak tracer kinetic model and MRI signal equations.
  3. Dual-Pathway Architecture: The CNN utilizes parallel pathways—a local pathway for fine details and a global pathway using dilated convolutions to capture broader spatial context.
Conceptual framework of the proposed direct estimation method. The forward physical model is integrated into the loss function to ensure the network respects the underlying kinetics.

The model was trained and validated using in vivo brain DCE-MRI data from patients with mild ischaemic stroke, retrospectively undersampled at a 10-fold rate.

Results

The proposed CNN demonstrated significant advantages in both speed and quality. Inference for a full 3D volume takes approximately 1.5 seconds, representing a $\approx 4 \times 10^3$ speedup over the 95 minutes required by state-of-the-art iterative methods.

Quantitatively, the CNN achieved high fidelity with an average Concordance Correlation Coefficient (CCC) of 0.88 and a Structural Similarity Index (SSIM) of 0.92 when balancing physical and parameter loss ($\lambda = 0.5$). Visually, the CNN effectively suppressed Gibbs artifacts that were present even in the fully-sampled "ground truth" data, leading to cleaner parameter maps than traditional methods.

Qualitative comparison of $K^{trans}$ and $v_p$ maps. The proposed CNN ($\lambda=0.5$) provides superior visual quality and higher SSIM scores compared to the traditional model-based approach.

Discussion and Conclusion

By combining data-driven deep learning with physics-based constraints, this framework enables near-instantaneous estimation of high-quality PK parameters from undersampled DCE-MRI. While the study is limited by a small patient cohort (n=3), the results suggest a powerful path forward for rapid clinical workflows. Future work will involve validating the model on larger, more diverse datasets and extending the framework to more complex tracer kinetic models.

Summary updated

BibTeX
@inproceedings{ulas2018direct,
  author = {Cagdas Ulas and Giles Tetteh and Michael J. Thrippleton and Paul A. Armitage and Stephen D. Makin and Joanna M. Wardlaw and Mike E. Davies and Bjoern H. Menze},
  title = {{Direct Estimation of Pharmacokinetic Parameters from DCE-MRI Using Deep CNN with Forward Physical Model Loss}},
  booktitle = {International Conference on Medical Image Computing and Computer-Assisted Intervention},
  year = {2018},
  doi = {10.1007/978-3-030-00928-1\_5},
}
Figure