Abstract
Mapping information from photographic images to volumetric medical imaging scans is essential for linking spaces with physical environments, such as in image-guided surgery. Current methods of accurate photographic image to computed tomography (CT) image mapping can be computationally intensive and/or require specialized hardware. For general purpose 3-D mapping of bulk specimens in histological processing, a cost-effective solution is necessary. Here, we compare the integration of a commercial 3-D camera and cell phone imaging with a surface registration pipeline. Using surgical implants and chuck-eye steak as phantom tests, we obtain 3-D CT reconstruction and sets of photographic images from two sources: Canfield Imaging’s H1 camera and an iPhone 14 Pro. We perform surface reconstruction from the photographic images using commercial tools and open-source code for Neural Radiance Fields (NeRF) respectively. We complete surface registration of the reconstructed surfaces with the iterative closest point (ICP) method. Manually placed landmarks were identified at three locations on each of the surfaces. Registration of the Canfield surfaces for three objects yields landmark distance errors of 1.747, 3.932, and 1.692 mm, while registration of the respective iPhone camera surfaces yields errors of 1.222, 2.061, and 5.155 mm. Photographic imaging of an organ sample prior to tissue sectioning provides a low-cost alternative to establish correspondence between histological samples and 3-D anatomical samples.
Introduction
In medical research, particularly within initiatives like the Human BioMolecular Atlas Program (HuBMAP), there is a critical need to register "micro-scale" histological data with "milli-scale" anatomical scans like CT. This registration allows researchers to precisely locate tissue biopsies within the broader context of an organ. However, high-accuracy mapping often relies on expensive, specialized hardware. This study investigates whether low-cost alternatives, specifically smartphone-based photogrammetry combined with Neural Radiance Fields (NeRF), can provide a viable solution for research sites with limited resources.
Methods
The researchers evaluated two distinct photogrammetry pipelines:
- Commercial 3D Camera: The VECTRA H1 by Canfield Technologies, which utilizes a binocular lens system to generate a calibrated 3D mesh.
- Smartphone + NeRF: An iPhone 14 Pro was used to record 1080p video of the specimens. These videos were processed using Neural Radiance Fields (NeRF) via the instant-ngp software to synthesize 3D scenes from 2D frames.
The study utilized three "phantom" objects with varying surface properties: a hip implant (non-reflective), a knee implant (highly reflective), and a chuck-eye steak (biological tissue proxy). Ground truth 3D volumes were obtained using a Philips Vereos PET/CT system. The registration pipeline included centroid and PCA alignment, volumetric scaling (for the NeRF models), and Iterative Closest Point (ICP) refinement to align the photographic surfaces with the CT data.
Results
The performance of each method was dictated by the geometry and surface characteristics of the objects. NeRF proved superior for objects with sharp edges and complex depth changes, such as the surgical implants. In contrast, the commercial H1 camera performed better on the smoother, more uniform surface of the steak.
The success of the registration was quantified using Average Fiducial Error (mm), measured between manually placed landmarks:
| Object | Canfield H1 Error (mm) | iPhone/NeRF Error (mm) |
|---|---|---|
| Hip Implant | 1.747 ± 0.445 | 1.222 ± 0.635 |
| Knee Implant | 3.932 ± 2.127 | 2.061 ± 0.917 |
| Chuck-eye Steak | 1.692 ± 0.639 | 5.155 ± 2.083 |
Discussion and Conclusion
While the iPhone/NeRF approach showed significant promise, particularly for complex geometries where it outperformed a $10,000+ camera system, it faces specific challenges. NeRF is sensitive to light reflections, which can create artifactual points that interfere with the ICP algorithm. Additionally, unlike the pre-calibrated commercial camera, NeRF reconstructions require an external scaling step to determine physical dimensions.
Despite these limitations, the results demonstrate that low-cost photogrammetry is a viable path for mapping histological samples to anatomical atlases. Future work will focus on automating the alignment process and addressing the challenges posed by deformable biological tissues.
Summary updated