Computational rheometry of yielding and viscoplastic flow in vane-and-cup rheometer fixtures

Emad Chaparian, Crystal E. Owens, Gareth H. McKinley

Abstract: A planar two-dimensional computational analysis is presented to qualify traditional and fractal vane-in-cup geometries for accurate rheometry of simple viscoplastic fluids with and without slip. Numerical simulations based on an adaptive augmented Lagrangian scheme are used to study the two-dimensional flow field of yield-stress fluids within and around vane tools with N=3 to 24 arms for a wide range of Bingham numbers, B. This allows for accurate calculations of the velocity and stress fields around vanes with various geometries, as well as direct comparison to experimental observations of the output torque measured by a rheometer, revealing sources of variation and error. We describe the impact of the vane structure on the fluid velocity field, from few-arm cruciform vanes (N < 6) that significantly perturb the flow away from ideal azimuthal kinematics, to many-arm fractal vanes (N > 12) in which the internal structural features are successfully "cloaked" by a yield surface. This results in the shearing of an almost-circular ring of viscoplastic fluid that is indistinguishable from the annular ring of fluid deformed around a slip-free rotating cylindrical bob and leads to more accurate rheometric measurements of the material flow curve.

Introduction

Measuring the flow properties (rheometry) of yield-stress fluids—such as concrete, mayonnaise, or drilling mud—is notoriously difficult. These materials behave as solids until a specific "yield stress" is applied. In standard rheometers using smooth cylindrical bobs, these fluids often "slip" at the wall, sliding like a solid block instead of shearing. This leads to massive errors in measuring viscosity and yield stress.

To address this, researchers use "vanes" (tools with blades) to trap the fluid and force it to shear against itself. However, traditional cruciform vanes introduce a new problem: they are not circular. Traditional analysis assumes the vane acts like a virtual cylinder (the "Couette analogy"), but the flow around 4 or 6 blades is actually quite non-axisymmetric. This paper quantifies these errors and validates a new generation of "fractal" vanes designed to provide the best of both worlds: no slip and perfectly circular flow.

Vane-and-cup rheometer setup (a) and various vane geometries (b, c), including traditional cruciform (N=4, 6) and fractal (N=12F, 24F) designs.

Computational Methodology

The study employs a planar two-dimensional computational analysis using an adaptive augmented Lagrangian scheme. Standard numerical methods often struggle with yield-stress fluids because the transition from "solid" to "liquid" is mathematically singular. The augmented Lagrangian method allows for solving "unregularized" constitutive equations, avoiding the need for high-viscosity approximations to simulate the solid-to-liquid transition.

Crucially, the authors utilized an adaptive meshing algorithm that automatically refines the computational grid specifically at the yield surface—the boundary between flowing and non-flowing fluid. This ensures high spatial resolution where it matters most for capturing the sharp gradients inherent in viscoplastic flows. The models were tested across a wide range of Bingham numbers ($B$), representing the ratio of yield stress to viscous stress, using both Bingham plastic and Herschel-Bulkley constitutive models.

Comparison of standard (left) and adaptive (right) computational meshes, demonstrating refinement at the yield surface for a fractal vane geometry.

Key Insights and Results

The "Cloaking" Effect

The paper's most significant finding is the "cloaking" effect. As the number of vane arms increases ($N \ge 12$), the fluid trapped between the arms rotates as a rigid plug. The yield surface becomes almost perfectly circular, effectively "cloaking" the complex internal vane geometry. To the rest of the fluid, the tool appears as a smooth, slip-free cylinder. This results in nearly ideal viscometric flow, which is essential for accurate rheology.

Vane vs. Cup Slip

Simulations demonstrate that slip on the vane surface itself has almost zero impact on measured torque because the fluid is mechanically trapped between the arms. However, slip on the smooth outer cup wall remains a major source of error. This highlights that a vane tool must be paired with a textured or "ribbed" cup to ensure measurement accuracy.

Accuracy and Validation

Fractal vanes with 24 arms were shown to reduce measurement errors from over 15% (standard 4-arm vane) to less than 5%. The authors also confirmed the accuracy of a new torque-to-stress conversion factor ($S_N$) that accounts for the number of vane arms, removing the need for the empirical "Newtonian calibration" required by the old Couette analogy.

Conclusion

This computational analysis proves that many-armed fractal vanes are superior to traditional cruciform designs for measuring yield-stress fluids. By creating a virtual circular boundary, they eliminate wall slip while maintaining measurement accuracy. While the current 2D study provides a rigorous foundation, future work involving 3D simulations could better capture complex secondary flows and "end effects" at the tool tips. Additionally, practical considerations like potential particle clogging in fine fractal features warrant further experimental investigation.

Summary updated

BibTeX
@article{chaparian2022computational,
  author = {Emad Chaparian and Crystal E. Owens and Gareth H. McKinley},
  title = {{Computational rheometry of yielding and viscoplastic flow in vane-and-cup rheometer fixtures}},
  journal = {Journal of Non-Newtonian Fluid Mechanics},
  year = {2022},
  doi = {10.1016/j.jnnfm.2022.104857},
}