Compressed Sensing on Multi-pinhole Collimator SPECT Camera for Sentinel Lymph Node Biopsy

Carlo Seppi, Uri Nahum, Peter A. von Niederhäusern, Simon Pezold, Michael Rissi, Stephan K. Haerle, Philippe C. Cattin

Abstract

State-of-the-art imaging devices for sentinel lymph node biopsy are either a 1-dimensional gamma probe or more recently 2-dimensional gamma cameras that locate the sentinel lymph node. These devices, however, share difficulties when multiple lymph nodes are close-by and do not allow the estimation of the distance to the lymph nodes, as the tracer activation is projected either to a 1- or 2-dimensional image plane. We propose a method, which reconstructs the tracer distribution using a single image of the detector resulting from a multi-pinhole collimator. Applying standard image processing tools on the detector’s image leads to a reduced, sparse system. Thus, we propose an efficient and reliable compressed sensing strategy, to reconstructs the 3-dimensional tracer distribution using a multi-pinhole collimator and a single detector image. This approach enables better estimation of lymph nodes position and improves the differentiation of close-by lymph nodes.
BibTeX
@inproceedings{seppi2017compressed,
  author = {Carlo Seppi and Uri Nahum and Peter A. von Niederhäusern and Simon Pezold and Michael Rissi and Stephan K. Haerle and Philippe C. Cattin},
  title = {{Compressed Sensing on Multi-pinhole Collimator SPECT Camera for Sentinel Lymph Node Biopsy}},
  booktitle = {International Conference on Medical Image Computing and Computer-Assisted Intervention},
  year = {2017},
  url = {https://openreview.net/forum?id=8fWjLWYBNf},
}

Loading related content...

Figure