Direct Calibration of a Laser Ablation System in the Projective Voltage Space

Adrian Schneider, Simon Pezold, Kyung-won Baek, Dilyan Marinov, Philippe C. Cattin

Abstract

Laser ablation is a widely adopted technique in many contemporary medical applications. However, it is new to use a laser to cut bone and perform general osteotomy surgical tasks with it. In this paper, we propose to apply the direct linear transformation algorithm to calibrate and integrate a laser deflecting tilting mirror into the affine transformation chain of a sophisticated surgical navigation system, involving next generation robots and optical tracking. Experiments were performed on synthetic input and real data. The evaluation showed a target registration error of 0.3mm ± 0.2mm in a working distance of 150 mm.
BibTeX
@inproceedings{schneider2015direct,
  author = {Adrian Schneider and Simon Pezold and Kyung-won Baek and Dilyan Marinov and Philippe C. Cattin},
  title = {{Direct Calibration of a Laser Ablation System in the Projective Voltage Space}},
  booktitle = {International Conference on Medical Image Computing and Computer-Assisted Intervention},
  year = {2015},
  doi = {10.1007/978-3-319-24553-9\_34},
}

Loading related content...

Figure