Comparing time and frequency domain estimation of neonatal respiratory\n rate using pressure-sensitive mats

Shermeen Nizami, Amente Bekele, Mohamed Hozayen, Kim Greenwood, JoAnn Harrold, James R. Green · DOI

TL;DR

This paper investigates the use of pressure-sensitive mats (PSMs) for estimating respiratory rate (RR) in neonates, a challenging area with limited research. They found that frequency-domain analysis of PSM data, especially after preprocessing, significantly outperforms time-domain methods, achieving near-perfect RR estimation with a confidence metric.

What Problem Does This Paper Address?

Estimating respiratory rate (RR) is crucial for monitoring patient health, particularly in vulnerable populations like neonates. While pressure-sensitive mats (PSMs) have shown promise for adult RR monitoring, their application in neonatal intensive care units (NICUs) is underexplored. Neonates present unique challenges due to their smaller size, lower pressure ranges, and susceptibility to various artifacts (e.g., grunting, electrical noise, patient movement). Existing non-invasive methods often lack spatial resolution or are difficult to integrate into NICU workflows. This research aims to bridge this gap by evaluating the effectiveness of PSM technology for neonatal RR estimation, comparing different signal processing techniques.

Methods

The study compares two primary methods for RR estimation from PSM data: time-domain and frequency-domain analysis.

  • Time-Domain Analysis: This method involves counting the number of times the pressure signal crosses a predefined threshold. The threshold is set based on the data's percentile values. RR is then calculated based on these crossings.
  • Frequency-Domain Analysis: This approach utilizes the Fast Fourier Transform (FFT) to analyze the signal's frequency components. The fundamental frequency (peak with the highest power) is identified as the RR. A confidence measure (Lc) is also derived from the ratio of the highest to the second-highest peaks in the periodogram, indicating the reliability of the RR estimation.

Crucially, the study also explores the impact of signal preprocessing on both methods. The preprocessing steps include:

  • Normalization: Removing DC bias caused by static pressure.
  • Detrending: Eliminating slow signal fluctuations due to sensor drift.
  • Median Filtering: Smoothing the signal to reduce high-frequency noise and recover the breathing pattern.

Datasets

The study utilized data collected from bench testing with a neonatal patient simulator, "SimNewB." This simulator was placed on two different types of mattresses (overhead warmer and crib mattress) and subjected to various experimental conditions:

  • Breathing Patterns: Normal breathing vs. breathing with grunting.
  • Simulator Position: Supine vs. prone.
  • Mattress Type: Overhead warmer vs. crib mattress.

The simulator was programmed to breathe at three distinct rates: 45, 60, and 75 breaths per minute (bpm), which fall within typical neonatal ranges. A total of twenty trials were conducted, with data acquired at a sampling rate of 20 frames/sec. This study does not appear to introduce a new benchmark dataset but rather uses a controlled simulation environment to evaluate existing methods.

Key Insights and Ideas

  • Frequency Domain Dominance: The frequency-domain approach consistently outperformed the time-domain analysis, especially after preprocessing. This suggests that analyzing the spectral content of the PSM signal is more robust for capturing the subtle respiratory patterns in neonates.
  • Preprocessing is Key: Normalizing, detrending, and median filtering significantly improved RR estimation accuracy for both time and frequency domains. This highlights the importance of mitigating artifacts for reliable PSM-based monitoring.
  • Confidence Metric (Lc): The introduction of the Lc metric in the frequency domain provides a valuable measure of confidence in the RR estimation. This is particularly useful when the error is zero, allowing for differentiation between experimental conditions.
  • Grunting's Impact: While grunting introduced noise and increased error in the time-domain analysis, the frequency-domain approach showed higher confidence (Lc) even with grunting. This suggests that the frequency domain might be able to distinguish between normal breathing and grunting, potentially leading to a novel grunting detection method.
  • Mattress Influence: The firmer and thicker crib mattress resulted in better RR estimation across all pre-processed datasets compared to the overhead warmer mattress, indicating that mattress properties can influence signal quality.

Key Results

  • Frequency Domain: Achieved zero percent error in RR estimation for all pre-processed datasets across all experimental conditions.
  • Time Domain: Achieved a lowest achievable RMS error of 1.57 percent for pre-processed data.
  • Confidence (Lc): Values of Lc remained relatively steady across pre-processed datasets, indicating consistent confidence in the frequency-domain estimations.
  • Position Impact: For raw data, the supine position showed significantly higher error than the prone position. Normalization improved the frequency-domain results for the supine position to zero error.
  • Grunting Impact: Time-domain analysis showed increased error with grunting, while frequency-domain analysis maintained high confidence.

Limitations and Future Work

  • Artifacts: The study acknowledges that it excluded data with artifacts from patient motion and other electrical, mechanical, environmental, and clinical variables. Future work will explore these factors.
  • Real Patient Data: This study used a simulator. The authors plan to assess RR estimation from real patient data in the NICU as part of a larger project.
  • Grunting Detection: The potential for using the difference between time- and frequency-domain estimates to detect grunting is highlighted as a novel avenue for future research.
  • Model-Based Approaches: The authors suggest exploring model-based time-domain approaches for potentially improved performance.
  • Sensor Fusion: Combining PSM data with audio data is proposed as a way to create a robust system for identifying grunting.

Summary updated

Abstract

Pressure-sensitive mats (PSM) have proved to be useful in the estimation of\nrespiratory rates (RR) in adult patients. However, PSM technology has not been\nextensively applied to derive physiologic parameters in infant and neonatal\npatients. This research evaluates the applicability of the capacitive XSensor\nPSM technology to estimate a range of RR in neonatal patient simulator trials\nconducted under several experimental conditions. PSM data are analyzed in both\nthe time and frequency domain and comparative results are presented. For the\nfrequency-domain approach, in addition to estimating RR, a measure of\nconfidence is also derived from the relative height of peaks in the\nperiodogram. The study demonstrates that frequency domain analysis of\nmean-shifted PSM data achieves the best possible RR estimation, with zero\npercent error, as compared to the lowest achievable RMS error of 1.57 percent\nin the time domain. The frequency domain approach outperforms the time domain\nanalysis whether examining raw data or those preprocessed by normalizing,\ndetrending and median filtering.\n
BibTeX
@inproceedings{nizami2017comparing,
  author = {Shermeen Nizami and Amente Bekele and Mohamed Hozayen and Kim Greenwood and JoAnn Harrold and James R. Green},
  title = {{Comparing time and frequency domain estimation of neonatal respiratory\textbackslash\{\}n rate using pressure-sensitive mats}},
  booktitle = {MeMeA},
  year = {2017},
  doi = {10.1109/memea.2017.7985882},
}
Figure