Detection of Lung Cancer Cells in Solutions Using a Terahertz Chemical Microscope

Yuichi Yoshida, Xue Ding, Kohei Iwatsuki, Katsuya Taniizumi, Hirofumi Inoue, Jin Wang, Kenji Sakai, Toshihiko KiwaSensors 2021 · OpenReview

TL;DR

This paper introduces a terahertz chemical microscope (TCM) capable of rapidly and quantitatively detecting lung cancer cells in solution. By leveraging an antibody-functionalized sensing plate and terahertz wave emission, the TCM offers a faster, less skill-dependent alternative to traditional pathological methods, paving the way for improved liquid biopsy diagnostics.

What Problem Does This Paper Address?

Accurate cancer diagnosis and personalized treatment rely heavily on understanding the proportion of cancer cells within a patient's sample. Current methods, like analyzing fixed and stained tissue slices, are time-consuming (taking at least two days) and require significant expertise from pathologists. This bottleneck hinders rapid analysis and personalized treatment strategies. The paper aims to address this by developing a faster, more automated method for quantifying cancer cells directly in solution, eliminating the need for complex sample preparation.

Methods

The core of this research is a Terahertz Chemical Microscope (TCM). The system utilizes a specialized sensing plate made of silicon oxide on a sapphire substrate. When a femtosecond laser pulse is directed at the sensing plate, it generates terahertz (THz) waves. The key innovation lies in functionalizing this sensing plate with antibodies specific to lung cancer cells.

Here's a breakdown of the methodology:

  • Sensing Plate Functionalization: Avidin is immobilized on the silicon oxide surface of the sensing plate via an amine coupling method. This is followed by a biotin-avidin reaction to attach biotin-labeled cytokeratin AE1/AE3 antibodies, which specifically bind to lung cancer cells. This avidin-biotin strategy aims for high-density antibody immobilization and improved alignment.
  • Cancer Cell Detection: A solution containing lung cancer cells (PC9) is introduced to the sensing plate. The cancer cells bind to the immobilized antibodies.
  • Terahertz Emission: The interaction between the cancer cells and the antibodies on the sensing plate alters the electric field within the silicon layer. This change in electric field directly influences the terahertz wave emitted from the sensing plate. The amplitude of this emitted terahertz wave is measured.
  • pH Normalization: A significant challenge identified is the variability in sensing plate sensitivity, which can be influenced by factors like pH. To account for this, the paper proposes a pH normalization step. The sensing plate's response to different pH levels (1.68 and 10.01) is measured and used to calibrate the terahertz signal. This ensures that changes in terahertz amplitude are primarily due to cancer cell presence, not environmental pH fluctuations.
  • Data Analysis: The terahertz amplitude change is calculated by subtracting the background signal (before cell reaction) from the signal after the reaction. This change is then plotted against cell concentration. The coefficient of variation (CV) is used to quantify the variability in measurements, both with and without pH correction.

Datasets

The study uses cultivated human lung adenocarcinoma cells (PC9) at concentrations of 10³, 10⁴, and 10⁵ cells/mL. These are not pre-existing datasets but rather experimental samples prepared by the researchers for evaluation.

Key Insights and Ideas

  • Terahertz Waves as a Sensitive Probe: The research demonstrates that terahertz waves can be effectively used to detect subtle changes in the electrical properties of a sensing surface caused by biological interactions, specifically the binding of cancer cells to antibodies.
  • Avidin-Biotin for Enhanced Immobilization: The use of the avidin-biotin system is crucial for achieving high-density and well-aligned antibody immobilization, which is essential for sensitive detection.
  • pH Variability is a Critical Factor: The paper highlights that the sensitivity of the sensing plate can be significantly affected by pH. This is a critical insight for any biosensing application relying on electrical or electromagnetic signal transduction.
  • pH Normalization Improves Accuracy: The proposed pH correction method significantly reduces the variability in measurements, leading to more reliable cancer cell detection. This suggests that such normalization strategies are vital for robust biosensing platforms.
  • Potential for Liquid Biopsies: The TCM's ability to detect cancer cells directly in solution without extensive sample preparation makes it a promising candidate for non-invasive liquid biopsies, which could revolutionize cancer diagnosis and monitoring.

Key Results

  • The TCM successfully detected lung cancer cells in solution, with the terahertz amplitude increasing proportionally to the cell concentration.
  • The measured changes in terahertz amplitude for PC9 concentrations of 10³, 10⁴, and 10⁵ cell/mL were approximately 0.047 ± 0.018 mV, 0.54 ± 0.22 mV, and 0.77 ± 0.38 mV, respectively.
  • Without pH correction, the coefficient of variation (CV) for cancer cell detection ranged from 36.9% to 49.2%.
  • With pH compensation, the CV was significantly reduced to between 19.1% and 34.7%, indicating a substantial improvement in measurement accuracy and reliability. The variation was reduced by a factor of 1.4 to 1.9 depending on the cell concentration.

Limitations and Future Work

Limitations:

  • Sensing Plate Variability: The paper acknowledges that variations in the properties of individual sensing plates can lead to differences in sensitivity, necessitating calibration.
  • Non-uniform Reactions: The terahertz distribution within each well showed some irregularity, attributed to non-uniform antibody immobilization or cell binding.

Future Work:

  • The authors suggest that the TCM is expected to be an excellent candidate for liquid biopsies in cancer diagnoses, implying further development and validation for clinical applications.
  • While not explicitly stated as future work, addressing the sensing plate variability and improving reaction uniformity would be logical next steps.

Summary updated

Abstract

Cancer genome analysis has recently attracted attention for personalized cancer treatment. In this treatment, evaluation of the ratio of cancer cells in a specimen tissue is essential for the precise analysis of the genome. Conventionally, the evaluation takes at least two days and depends on the skill of the pathologist. In our group, a terahertz chemical microscope (TCM) was developed to easily and quickly measure the number of cancer cells in a solution. In this study, an antibody was immobilized on a sensing plate using an avidin-biotin reaction to immobilize it for high density and to improve antibody alignment. In addition, as the detected terahertz signals vary depending on the sensitivity of the sensing plate, the sensitivity was evaluated using pH measurement. The result of the cancer cell detection was corrected using the result of pH measurement. These results indicate that a TCM is expected to be an excellent candidate for liquid biopsies in cancer diagnosis.
BibTeX
@inproceedings{yoshida2021detection,
  author = {Yuichi Yoshida and Xue Ding and Kohei Iwatsuki and Katsuya Taniizumi and Hirofumi Inoue and Jin Wang and Kenji Sakai and Toshihiko Kiwa},
  title = {{Detection of Lung Cancer Cells in Solutions Using a Terahertz Chemical Microscope}},
  booktitle = {Sensors},
  year = {2021},
  url = {https://openreview.net/forum?id=odtINZgUNl},
}
Figure