Comprehensive preclinical evaluation of a multi-physics model of liver tumor radiofrequency ablation
Abstract
Purpose We aim at developing a framework for the validation of a subject-specific multi-physics model of liver tumor radiofrequency ablation (RFA). Methods The RFA computation becomes subject specific after several levels of personalization: geometrical and biophysical (hemodynamics, heat transfer and an extended cellular necrosis model). We present a comprehensive experimental setup combining multimodal, pre- and postoperative anatomical and functional images, as well as the interventional monitoring of intra-operative signals: the temperature and delivered power. Results To exploit this dataset, an efficient processing pipeline is introduced, which copes with image noise, variable resolution and anisotropy. The validation study includes twelve ablations from five healthy pig livers: a mean point-to-mesh error between predicted and actual ablation extent of 5.3 ± 3.6 mm is achieved. Conclusion This enables an end-to-end preclinical validation framework that considers the available dataset.
BibTeX
@article{audigier2017comprehensive,
author = {Chloé Audigier and Tommaso Mansi and Hervé Delingette and Saikiran Rapaka and Tiziano Passerini and Viorel Mihalef and Marie-Pierre Jolly and Raoul Pop and Michele Diana and Luc Soler and Ali Kamen and Dorin Comaniciu and Nicholas Ayache},
title = {{Comprehensive preclinical evaluation of a multi-physics model of liver tumor radiofrequency ablation}},
journal = {Int. J. Comput. Assist. Radiol. Surg.},
year = {2017},
doi = {10.1007/s11548-016-1517-x},
}Ask Althea to put this paper to the test.
Loading related content...