Design and Analysis of a Metamaterial-Inspired Absorber for data rate in 52% RF-to-DC conversion Efficiency Dual-band SWIPT system

Z. W. Dong, Xin Jiang, Adel Barakat, Ramesh K. PokharelarXiv 2024 · DOI · arXiv

TL;DR

Researchers have developed a metamaterial-inspired absorber that significantly boosts the performance of Simultaneous Wireless Information and Power Transfer (SWIPT) systems operating through biological tissue. By using split-ring resonators to achieve negative permeability, the design improves channel isolation by 5dB, increases data rates by 7 MB/s, and maintains a high RF-to-DC conversion efficiency of 52%.

What Problem Does This Paper Address?

Wireless medical implants and wearables need two things: power and data. SWIPT systems aim to provide both simultaneously. However, a major hurdle in compact, multi-port SWIPT designs is interference (or poor isolation) between the power channel (usually a lower frequency) and the data channel (a higher frequency).

When these channels are close together in a small device, the magnetic fields leak into one another. This leakage forces the system to use higher transmit power to overcome noise, which is particularly problematic when sending signals through biological tissue, where safety limits (like Specific Absorption Rate, or SAR) and signal attenuation are critical concerns.

Methods

The authors propose a 4-port SWIPT system that incorporates a novel metamaterial-inspired absorber at the transmitter (Tx) side.

  • The Absorber: The team etched two split-ring resonators (SRRs) between the ports. Using Composite Right/Left-Handed Transmission Line (CRLH-TL) theory, they designed these SRRs to exhibit a real permeability ($\mu$) close to -1 and an imaginary part near 0 at the 40 MHz operating frequency.
  • Reverse Refraction: At $\mu \approx -1$, the material induces "reverse refraction" of the magnetic field. This effectively traps or redirects the field, preventing it from leaking from the signal port to the power port.
  • Rectifier Design: On the receiver (Rx) side, they implemented a voltage doubler rectifier (using HSMS-2860 diodes) specifically tuned via conjugate impedance matching to maximize power conversion at 13.56 MHz.
  • Testing Environment: To simulate real-world conditions, the researchers used a 5mm thick human body equivalent "phantom" between the Tx and Rx units.

Datasets

As this is an experimental hardware paper, standard software datasets were not used. Instead, the authors generated original experimental data through:

  • Electromagnetic Simulations: Using Ansys HFSS to model H-field distributions and S-parameters.
  • Physical Measurements: Using a PNA analyzer for S-parameters, a vector signal generator (VSG) for data transmission, and a spectrum analyzer to measure Bit Error Rate (BER) and data rates.

Key Insights and Ideas

  • Negative Permeability as an Isolator: The "aha moment" is using the unique properties of metamaterials—specifically negative permeability—to solve a near-field interference problem. By making the absorber act like an open circuit at specific frequencies, they physically block unwanted signal propagation without needing bulky shielding.
  • Efficiency vs. Isolation Trade-off: Usually, adding components to improve isolation might degrade Power Transfer Efficiency (PTE). However, because this absorber reduces magnetic leakage, it actually helps maintain a high PTE (52%) by ensuring more energy reaches the intended power port.
  • Power Sensitivity: The system demonstrated stable communication at much lower power levels (-23dBm to -29dBm), representing a 6dBm improvement over conventional designs. This is vital for implantable devices where minimizing heat and power consumption is a priority.

Key Results

  • Isolation Boost: The proposed absorber improved isolation between ports by approximately 5dB at a 9mm transfer distance.
  • Data Rate Increase: The system achieved a substantial data rate jump from 13.5 MB/s to 20.5 MB/s (a 7 MB/s improvement).
  • Conversion Efficiency: The rectifier achieved an RF-to-DC conversion efficiency of 52% at the 9mm distance.
  • Signal Integrity: The system could transmit signals with 5dBm weaker power than a standard SWIPT system while maintaining the same Bit Error Rate.

Limitations and Future Work

  • Distance Sensitivity: The authors noted that when the distance between Tx and Rx exceeds 12mm, the PTE drops by about 2% due to the low Q-factor of the split-ring resonators. Future work could focus on high-Q metamaterials to extend the effective range.
  • Tissue Variability: While a 5mm phantom was used, real human anatomy is heterogeneous (fat, muscle, bone). Testing the system's robustness across different tissue types and thicknesses would be a logical next step.
  • Miniaturization: While the current design is compact, further integration into a fully biocompatible, flexible package for long-term implantation remains a challenge for future engineering.

Summary updated

Abstract

This paper proposes a novel metamaterial-inspired absorber designed to enhance the data rate in 52% RF to DC conversion simultaneous wireless information and power transfer system (SWIPT) through biological tissue. The proposed absorber includes split-ring resonators(SRRs) and demonstrates significant permeability characteristics, with both the real and imaginary parts being negative and close to -1. It also improves isolation by around 5dB in a WPT distance of 9mm. A 5mm thick phantom is used for biological tissue in this study. Experimental results exhibits that the SWIPT systems including a rectifier that converts 52% RF to DC efficiency in a WPT distance of 9mm embedding this absorber between power and signal ports at Tx side results in a 5dB improvement in isolation performance. By using proposed absorber, it enables a 7MB/s improvement of data rate and allows signals to be transmitted with 5dBm weaker power than without absorber SWIPT system.
BibTeX
@misc{dong2024design,
  author = {Z. W. Dong and Xin Jiang and Adel Barakat and Ramesh K. Pokharel},
  title = {{Design and Analysis of a Metamaterial-Inspired Absorber for data rate in 52\% RF-to-DC conversion Efficiency Dual-band SWIPT system}},
  year = {2024},
  eprint = {2410.12431},
  archivePrefix = {arXiv},
  url = {https://arxiv.org/abs/2410.12431},
}
Figure