Development of an in-situ apparatus for synchrotron-based surface scattering measurements at mineral–water interfaces in elevated pressure and temperature conditions

Sang Soo Lee, Changyong Park, Jacquelyn N. Bracco, Peng Yang, Andrew G. Stack, Paul Fenter2023 · DOI

Abstract

Mineral-water interfaces are primary sites for various chemical reactions in nature. Many of these reactions occur in subsurface environments where elevated pressure and temperature can alter both the chemistry of solutes and the stability of mineral surfaces, which in turn influence interfacial reactivity. Here, we report a new in-situ apparatus that is designed for synchrotron-based surface scattering measurements in these nonambient conditions. This custom-built cell, made of an anti-corrosive Ni-based alloy (Alloy 625), has a miniaturized reaction chamber that contains a small volume (~10 μL) of a fluid in contact with a small active area (~5 mm²) of a mineral surface. This interfacial environment is encapsulated with a pair of diamond discs as X-ray windows, which are stable under operation pressures up to ~40 MPa (~6,000 psi) and temperatures up to ~400 °C. The cell performance is validated by collecting a series of in-situ high-resolution X-ray reflectivity (XR) at the barite (001)-water interface as a function of temperature from 25 °C to 200 °C. Challenges in the measurement and data analysis are also discussed based on the comparison with those using other XR cell setups. Overall, the development provides us a new opportunity for the research on this currently unexplored area of geochemistry at mineral-water interfaces in extreme conditions.

Introduction

Many of the most significant chemical reactions on Earth occur where water meets rock deep underground. These subsurface environments are characterized by intense pressure and high temperatures, conditions that fundamentally alter the behavior of solutes and the stability of mineral surfaces. Understanding these "interfacial reactivities" is essential for critical applications such as carbon sequestration, nuclear waste storage, and the study of natural ore deposits.

Despite their importance, studying these interfaces at the atomic scale is exceptionally difficult. High-resolution tools like synchrotron-based X-ray reflectivity (XR) are typically optimized for ambient laboratory conditions. Engineering a device that can withstand the corrosive, high-pressure environment of the deep crust while remaining transparent to X-rays represents a significant technical hurdle. This paper introduces a specialized in-situ apparatus designed to bridge this gap, allowing for real-time observation of mineral-water interfaces under conditions mimicking the Earth's upper crust.

Design and Methodology

The researchers developed a custom-built, high-pressure, high-temperature (HPHT) reaction cell specifically for surface scattering measurements at the Advanced Photon Source (APS). The design focuses on three key engineering pillars:

1. Material Selection

The cell is constructed from Alloy 625, a nickel-based superalloy. This material was chosen for its exceptional mechanical strength and its resistance to corrosion, which is vital when the cell is filled with chemically active fluids at high temperatures.

2. Miniaturization and Safety

To manage the inherent risks of high-pressure operations, the reaction chamber was miniaturized. It holds a tiny fluid volume of approximately 10 μL against a mineral surface area of roughly 5 mm². This small volume minimizes the potential energy release in the event of a structural failure while ensuring the chemical environment remains highly responsive to temperature adjustments.

3. X-ray Transparency

The "windows" of the device are comprised of a pair of diamond discs. Diamond was selected because it possesses the structural integrity to withstand pressures up to 40 MPa (6,000 psi) while remaining nearly transparent to the high-energy X-rays required for high-resolution reflectivity measurements.

Experimental Validation

To validate the performance of the apparatus, the team conducted high-resolution XR measurements on a barite (001) crystal interface. The experiment involved tracking structural changes as the temperature was increased from 25 °C to 200 °C. This test served to demonstrate that the signal-to-noise ratio achieved through the diamond windows was sufficient to resolve atomic-scale surface features even under non-ambient conditions.

Key Results and Discussion

Operational Envelope

The device successfully demonstrated stable operation at pressures up to 40 MPa and temperatures up to 400 °C. This range covers a significant portion of the conditions found in the Earth's upper crust, making it a versatile tool for geophysical research.

High-Resolution Data Acquisition

The successful capture of the barite-water interface structure at 200 °C confirms that the apparatus can provide high-quality data. This allows researchers to move beyond "snapshots" taken after an experiment and instead view a "movie" of geochemical processes as they evolve in real-time.

Analytical Challenges

While the cell is a breakthrough, it introduces unique challenges in data analysis. The presence of diamond windows and the high-pressure casing introduces background noise and geometric constraints that are not present in standard ambient XR cells. The authors emphasize that specific data analysis strategies are required to accurately extract surface structures from the data collected in this environment.

Conclusion and Future Directions

The development of this HPHT reaction cell opens a new frontier in subsurface geochemistry. By integrating extreme environment engineering with high-precision X-ray optics, researchers can now explore mineral-water interfaces in conditions that were previously inaccessible.

Future work will focus on refining the software and models used to account for the systematic errors and noise introduced by the HPHT hardware. Additionally, the researchers plan to expand the scope of study to include a wider variety of minerals and complex fluids, such as brines or CO₂-saturated water, to further our understanding of the deep Earth's chemical landscape.

Summary updated

Abstract

Mineral-water interfaces are primary sites for various chemical reactions in nature.Many of these reactions occur in subsurface environments where elevated pressure and temperature can alter both the chemistry of solutes and the stability of mineral surfaces, which in turn influence interfacial reactivity.Here, we report a new in-situ apparatus that is designed for synchrotron-based surface scattering measurements in these nonambient conditions.This custom-built cell, made of an anti-corrosive Ni-based alloy (Alloy 625), has a miniaturized reaction chamber that contains a small volume (~10 L) of a fluid in contact with a small active area (~5 mm 2 ) of a mineral surface.This interfacial environment is encapsulated with a pair of diamond discs as X-ray windows, which are stable under operation pressures up to ~40 MPa (= ~6,000 psi) and temperatures up to ~400 C.The cell performance is validated by collecting a series of in-situ high-resolution X-ray reflectivity (XR) at the barite (001)-water interface as a function of temperature from 25 C to 200 C.Challenges in the measurement and data analysis are also discussed based on the comparison with those using other XR cell setups.Overall, the development provides us a new opportunity for the research on this currently unexplored area of geochemistry at mineral-water interfaces in extreme conditions.
BibTeX
@misc{lee2023development,
  author = {Sang Soo Lee and Changyong Park and Jacquelyn N. Bracco and Peng Yang and Andrew G. Stack and Paul Fenter},
  title = {{Development of an in-situ apparatus for synchrotron-based surface scattering measurements at mineral–water interfaces in elevated pressure and temperature conditions}},
  year = {2023},
  doi = {10.7185/gold2023.18735},
}
Figure